Open-access An audit of the adequacy of contrast enhancement in CT pulmonary angiograms in a South African tertiary academic hospital setting

Abstract

BACKGROUND: Undiagnosed pulmonary embolism carries high mortality and morbidity. Computed tomography pulmonary angiogram (CTPA) is the diagnostic method of choice for accurate diagnosis. Inadequate contrast opacification is the second most common cause of indeterminate CTPAs. OBJECTIVES: Audit the adequacy of CTPA contrast enhancement and determine whether inadequate enhancement is affected by the size and site of the intravenous cannula, flow rate, contrast volume, contrast leakage and day shift versus after hours services. METHOD: Retrospective and prospective audits of the adequacy of contrast enhancement of CTPAs at the Charlotte Maxeke Johannesburg Academic Hospital were conducted using the Royal College of Radiologists guidelines (≤ 11% of studies with < 210 HU). Protocol variables were collected prospectively from questionnaires completed by radiographers performing the CTPAs. Adequate versus inadequate groups were analysed. RESULTS: A total of 63 (retrospective) and 130 (prospective) patients were included with inadequate contrast enhancement rates of 19% (12/63) and 20.8% (27/130), respectively. The majority of CTPAs were performed during the day 56.2% (73/130) with a 20G cannula 66.2% (86/130) in the forearm 33.8% (44/130) injecting 100 mL - 120 mL contrast 43.1% (56/130) at 3 mL/s 63.1% (82/130). The median flow rate (3 mL/s) and contrast volume (80 mL) were identical in both adequate and inadequate groups, while the remaining variables showed no statistical difference. CONCLUSION: The rate of inadequately enhanced CTPAs in this study was high. The protocol variables did not have a significant influence on the rate of inadequate enhancement. Further research, particularly using flow rates > 4 mL/s, is required for protocol optimisation

contrast enhancement; audit; CT pulmonary angiogram; pulmonary embolism; flow rate


Introduction

Accurate diagnosis of pulmonary embolism (PE) is imperative as untreated or undiagnosed PE carries a high mortality and morbidity.12,344,56

The Royal College of Radiologists (RCR) recommend audits of CTPAs in order to limit indeterminate CTPAs due to inadequate enhancement. The RCR guideline (2013) advocates that a maximum of 11.0% of CTPAs can have inadequate contrast enhancement (< 210 Hounsfield units [HU]).78,9,10,119,11

To the best of our knowledge, no such audit has been performed in South Africa. This audit primarily assessed the adequacy of contrast enhancement of CTPA examinations at a South African tertiary radiology department and secondarily assessed the influence of certain technical factors.

Materials and methods

A single-centre retrospective and prospective audits were performed at a large tertiary hospital in South Africa, which included routine in- and outpatient CTPA referrals from this hospital as well as from a network of referral hospitals and clinics.

The retrospective audit was conducted using all consecutive CTPA studies for suspected PE in adult patients (18 years or older) from 01 December 2019 to 31 December 2019, while the prospective audit included all consecutive CTPA studies from 01 January 2020 to 31 March 2020. Informed consent was obtained from the patients for the prospective audit. Additionally, the prospective audit incorporated a questionnaire that was completed by the attending radiographer acquiring the scan, which included technical factors such as intravenous (IV) cannula size, location of the IV cannula, flow rate of the injector, volume of contrast, presence of contrast leakage and the time at which the scan was acquired. Data from the radiographer checklists were captured.

CT pulmonary angiograms were performed on either a Phillips 64 slice Brilliance, 128 slice Ingenuity or Siemens 64 slice Somatom CT scanner. Examinations were performed with the patient in the supine position with arms placed above the head and scans acquired in a cranio-caudal direction, from the lung apices to the mid-low liver. Patients were instructed to inspire and hold their breath, after which scanning was performed at full inspiration. A dose of 80 mL - 100 mL 350 mg iodine/mL of non-ionic iodinated intravenous contrast material (Omnipaque) was administered through the IV cannula, via a Guerbet Optivantage pump injector at a rate determined by the radiographer (preferably > 3 mL/s), depending on the IV-line size and position. Bolus tracking was used with a region of interest (ROI) centred within the main pulmonary artery (MPA), at the level of its bifurcation, with scanning to be initiated after a 6.2 s delay once a level of 110 HU was reached within the ROI.

