Open-access Implementation and evaluation of the Presto combined qualitative real-time assay for Chlamydia trachomatis and Neisseria gonorrhoeae in Rwanda

Abstract

BACKGROUND: The Presto combined qualitative real-time assay for Chlamydia trachomatis and Neisseria gonorrhoeae (Presto CT/NG PCR assay) is appealing for developing countries, because it can be used with multiple DNA extraction methods and polymerase chain reaction (PCR) platforms. OBJECTIVES: The objective of the study was to implement and evaluate the Presto CT/NG PCR assay at the National Reference Laboratory (NRL) in Kigali, Rwanda, where no real-time PCR assays for the detection of C. trachomatis or N. gonorrhoeae were available. METHODS: The Presto CT/NG PCR assay was first evaluated at the Institute of Tropical Medicine (ITM) in Antwerp, Belgium. Next, NRL laboratory technicians were trained to use the assay on their ABI PRISM 7500 real-time PCR instrument and their competencies were assessed prior to trial initiation. During the trial, endocervical swabs were tested at the NRL, with bi-monthly external quality control testing monitored by the ITM. The final NRL results were evaluated against extended gold standard testing at the ITM, consisting of the Abbott m2000 RealTime System with confirmation of positive results by an in-house real-time PCR assay for C. trachomatis or N. gonorrhoeae. RESULTS: Of the 192 samples analysed using the Presto assay at the NRL, 16 samples tested positive for C. trachomatis and 17 tested positive for N. gonorrhoeae; four of these were infected with both. The sensitivity and specificity of the Presto assay were 93.3% (95% confidence interval [CI]: 68.1% - 99.8%) and 99.4% (95% CI: 96.8% - 100%) for C. trachomatis and 100% (95% CI: 76.8% - 100%) and 98.8% (95% CI: 95.8% - 99.9%) for N. gonorrhoeae. CONCLUSION: C. trachomatis and N. gonorrhoeae testing with the Presto assay was feasible in Kigali, Rwanda, and good performance was achieved.

qPCR; Chlamydia trachomatis; Neisseria gonorrhoeae


Introduction

Chlamydia trachomatis and Neisseria gonorrhoeae infections are the most common sexually transmitted bacterial infections worldwide.11,2C. trachomatis and N. gonorrhoeae infections adequately, they need to be detected early and accurately.3

Highly sensitive and specific nucleic acid amplification assays to identify C. trachomatis and N. gonorrhoeae infections are commercially available.4,5Chlamydia intracellular culture.

Over the past decades, dual detection systems for C. trachomatis and N. gonorrhoeae based on polymerase chain reaction (PCR) have been developed. Among them is the Presto combined qualitative real-time assay for C. trachomatis and N. gonorrhoeae (Presto CT/NG PCR assay; Goffin Molecular Technologies, Houten, the Netherlands). The Presto CT/NG PCR assay is characterised by its flexibility in use of sample types (with or without transport medium), DNA can be isolated using any system and it can be implemented on different real-time PCR amplification platforms.6,7

The assay includes an internal amplification control (IAC) to monitor both nucleic acid extraction and amplification efficiency. Clinical samples may contain inhibitory substances which are not always reliably removed during sample preparation. The specifically designed IAC will not only detect PCR inhibition but also inefficient DNA isolation from each individual sample. The IAC consists of an inactivated Escherichia coli containing a modified genomic DNA fragment with primer binding sites identical to the C. trachomatis and N. gonorrhoeae sequences. Detection of the amplified sequence, which is different from the C. trachomatis and N. gonorrhoeae amplicon fragments is done by a probe sequence with a different reporter dye. The Presto assay also includes a C. trachomatis and a N. gonorrhoeae selector. These selectors facilitate the detection of concurrent infections, especially when the high concentration of one infection (or target) masks the presence of the other. In other words, they are meant to select for either C. trachomatis or N. gonorrhoeae in a clinical sample with a strong positive C. trachomatis or N. gonorrhoeae result. The selector blocks the highly concentrated target present in the specimen and enhances the amplification of the other not-blocked target, if present.

