Abstract
OBJECTIVES: To establish local diagnostic reference levels (LDRLs) for emergency paediatric head computed tomography (CT) scans performed at a South African (SA) tertiary-level hospital and to compare these with published data. MATERIALS AND METHODS: A retrospective analysis was conducted of volume-based CT dose index (CTDIvol) and dose length product (DLP) data from uncontrasted paediatric head CT scans performed in the Trauma and Emergency Unit of a tertiary-level SA hospital from January to June 2013. A random sample of 30 patients in each of 3 age groups (0-2, >2-5 and >5-10 years) was used. LDRL values were compared with several national DRLs from Europe and Australia. RESULTS: Mean CTDIvol and DLP values were: 30 mGy and 488 mGy.cm for the 0-2 years age group; 31 mGy and 508 mGy.cm for the >2-5 years group, and 32 mGy and 563 mGy.cm for the >5-10 years group, respectively. The mean DLP for 0-2 year-olds was the only parameter outside the range of corresponding published reference data. Stratification into narrower age groupings showed an increase in DLP values with age. CONCLUSION: An institutional review of the head CT scanning technique for emergency studies performed on children less than 2 years of age is recommended. The current study highlights the role of LDRLs in establishing institutional dosimetry baselines, in refining local imaging practice, and in enhancing patient safety. Standard age stratification for DRL and LDRL reporting is recommended.
Introduction
There is a burgeoning global demand for computed tomography (CT). Compared with plain-film radiography, CT accounts for relatively large doses of ionising radiation, with CT exposure currently representing the largest manmade contribution of absorbed dose to the general population. Monitoring of CT radiation dosage is therefore of increasing importance, especially in paediatric imaging, as children are more vulnerable to the harmful effects of ionising radiation; this is particularly true in low- and middle-income countries where there may be constraints on equipment upgrades and maintenance.1,2,3,4,5,6
As with all diagnostic studies, CT scans should be clinically justified, provide potential patient benefit and utilise appropriate imaging protocols, to keep radiation doses as low as reasonably achievable (ALARA).34,78
National DRLs for specific examinations and patient groups are based on dose distributions observed in national surveys, with the third quartile (the level below which 75% of all dose data falls) most commonly adopted. Conversely, by definition, 25% of sampled dose data will be above the DRL, thereby identifying relatively high dosage.8910
For CT examinations, DRLs are defined in terms of two established dose indicators, namely, volume-based computed tomography dose index (CTDIvol) and dose length product (DLP). The CTDIvol represents the average dose per slice, whilst the DLP reflects the total energy absorbed along the scan length, and is the product of the CTDIvoland scan length.1vol and DLP values, which are displayed on the CT workstation and stored with the study images in DICOM format on a picture archiving and communication system (PACS).2
In high-income countries such as those in Europe, North America and Australasia, governments have introduced regulations requiring the establishment and maintenance of DRLs in all radiological clinics and practices. However, in low- and middle-income countries, the establishment of DRLs for X-ray imaging practices has not been widely implemented, mainly owing to the constraints under which radiological professionals operate.11
Although many countries and institutions are implementing quality assurance measures that include establishment of DRLs for common imaging examinations, DRLs have not played a meaningful role in evaluating the quality and safety of South African (SA) radiological services. There is currently no published national DRL data for South Africa.