All images were viewed on a Philips Enterprise PACS workstation. The contrast enhancement of every CTPA was assessed by placing a circular ROI within the MPA at its largest axial diameter on each study at a slice thickness of 1 mm. The ROI diameter was 50% of the vessel diameter. The HU obtained at this level was recorded as an average of three measurements (Figure 1). Hounsfield unit readings in the MPA were captured, and the HU was further categorised as adequate (≥ 210 HU) or inadequate (< 210 HU).

Statistical analysis

Descriptive statistics (frequencies and percentages) were calculated for categorical data, and medians and percentiles for numerical data. The Shapiro-Wilk test was used to investigate whether numerical variables followed a normal distribution. The Chi-Squared or Fisher's exact tests were used to compare frequencies of adequate versus inadequate HU values. The Mann-Whitney U-test was used to compare median values for the two independent groups: adequate HU versus inadequate HU. A significance level (α) of p < 0.05 was used. All data analysis was performed using SAS version 9.2.

Ethical considerations

Ethics approval was provided by the Committee of Human Research of the Faculty of Health Sciences, University of the Witwatersrand (ethics clearance number M190454). Informed consent was obtained from the patients for the prospective audit. The identifying data of patients were anonymised and stored on a password-protected computer.

Results

A total of 63 (retrospective) and 149 patients (prospective) were included. Nineteen CTPAs from the prospective audit were excluded due to incomplete informed consent. Inadequate contrast enhancement was identified in 19% (12/63) of the retrospective and 20.8% (27/130) of the prospective cases.

The majority of CTPAs were performed with a 20 G IV cannula (86/130, 66.2%), mostly in the forearm (44/130, 33.8%), followed by the hand and antecubital fossa. The IV cannula sizes were not significantly different between the adequate and inadequate contrast enhancement groups (p = 0.20). Similarly, the IV cannula sites were not significantly different between the two groups (p = 0.65). (Table 1).

The majority of CTPAs were performed with 100 mL - 120 mL of contrast (56/130, 43.1%) (Figure 2) at 3 mL/s (82/130, 63.1%) (Figure 3). Both adequate and inadequate groups had a median flow rate of 3 mL/s, while both groups had a median contrast volume of 80 mL, with no significant differences (p = 0.94 and 0.55, respectively) (Table 2). One hundred and twenty-four studies were performed at < 4 mL/s compared to six studies at ≥ 4 mL/s (Table 1). The number of inadequate contrast opacified studies between flow rates of ≥ 4 mL/s and < 4 mL/s was not significantly different (p = 0.59) (Table 1).

Table 2

The majority of the CTPAs were performed during the day (73/130, 56.2%). Additionally, the time of the scan acquisition, that is, day versus afterhour shifts and presence of contrast leakage made no significant difference to the rate of inadequate contrast enhancement (p = 0.21 and 0.60, respectively) (Table 1).

Discussion

This audit established that the rate of inadequate CTPAs was higher than the international guideline set by the RCR78,9,10,11

The effect of the variables (position of IV cannula, size of IV cannula, flow rate, volume of contrast, time of study and presence of contrast leakage) on the rate of inadequate contrast enhancement was found to be non-significant. These findings provide some insights into what does not contribute to inadequate contrast opacification but have to be interpreted in light of the limitations subsequently discussed.