Another concern that has arisen in recent years is the detection of the C. trachomatis Swedish variant strain. This is a C. trachomatis variant with a deletion in the cryptic plasmid.8C. trachomatis Swedish variant.8C. trachomatis Swedish variant strain.

Real-time PCR assays for the detection of C. trachomatis and N. gonorrhoeae were not available at the National Reference Laboratory (NRL) in Kigali, Rwanda, in 2013. We installed the Presto assay at the NRL in the context of a clinical trial (Ring Plus study) funded by the European & Developing Countries Clinical Trials Partnership.9,10

Methods

Ethical considerations

The evaluation of the Presto assay was embedded in the clinical trial (Ring Plus study; ClinicalTrials.gov NCT01796613), which was conducted at Rinda Ubuzima Research Center, Kigali, Rwanda, from May 2013 until March 2014. The clinical trial protocol has been published elsewhere.9

Evaluation at the Institute of Tropical Medicine, Antwerp, Belgium

Prior to the start of the clinical trial in Rwanda, the Presto assay was evaluated at the ITM in Antwerp, Belgium.

A 1:10 dilution series of a C. trachomatis serovar L2 and a N. gonorrhoeae strain in diluted phosphate buffered saline (dPBS) was tested using the Presto assay, as well as an extended gold standard consisting of the Abbott m2000 RealTime System (Lake Forest, Illinois, United States) with confirmation of positive results by in-house real-time PCR assays for C. trachomatis and N. gonorrhoeae. The added value of the selectors in the Presto assay was evaluated on five aliquots containing N. gonorrhoeae strains in dPBS in a final concentration of 3 × 105C. trachomatis serovar L2 strains diluted in dPBS in a final concentration of 9 × 104C. trachomatis L2 (range: 102-2 EB per PCR reaction) and N. gonorrhoeae (range: 3 × 102-2 CFU per PCR reaction). The C. trachomatis selector was added to the highly concentrated N. gonorrhoeae-positive suspensions (3 × 105N. gonorrhoeae selector was added to the highly concentrated C. trachomatis-positive suspensions (9 ×******* 104C. trachomatis (QCMD 2011) and 10 samples for N. gonorrhoeae (QCMD 2010), was also tested with the Presto assay and the extended gold standard. Finally, the analytical specificity for N. gonorrhoeae was evaluated by testing eight different non-gonococcal Neisseria species: N. meningitidis, N. lactamica, N. pharingis, N. animalis, N. caviae, N. ovis, N. subflava and N. perflava.

Specimen collection, processing and transport

Several sexually transmitted infection assays were performed at the clinical trial's baseline visit,10

The dry swabs were eluted in the on-site Rinda Ubuzima laboratory by adding 1.2 mL dPBS directly onto the swabs. The dPBS was prepared by dissolving 0.96 g phosphate buffered saline (PBS) and 7.65 g NaCl in 1L molecular biology water (MBW) and sterilised by filtration. After vortexing the swabs for 15 seconds, two aliquots of 550 µL were prepared: one was stored at 2 °C - 8 °C until transportation in a temperature-controlled cool box with cooling elements to NRL within 1 week of sample collection (or at -20 °C in case transportation was delayed) and the other at -20 °C until transportation in a temperature-controlled dry shipper to the ITM. The Presto assay was performed at the NRL within two working days after sample receipt according to the manufacturer's instructions.

Training and external quality control

ITM and NRL staff wrote a standard operating procedure together, and hands-on training was conducted by ITM staff in the NRL laboratory. The competence of the trained NRL technicians was evaluated using a blind specimen panel consisting of a positive C. trachomatis sample, a positive N. gonorrhoeae sample, a sample positive for both C. trachomatis and N. gonorrhoeae and a negative sample.

External quality control panels were provided by the ITM on a bi-monthly basis throughout the study. Each panel contained a strong and a weak-positive C. trachomatis sample, a strong and a weak-positive N. gonorrhoeae sample, a sample concurrently infected with both C. trachomatis and N. gonorrhoeae and a sample negative for both C. trachomatis and N. gonorrhoeae. All samples with the exception of the weak-positive samples were expected to be correctly identified by the NRL. Weak-positive samples were allowed to be missed: they were included in the panel to encourage the NRL to excel in their competence.