The aim of the present study was to establish and describe LDRLs for emergency paediatric head CT scans at a tertiary-level South African hospital, and to compare these with DRL data published in Europe and Australia.12,13,14,15
Methodology
Setting
The study was conducted at a 1386-bed tertiary-level SA teaching hospital, which performs approximately 1130 paediatric head CT scans annually. All paediatric emergency scans are performed on the hospital's Trauma and Emergency Unit's scanner, a Somatom Emotion™6 multidetector CT scanner (MDCT) (Siemens, Erlangen, Germany), which utilises automatic tube current modulation (CARE Dose™4D, Siemens, Erlangen Germany). The primary objective of this modulation is to maintain consistent image quality by compensating for patient size and differential attenuation within the scanned body part.9
Patient population
A retrospective audit was undertaken of all uncontrasted emergency paediatric head CT scans performed from January to June 2013. Patients up to 10 years of age were included and stratified by age into groups of younger than 2 years; >2-5 years; and >5-10 years. In the hospital where the study was conducted, the scan protocol for children older than 10 years was determined by patient size. Some children older than 10 were therefore scanned on an adult protocol. To maintain uniformity, this latter age group was therefore excluded. Random samples of 30 patients were evaluated in each age group.9,10
CT protocol and audit
All scans were performed according to the standard institutional Trauma and Emergency Unit paediatric head CT protocol, which images the brain and upper cervical spine, up to and including the level of C2. Scan parameters were based on a scanner-specific 16-cm diameter plastic phantom and are listed in Table 1. The CTDIvol and DLP values for each study were recorded on a customised spreadsheet. Scan length was derived from the quotient of the DLP and CTDIvol.2
Parameters such as tube potential, beam collimation and pitch are specific to the scanning protocol and are outlined in Table 1. In the present study, the scan reference was set at 230 mAs for the paediatric protocol (based on a 16-cm diameter phantom). The reference mAs is a user-specified parameter that drives the automatic adjustment of the tube current, and is usually preset by the vendor application specialist.
For interdisciplinary quality audit purposes, local DRLs were established by calculating mean CTDIvol and DLP values for each age group. Comparisons with international values were performed using published DRL data from Australia, Switzerland, Germany and the United Kingdom (UK).12,13,14,15
Results
Results obtained in the study are summarised in Table 2. Mean CTDIvol values were relatively constant across the age groups, ranging from 30 to 32 mGy. Mean DLP values increased with patient age from 488 to 563 mGy.cm.
Comparison with published DRLs showed that mean CTDIvol values were either lower than, or within the range of, other published third quartile national DRLs.12,13,14,15
For patients under 2 years of age, the mean DLP value (488 mGy.cm) exceeded the third quartile DLP values (270 mGy.cm - 390 mGy.cm) of studies from Australia, Switzerland, Germany and the UK, whilst DLP values for older children fell within the corresponding range.12,13,14,15
Discussion
The finding that the mean institutional CTDIvol values, as listed in Table 2, compare favourably with those published for Australia, Switzerland, Germany and the UK12,13,14,15
In the newborn to 2-years-old age category, the mean DLP was higher than the 75th percentile for the DRLs reported elsewhere.12,13,14,1512,13,14,151612,13,14,15
The comparative component of the present study was limited by lack of international uniformity in age stratification for DRL data. We advocate standardised age stratification to facilitate interpretation and comparison of data. With this in mind, CTDIvol and DLP values from our study were plotted against age in 1-year age increments (Figure 1). CTDIvol values showed small increases to age 4 years, and then stabilised. For CTDIvol, stratification beyond 4 years of age is apparently not required, whilst narrow groupings may be appropriate below 4 years of age. DLP values increase with age for the entire study group (Figure 1). It thus appears that the broad age groupings currently reflecting DLP data in the international literature merit review.
As CTDIvol and DLP are standard parameters computed for each examination on all modern scanners, they represent a readily available resource for ongoing comparative quality assurance at local, regional, national and international levels.9,10
Conclusion
The present study successfully established local DRLs for emergency paediatric CT scans in a hospital operating under resource-constrained conditions. Excellent agreement of these baseline DRLs with international values attested to the safety and efficiency of institutional practice. Furthermore, the role of LDRLs in establishing institutional dosimetry baselines that would optimise local imaging practice, and enhance patient safety, is also highlighted. If a multinational survey of several South African hospitals were to follow the present article, with inclusion of teenagers and adults, it may provide sufficient appropriate data to establish national DRLs. Standard international age stratification for paediatric DRLs are recommended.
Competing interests
The authors declare that they have no financial or personal relationships which may have inappropriately influenced them in writing this article.
Authors' contributions
Z.V. (Stellenbosch University) was project leader, wrote the study protocol, obtained ethical approval, collected data, analysed the data and wrote the article. R.D.P. and J.M.A. (Stellenbosch University) contributed to design, reviewed and gave input on the final article. W.G. (Stellenbosch University) was the supervisor, reviewed and gave input on the final article.