There were higher rates of inadequate contrast enhancement when IV cannulas were positioned in the hand (28.6%), compared to 21.6% in the antecubital fossa and 15.9% in the forearm; however, this small difference was not significant (p = 0.65). This is comparable to findings by Marshall et al.12137,12,141516,171413,18,19,202122

This study only had six CTPAs with a flow rate of ≥ 4 mL/s, and although none of them had inadequate contrast opacification, we could not prove any significant difference in the rate of inadequately contrast-enhanced CTPAs between this and flow rates < 4 mL/s (p = 0.59). The significance is limited by the small number of CTPAs with flow rates ≥ 4 mL/s. Several international studies have achieved lower rates of inadequate contrast opacification, compared to ours, utilising flow rates higher than 4 mL/s. Lloyd et al.11232425

The contrast volume did not have any significant effect on the rate of inadequate contrast opacification. This is similar to the finding by Goble et al.171624,26

Other possible explanations for the high rate of inadequately opacified studies can include factors that were not evaluated in this audit such as cardiovascular status, body habitus, patient age, respiratory rate and IV cannula size to access site mismatch. Scanning at maximal inspiration, as utilised in our protocol, has shown inferior contrast enhancement compared to minimal inspiration and expiration.27,28,2927,30

A circulation adjusted protocol with scan initiation of 10 s after bolus triggering at 150 HU or a fixed scan delay of 19 s after injection was suggested as adequate.3117

Limitations of the study

The small sample size limited the significance, especially the low number of CTPAs with high flow rates (≥ 4 mL/s) and a single CTPA with a 22 G IV cannula. As this was an audit, variables could not be controlled, which led to an unequal distribution within variables. This audit did not account for patient age, weight, cardiovascular status and respiratory motion, which have all been shown to affect contrast enhancement.4,13

The use of one reader limits reliability; however, the HU measurement in the MPA is an objective measurement that requires minimal expertise and the addition of the retrospective audit improved the reliability. The amount of contrast leakage was not quantified.

Recommendations

The CTPA protocol at this hospital should be assessed and adjusted to achieve a lower rate of inadequate contrast enhancement. This will lead to less indeterminate studies, less repeated studies and improved patient care. Alterations that have been advised and shown to be helpful include scanning at minimal inspiration or expiration,27,28,297

The use of protocols adjusting contrast volume according to body weight and using a multiphasic injection can be investigated. Following any changes to current practice, further audits should be performed to assess any subsequent improvement in the rate of inadequate opacification.

We recommend that similar audits be performed at other institutions to ensure that adequate contrast enhancement is achieved and local guidelines are established.

Conclusion

This audit proves that the rate of inadequately enhanced CTPAs at this hospital, according to the RCR recommendations, was high. Causes for inadequate contrast enhancement are multifactorial, and we could not prove any significant impact with regard to the size of IV cannula, position of IV cannula, volume of contrast, time of study acquisition and presence of contrast leakage on the rate of inadequately opacified CTPAs. Although evidence from this study is limited, the use of high flow rates (≥ 4 mL/s) should be explored in an attempt to reduce inadequate opacification.

Competing interests

The authors declare that they have no financial or personal relationships that may have inappropriately influenced them in writing this article.

Authors' contributions

D.J.B. was the guarantor of the integrity of the entire study and responsible for the literature review, data analysis, result compilation and manuscript preparation. H.M. contributed to study concepts, design and manuscript editing.

Funding information

This research did not receive any specific grant from funding agencies in the public, commercial or not-for-profit sectors.

Data availability

Data that support the findings of this study are available upon request from the corresponding author D.J.B.

Disclaimer

The views and opinions expressed in this article are those of the authors and do not necessarily reflect the official policy or position of any affiliated agency of the authors.

Acknowledgements

The authors would like to thank the CT radiography staff for performing the studies at the Charlotte Maxeke Johannesburg Hospital and completing the questionnaires as well as Maryn Viljoen for assistance with the statistical analysis. We further acknowledge the British Institute of Radiology for clarifying that this work was previously presented as an electronic poster at the British Institute of Radiology Congress 2020 and that the authors of this article maintain the right to publish this work.