Chlamydia trachomatis and Neisseria gonorrhoeae testing using the Presto assay

The Presto assay targets the C. trachomatis cryptic plasmid and the opa gene of N. gonorrhoeae. Assay procedures were performed in physically separated rooms including a sample preparation room, a DNA-free area, and an amplification room. The QIAamp DNA Mini Kit (Qiagen, Hilden, Germany) was used for the DNA extraction of 400 µL of aliquot, following the buccal swab spin protocol. Before extraction, 5 µL of the IAC was added to every sample tube.

After adding DNA of samples and controls to the master mix, the reaction plate was loaded into the ABI PRISM 7500 instrument (Applied Biosystems™, Foster City, California, United States) with the following settings: fixed threshold, 0.01; activation polymerase, 30 sec at 95 °C; number of cycles, 40; denaturation, 3 sec at 95 °C; annealing, extension and exonuclease activity, 30 sec at 60 °C.

If a positive signal with cycle threshold (Ct) value lower than 35 was obtained for either C. trachomatis or N. gonorrhoeae, the PCR was repeated with 1 µL of a selector for N. gonorrhoeae or C. trachomatis and 9 µL of DNA to check for the presence of the other pathogen.

Specimens were considered negative when the Ct value of the specimen was greater than 40 and IAC was 37 or less, positive when the Ct value of the specimen was 35 or less for C. trachomatis or N. gonorrhoeae with any Ct value of the IAC, and equivocal when the Ct value of the specimen was between 35 and 40 and a Ct value of the IAC was less than 40. Specimens were considered inhibited when the Ct value of the specimen was greater than 40 and the Ct value of the IAC was greater than 37.

The assay kit contained two positive and one negative control for test-run validation which were interpreted according to the manufacturer's instructions (CG 160501, rev05, 10/2011).

Extended gold standard

At the ITM, DNA extraction and subsequent amplification was performed using the Abbott m2000 Real Time System (Lake Forest, Illinois, United States) according to the manufacturer's instructions. C. trachomatis-positive samples were confirmed with an in-house real-time PCR based on the publication by Chen et al.11N. gonorrhoeae-positive samples were confirmed with an in-house real-time PCR based on the publication by Hopkins et al.12

Statistical methods

For both C. trachomatis and N. gonorrhoeae, samples were defined as true positive if identified positive with both the Abbott assay and the in-house real-time PCR. Samples were defined as true negative when found to be negative in the Abbott assay or when a positive result could not be confirmed by the in-house real-time PCR. Prevalence at baseline was calculated for C. trachomatis and N. gonorrhoeae based on the extended gold standard. Sensitivity, specificity, positive predictive value (PPV) and negative predictive value (NPV) of the Presto assay were calculated at baseline with the extended gold standard as reference. Equivocal results in the Presto assay were considered as negative in the above mentioned calculations. Wilson Binomial 95% confidence intervals (CI) were calculated for the sensitivity, specificity, PPV and NPV.

Results

Evaluation at the Institute of Tropical Medicine

Evaluation of the Presto assay against the extended Gold Standard at the ITM showed that the analytical sensitivities of the Presto assay were in the same order of magnitude (100 EB/mL for C. trachomatis, 330 CFU/mL for N. gonorrhoeae versus 86 EB/mL for C. trachomatis, 280 CFU/mL for N. gonorrhoeae).

The analytical sensitivity of the Presto PCR increased by 2 log10 or 100 fold when a C. trachomatis selector was added to the strong-positive N. gonorrhoeae suspensions. The Ct value of the IAC decreased, reducing the inhibition. No difference in sensitivity was obtained by using a N. gonorrhoeae selector on strong-positive C. trachomatis suspensions, although the Ct value of the IAC also decreased. No effect of the use of selectors was obtained when strong C. trachomatis-positive clinical samples were spiked with low concentrations of N. gonorrhoeae or when strong N. gonorrhoeae-positive clinical samples were spiked with low concentrations of C. trachomatis.