Acknowledgements
The authors give special thanks to IT specialist, Deborah Purdy, for assistance in data mining from the Tygerberg Hospital PACS, and Anne-Marie du Plessis for valuable input regarding clinical protocols for paediatric CT at Tygerberg Hospital.
References
- 1. American Association of Physicists in Medicine. Report of AAPM task group 23. The measurement, reporting, and management of radiation dose in CT. College Park, MD: American Association of Physicists in Medicine; 2008.
- 2. Coursey CA, Frush DP. CT and radiation: What radiologists should know? Appl Radiol. 2008;37:22-29.
-
3. Sadetzki S, Mandelzweig L. Childhood exposure to external ionising radiation and solid cancer risk. Br J Cancer. 2009;100:1021-1025. PMID: 19337255,http://dx.doi.org/10.1038/sj.bjc.6604994
» http://dx.doi.org/10.1038/sj.bjc.6604994 - 4. European Commission. Guidance on diagnostic reference levels (DRLs) for medical exposures. c1999 [cited 2014 Apr 25].
- Available from:http://ec.europa.eu/energy/nuclear/radiation_protection/doc/publication/109_en.pdf
- 5. International Commission on Radiological Protection (IRCP). Radiological protection in paediatric diagnostic and interventional radiology. IRCP publication 121. c2013 [cited 2014 Apr 25].
- Available from: http://www.icrp.org/publication.asp?id=ICRP%20Publication%20121
-
6. Brenner DJ, Hall EJ. Computed tomography - An increasing source of radiation exposure. N Engl J Med. 2007;357:2277-2284. PMID: 18046031,http://dx.doi.org/10.1056/NEJMra072149
» http://dx.doi.org/10.1056/NEJMra072149 - Available from: http://www.icrp.org/docs/2001_ann_rep.pdf
-
8. Brady Z, Ramanauskas F, Cain TM, Johnston PN. Assessment of paediatric CT dose indicators for the purpose of optimisation. Br J Radiol. 2012;85:1488-1498. PMID: 22844033, http://dx.doi.org/10.1259/bjr/28015185
» http://dx.doi.org/10.1259/bjr/28015185 -
Available from: http://www.ipem.ac.uk/Publications/IPEMReportSeries.aspx
» http://dx.doi.org/10.1259/bjr/93277434 -
10. Al Suwaidi JS, Al Balooshi LG, Al Awadhi HM, et al. Continuous monitoring of CT dose indexes at Dubai Hospital. Am J Roentgenol. 2013;201:858-864. PMID: 24059376, http://dx.doi.org/10.2214/AJR.12.10233
» http://dx.doi.org/10.2214/AJR.12.10233 - 11. World Bank Countries and Economies. c2014 [cited 2014 Nov 31].
- Available from: http://data.worldbank.org/country
- Available from: http://www.arpansa.gov.au/services/ndrl/paediatric.cfm
-
13. Verdun FR, Gutierrez D, Vader JP, et al. CT radiation dose in children: A survey to establish age-based diagnostic reference levels in Switzerland. Eur Radiol. 2008;18:1980-1986. PMID: 18389242, http://dx.doi.org/10.1007/s00330-008-0963-4
» http://dx.doi.org/10.1007/s00330-008-0963-4 - 14. Galanski M, Nagel HD, Stamm G, Paediatric CT exposure practice in the Federal Republic of Germany: Results of a nationwide survey in 2005-2006. Hannover: Medizinische Hochschule. c2007 [cited 2014 Apr 30].
- Available from:https://www.mhhannover.de/fileadmin/kliniken/diagnostische_radiologie/download/Report_German_Paed-CT-Survey_2005_06.pdf
-
16. Strauss KJ, Goske MJ, Kaste SC, et al. Image gently: Ten steps you can take to optimize image quality and lower CT dose for paediatric patients. Am J Roentgenol. 2010;194:868-873. PMID: 20308484, http://dx.doi.org/10.2214/AJR.09.4091
» http://dx.doi.org/10.2214/AJR.09.4091
Accepted: 07 Oct. 2015
Published: 30 Nov. 2015