References

  • 1.Sin D, McLennan G, Rengier F, et al. Acute pulmonary embolism multimodality imaging prior to endovascular therapy. Int J Cardiovasc Imaging. 2021;37(1):343-358. https://doi.org/10.1007/s10554-020-01980-9
    » https://doi.org/10.1007/s10554-020-01980-9
  • 2.Patel H, Sun H, Hussain AN, Vakde T. Advances in the diagnosis of venous thromboembolism: A literature review. Diagnostics (Basel, Switzerland). 2020;10(6):365. https://doi.org/10.3390/diagnostics10060365
    » https://doi.org/10.3390/diagnostics10060365
  • 3.Moore AJE, Wachsmann J, Chamarthy MR, Panjikaran L, Tanabe Y, Rajiah P. Imaging of acute pulmonary embolism: An update. Cardiovasc Diagn Ther. 2018;8(3):225-243. https://doi.org/10.21037/cdt.2017.12.01
    » https://doi.org/10.21037/cdt.2017.12.01
  • 4.Jones SE, Wittram C. The indeterminate CT pulmonary angiogram: Imaging characteristics and patient clinical outcome. Radiology. 2005;237(1):329-337. https://doi.org/10.1148/radiol.2371041520
    » https://doi.org/10.1148/radiol.2371041520
  • 5.Bates DDB, Tkacz JN, LeBedis CA, Holalkere N. Suboptimal CT pulmonary angiography in the emergency department: A retrospective analysis of outcomes in a large academic medical center. Emerg Radiol. 2016;23(6):603-607. https://doi.org/10.1007/s10140-016-1425-y
    » https://doi.org/10.1007/s10140-016-1425-y
  • 6.Statistics South Africa. Mortality and causes of death in South Africa, 2016: Findings from death notification [homepage on the Internet]. Statistical release P0309.3. Pretoria: Statistics South Africa; 2013 [cited 2019 Jan 15]. Available from: http://www.statssa.gov.za
    » http://www.statssa.gov.za
  • 7.Beattie A, Muller M. Adequate contrast enhancement of CT pulmonary angiograms [homepage on the Internet]. The Royal College of Radiologists. [cited 2021 Jun 8]. Available from: https://www.rcr.ac.uk/audit/adequate-contrast-enhancement-ct-pulmonary-angiograms
    » https://www.rcr.ac.uk/audit/adequate-contrast-enhancement-ct-pulmonary-angiograms
  • 8.Afzal I, Colapinto M, Haider E. CT pulmonary angiogram audit [homepage on the Internet]. [cited 2019 Jan 19]. Available from: https://car.ca/wp-content/uploads/AP004_CT_Pulmonary_Angiography_Audit_Afzal.pdf
    » https://car.ca/wp-content/uploads/AP004_CT_Pulmonary_Angiography_Audit_Afzal.pdf
  • 9.Abdullah M, Umer US, Khan AN, et al. Adquacy of CT pulmonary angiogram for diagnosis of pulmonary embolism. Pakistan J Med Heal Sci. 2021;15(4):2020-2021.
  • 10.Bin Musarraf AA, Sapundzieski M, Rogers T. Adequate contrast enhancement of CT pulmonary angiograms [homepage on the Internet]. [cited 2021 Jun 9]. Available from: https://cdn.technologynetworks.com/ep/pdfs/adequate-contrast-enhancement-of-ct-pulmonary-angiograms.pdf
    » https://cdn.technologynetworks.com/ep/pdfs/adequate-contrast-enhancement-of-ct-pulmonary-angiograms.pdf
  • 11.Lloyd S, Sahu A, Riordan R. Diagnosis of pulmonary embolism by computed tomographic pulmonary angiography with and without optimal contrast enhancement: A prospective single centre audit. Hong Kong J Radiol. 2012;15(3):162-169.
  • 12.Marshall T, Chen NM, Nguyen E, Slattery DE, Zitek T. Rethinking intravenous catheter size and location for computed tomography pulmonary angiography. West J Emerg Med. 2019;20(2):244-249. https://doi.org/10.5811/westjem.2018.11.40930