The QCMD panel included eight positive and two negative C. trachomatis samples. Six samples tested C. trachomatis-positive by the Presto assay. Of the two low-positive C. trachomatis samples that were not detected by the Presto assay, one was also missed using the extended gold standard. The QCMD panel included a C. trachomatis Swedish variant, which was successfully detected with the Presto assay. The QCMD panel for N. gonorrhoeae included seven positive and three negative N. gonorrhoeae samples, of which one was Neisseria cinerea. Six samples tested positive in the Presto assay. One sample was not tested due to insufficient volume to perform the test. The three N. gonorrhoeae-negative samples tested negative in the Presto assay.

No cross reactivity was detected with the other eight Neisseria species.

Implementation of the Presto assay at the National Reference Laboratory

The results of the blind specimen panel used for competency check were 100% concordant. The bi-monthly external quality control panels also produced satisfactory results: two C. trachomatis weak-positive samples were missed with the Presto assay, and one N. gonorrhoeae-positive sample result was equivocal.

Clinical trial results

Baseline samples from 185 women were tested for C. trachomatis and N. gonorrhoeae by the Presto assay. Additional samples from seven women were tested at follow-up visits. The results are summarised in Table 1.

All study samples (n = 192) were retested according to the extended gold standard algorithm at the ITM. Two samples were excluded from data analysis as they contained PCR inhibitors: one sample was inhibited in the Presto assay and one sample was inhibited in the Abbott assay. The Presto assay missed one C. trachomatis-positive sample, although this was equivocal in the assay. It falsely detected three N. gonorrhoeae and one C. trachomatis. The latter C. trachomatis sample was positive on the Abbott assay but could not be confirmed using the in-house real-time PCR (Table 2).

At baseline, the overall prevalence was 8.1% for C. trachomatis and 7.0% for N. gonorrhoeae. The sensitivity, specificity, PPV and NPV of the Presto assay are summarised in Table 3.

Discussion

We successfully implemented the Presto combined qualitative real-time assay for Chlamydia trachomatis and Neisseria gonorrhoeae at the NRL, Kigali, Rwanda. The implementation was a stepwise process covering hands-on training and writing of standard operating procedures. The competency of the laboratory technicians in the execution of the Presto assay was confirmed through the 100% concordant results obtained for the testing of the blind specimen panel. The testing of the bi-monthly external quality control panels allowed us to monitor and advise the NRL on testing for C. trachomatis and N. gonorrhoeae using the Presto assay. In the end, we were able to successfully install the Presto assay at the NRL and to guarantee the quality of the results during the clinical trial. In addition, the quality control procedures and retesting of samples at ITM may contribute to the method validation that is required for laboratory accreditation. The NRL may also use a similar implementation approach in the event the amplification testing for C. trachomatis and N. gonorrhoeae should be used in other settings such as for diagnosis or for research or clinical trials.

The analytical sensitivity of the Presto assay for C. trachomatis and N. gonorrhoeae was comparable to that of the extended gold standard. Previous versions of PCR assays for N. gonorrhoeae showed cross reactivity and amplification of non-gonorrhoeae Neisseria species,13N. gonorrhoeae detection and resulted in unnecessary treatment. Superfluous antibiotic use increases the risk of untreatable N. gonorrhoeae due to multidrug resistance and should be maximally prevented.14Neisseria species was amplified by the Presto assay, confirming its analytical specificity. However, the PPV for N. gonorrhoeae detection using the Presto assay was under 90%, which necessitates a confirmation test for N. gonorrhoeae as recommended by the International Union against Sexually Transmitted Infections (IUSTI) guidelines.15N. gonorrhoeae false-positive samples.