    » https://doi.org/10.5811/westjem.2018.11.40930
  • 13.Roggenland D, Peters SA, Lemburg SP, Holland-Letz T, Nicolas V, Heyer CM. CT angiography in suspected pulmonary embolism: Impact of patient characteristics and different venous lines on vessel enhancement and image quality. Am J Roentgenol. 2008;190(6):351-359. https://doi.org/10.2214/AJR.07.3402
    » https://doi.org/10.2214/AJR.07.3402
  • 14.Rubin GD, Browne LP, Fishman EK, et al. ACR-NASCI-SIR-SPR Practice parameter for the performance and interpretation of body computed tomography angiography (CTA) [homepage on the Internet]. 2016 [cited 2021 Jun 9]. Available from: https://www.acr.org/-/media/acr/files/practice-parameters/body-cta.pdf
    » https://www.acr.org/-/media/acr/files/practice-parameters/body-cta.pdf
  • 15.Schoellnast H, Deutschmann HA, Berghold A, Fritz GA, Schaffler GJ, Tillich M. MDCT angiography of the pulmonary arteries: Influence of body weight, body mass index, and scan length on arterial enhancement at different iodine flow rates. Am J Roentgenol. 2006;187(4):1074-1078. https://doi.org/10.2214/AJR.05.0714
    » https://doi.org/10.2214/AJR.05.0714
  • 16.Chen M, Mattar G, Abdulkarim JA. Computed tomography pulmonary angiography using a 20% reduction in contrast medium dose delivered in a multiphasic injection. World J Radiol. 2017;9(3):143. https://doi.org/10.4329/wjr.v9.i3.143
    » https://doi.org/10.4329/wjr.v9.i3.143
  • 17.Goble EW, Abdulkarim JA. CT pulmonary angiography using a reduced volume of high-concentration iodinated contrast medium and multiphasic injection to achieve dose reduction. Clin Radiol. 2014;69(1):36-40. https://doi.org/10.1016/j.crad.2013.07.023
    » https://doi.org/10.1016/j.crad.2013.07.023
  • 18.Gossner J. Feasibility of computed tomography pulmonary angiography with low flow rates. J Clin Imaging Sci. 2012;2(3):57. https://doi.org/10.4103/2156-7514.100999
    » https://doi.org/10.4103/2156-7514.100999
  • 19.Wittram C. How I do it: CT pulmonary angiography. Am J Roentgenol. 2007;188(5):1255-1261. https://doi.org/10.2214/AJR.06.1104
    » https://doi.org/10.2214/AJR.06.1104
  • 20.Ramadan SU, Kosar P, Sonmez I, Karahan S, Kosar U. Optimisation of contrast medium volume and injection-related factors in CT pulmonary angiography: 64-slice CT study. Eur Radiol. 2010;20(9):2100-2107. https://doi.org/10.1007/s00330-010-1782-y
    » https://doi.org/10.1007/s00330-010-1782-y
  • 21.Schwab SA, Uder M, Anders K, Heinrich MC, Kuefner MA. Peripheral intravenous power injection of iodinated contrast media through 22G and 20G cannulas: Can high flow rates be achieved safely? A clinical feasibility study. Rofo. 2009;181(4):355-361. https://doi.org/10.1055/s-0028-1109181
    » https://doi.org/10.1055/s-0028-1109181
  • 22.Chaturvedi A, Oppenheimer D, Rajiah P, Kaproth-Joslin KA, Chaturvedi A. Contrast opacification on thoracic CT angiography: Challenges and solutions. Insights Imag. 2017;8(1):127-140. https://doi.org/10.1007/s13244-016-0524-3
    » https://doi.org/10.1007/s13244-016-0524-3
  • 23.Menon A, Abdulkarim J. CT pulmonary angiography using 45 mL of iodinated contrast media [homepage on the Internet]. ECR 2018/C-1402. 2018 [cited 2021 Jun 8]. Available from: https://epos.myesr.org/poster/esr/ecr2018/C-1402
    » https://epos.myesr.org/poster/esr/ecr2018/C-1402
  • 24.Hendriks BMF, Kok M, Mihl C, Bekkers SCAM, Wildberger JE, Das M. Individually tailored contrast enhancement in CT pulmonary angiography. Br J Radiol. 2016;89(1061):20150850. https://doi.org/10.1259/bjr.20150850
    » https://doi.org/10.1259/bjr.20150850
  • 25.Ozawa Y, Hara M, Shibamoto Y. The frequency of insufficient contrast enhancement of the pulmonary artery in routine contrast-enhanced chest CT and its improvement with an increased injection rate: A prospective study. J Thorac Imaging. 2011;26(1):42-47. https://doi.org/10.1097/RTI.0b013e3181ced07a
    » https://doi.org/10.1097/RTI.0b013e3181ced07a
  • 26.Jamali L, Alikhani B, Getzin T, Ringe K, Wacker F, Raatschen H-J. Arterial attenuation in individualized computed tomography pulmonary angiography injection protocol adjusted based on the patient's body mass index. J Res Med Sci. 2020;25(1):94. https://doi.org/10.4103/jrms.JRMS_690_19
    » https://doi.org/10.4103/jrms.JRMS_690_19
  • 27.Li YYJ, Lau KK, Ardley N, Lau T. Efficacy of 'breath holding at ease' during CT pulmonary angiography in the improvement of contrast enhancement in pulmonary arteries. J Med Imaging Radiat Oncol. 2013;57(4):415-422. https://doi.org/10.1111/1754-9485.12039
    » https://doi.org/10.1111/1754-9485.12039
  • 28.Chen YH, Velayudhan V, Weltman DI, et al. Waiting to exhale: Salvaging the nondiagnostic CT pulmonary angiogram by using expiratory imaging to improve contrast dynamics. Emerg Radiol. 2008;15(3):161-169. https://doi.org/10.1007/s10140-007-0695-9
    » https://doi.org/10.1007/s10140-007-0695-9
  • 29.Kay FU, Macedo ACS, Chate RC, et al. Reduction of poor contrast enhancement of the pulmonary artery in computed tomography angiography using an alternative respiratory maneuver. J Thorac Imaging. 2014;29(2):107-112. https://doi.org/10.1097/RTI.0000000000000050
    » https://doi.org/10.1097/RTI.0000000000000050
  • 30.Liu D, Cai X, Che X, Ma Y, Fu Y, Li L. Visibility and image quality of peripheral pulmonary arteries in pulmonary embolism patients using free-breathing combined with a high-threshold bolus-triggering technique in CT pulmonary angiography. J Int Med Res. 2020;48(8):300060520939326. https://doi.org/10.1177/0300060520939326
    » https://doi.org/10.1177/0300060520939326
  • 31.Zhu J, Wang Z, Kim Y, et al. Analysis of contrast time-enhancement curves to optimise CT pulmonary angiography. Clin Radiol. 2017;72(4):340.e9-340.e16. https://doi.org/10.1016/j.crad.2016.11.018
    » https://doi.org/10.1016/j.crad.2016.11.018

Correspondence:
Derik Basson

Received: 18 Nov. 2021
Accepted: 16 Jan. 2022
Published: 24 Mar. 2022

Correspondence

derikjbasson@gmail.com

Publication Dates

  • Publication in this collection
    01 Feb 2023
  • Date of issue

History

  • Received
    18 Nov 2021
  • Accepted
    16 Jan 2022
location_on
Radiological Society of South Africa AOSIS Publishing, Postnet Suite #110, Private Bag X19, Durbanville, Cape Town, Western Cape Province, ZA, 7551, Tel: +27 21 975 2602, Tel: 086 1000 381 - E-mail: publishing@aosis.co.za
rss_feed Acompanhe os números deste periódico no seu leitor de RSS
Ir para o topo