The testing for C. trachomatis and N. gonorrhoeae described in this manuscript was part of a clinical trial which was conducted according to Good Clinical and Laboratory Practice requirements.9

Our study population was characterised by a high prevalence of both C. trachomatis and N. gonorrhoeae. Almost one quarter of the infections were dual infections, highlighting the need for a dual detection system in this setting. The advantages of a dual detection system are staff time and assay cost reduction while two detection results are delivered. Notwithstanding this study's small sample size, the Presto assay's sensitivity and specificity for C. trachomatis detection was in line with two previous studies.6,7N. gonorrhoeae detection; we did not obtain cross reactions with the non-gonorrhoeae Neisseria species, and we therefore advise that a possible lower specificity should be confirmed in future studies. A total of eight equivocal results (4.2%) were found during this small evaluation. In clinical practice, clinicians should be instructed to collect another sample for retesting. The result will be reported as equivocal in the event of a second equivocal result and conclusion regarding the diagnosis cannot be drawn, or as positive or negative, if the second test result is positive or negative.

To date no reports have been found on the current presence of the C. trachomatis Swedish variant strain in Rwanda or Africa, but it has become a global requirement that newly designed PCR assays for C. trachomatis detection also detect the C. trachomatis Swedish variant.8C. trachomatis target of the Presto assay lies within the C. trachomatis cryptic plasmid, the manufacturer states that the C. trachomatis Swedish variant will be detected. This statement was confirmed by testing a QCMD panel which included a sample with the C. trachomatis Swedish variant.

The laboratory and etiological diagnosis of C. trachomatis and N. gonorrhoeae is scarcely available and problematic in resource-poor settings. To date many settings have adopted the syndrome-based management approach.1617,18C. trachomatis and N. gonorrhoeae based on symptoms and signs.14,1920C. trachomatis and N. gonorrhoeae, none, except one point-of-care test, the Cepheid GeneXpert CT/NG assay, is sufficiently accurate to be useful in an etiological diagnosis.21,22C. trachomatis and N. gonorrhoeae that may outcompete the Presto assay in developing country settings. It is faster, but not as flexible, and it requires its own GeneXpert equipment, which is widely used in developing countries' laboratories for tuberculosis diagnostic testing. Still, the reagent for the GeneXpert CT/NG assay is more expensive compared to the Presto assay.

Limitations

An initial evaluation of the Presto assay was done in Belgium at the ITM. Ideally, this should have been performed in Rwanda, the country of use. However, at that time a PCR assay for the detection of C. trachomatis and N. gonorrhoeae was not applied at the NRL in Kigali, and well-characterised reference, clinical or quality control panels were not available. We acknowledge that the inability to initially evaluate the Presto assay in Rwanda is a limitation of the study.

Conclusions

In conclusion, testing for C. trachomatis and N. gonorrhoeae with the Presto assay was easily implemented and feasible in Kigali, Rwanda. Overall, a good performance of the assay was achieved, but our results suggest that positive Presto N. gonorrhoeae results may need confirmation using a nucleic acid amplification assay with a different amplification target. Further field evaluations are recommended to confirm our findings.

Competing interests

The authors declare that they have no financial or personal relationships that may have inappropriately influenced them in writing this article.

Sources of support

The main study was funded by the European & Developing Countries Clinical Trials Partnership through a project entitled 'Preparing for clinical trials with vaginal rings that protect women from HIV and unintended pregnancy' (grant code SP.2011.41304.043), the University of Liverpool and the Institute of Tropical Medicine.

Authors' contributions

V.C. and T.C. wrote the first draft of the manuscript. I.D.B., C.M.M., J.R. and J.v.d.W. revised and edited the text. H.S. designed and performed the evaluation at Institute of Tropical Medicine. V.C. provided the hands-on training at the NRL and supervised the elaboration of the standard operating procedures and quality control procedures. L.M. and V.M. generated the data. All authors revised and approved the present version of the manuscript.

Acknowledgements

We would like to thank all women who participated in the Ring Plus study and all the Ring Plus staff and laboratory technicians of National Reference Laboratory and Institute of Tropical Medicine for their dedicated collaboration.

References

  • 1.Unemo M, Bradshaw CS, Hocking JS, et al. Sexually transmitted infections: Challenges ahead. Lancet Infect Dis. 2017;17(8):e235-e279. https://doi.org/10.1016/S1473-3099(17)30310-9
    » https://doi.org/10.1016/S1473-3099(17)30310-9
  • 2.Tsevat DG, Wiesenfeld HC, Parks C, Peipert JF. Sexually transmitted diseases and infertility. Am J Obstet Gynecol. 2017;216(1):1-9. https://doi.org/10.1016/j.ajog.2016.08.008
    » https://doi.org/10.1016/j.ajog.2016.08.008
  • 3.Rivard KR, Dumkow LE, Draper HM, Brandt KL, Whalen DW, Egwuatu NE. Impact of rapid diagnostic testing for chlamydia and gonorrhea on appropriate antimicrobial utilization in the emergency department. Diagn Microbiol Infect Dis. 2017 Feb;87(2):175-179. https://doi.org/10.1016/j.diagmicrobio.2016.10.019
    » https://doi.org/10.1016/j.diagmicrobio.2016.10.019
  • 4.Chernesky MA, Jang D, Gilchrist J, et al. Comparison of cobas 4800, m2000, Viper XTR, and Infinity 80 Automated Instruments when processing urine specimens for the diagnosis of Chlamydia trachomatis and Neisseria gonorrhoeae. Sex Transm Dis. 2017 Jan;44(3):1. https://doi.org/10.1097/OLQ.0000000000000570
    » https://doi.org/10.1097/OLQ.0000000000000570
  • 5.Chernesky M, Jang D, Gilchrist J, et al. Head-to-head comparison of second-generation nucleic acid amplification tests for detection of Chlamydia trachomatis and Neisseria gonorrhoeae on urine samples from female subjects and self-collected vaginal swabs. J Clin Microbiol. 2014;52(7):2305-2310. https://doi.org/10.1128/JCM.03552-13
    » https://doi.org/10.1128/JCM.03552-13
  • 6.Schuurs TA, Verweij SP, Weel JFL, Ouburg S, Morré SA. Detection of Chlamydia trachomatis and Neisseria gonorrhoeae in an STI population: Performances of the Presto CT-NG assay, the Lightmix Kit 480 HT CT/NG and the COBAS Amplicor with urine specimens and urethral/cervicovaginal samples. BMJ Open. 2013;3(12):e003607. https://doi.org/10.1136/bmjopen-2013-003607
    » https://doi.org/10.1136/bmjopen-2013-003607
  • 7.de Waaij DJ, Dubbink JH, Peters RPH, Ouburg S, Morré SA. Comparison of GMT presto assay and Roche cobas® 4800 CT/NG assay for detection of Chlamydia trachomatis and Neisseria gonorrhoeae in dry swabs. J Microbiol Meth. 2015 Nov;118:70-74. https://doi.org/10.1016/j.mimet.2015.08.020
    » https://doi.org/10.1016/j.mimet.2015.08.020
  • 8.Unemo M, Clarke IN. The Swedish new variant of Chlamydia trachomatis. Curr Opin Infect Dis. 2011;24:62-69. https://doi.org/10.1097/QCO.0b013e32834204d5
    » https://doi.org/10.1097/QCO.0b013e32834204d5
  • 9.Schurmans C, De Baetselier I, Kestelyn E, et al. The ring plus project: Safety and acceptability of vaginal rings that protect women from unintended pregnancy. BMC Public Health. 2015;15(348). https://doi.org/10.1186/s12889-015-1680-y
    » https://doi.org/10.1186/s12889-015-1680-y
  • 10.Kestelyn E, Agaba S, Van Nuil JI, et al. A randomised trial of a contraceptive vaginal ring in women at risk of HIV infection in Rwanda: Safety of intermittent and continuous use. PLoS One. 2018 Jun 1;13(6):e0197572. https://doi.org/10.1371/journal.pone.0197572
    » https://doi.org/10.1371/journal.pone.0197572
  • 11.Chen C-Y, Chi KH, Alexander S, Ison CA, Ballard RC. A real-time quadriplex PCR assay for the diagnosis of rectal lymphogranuloma venereum and non-lymphogranuloma venereum Chlamydia trachomatis infections. Sex Transm Infect. 2008;84(4):273-276. https://doi.org/10.1136/sti.2007.029058
    » https://doi.org/10.1136/sti.2007.029058
  • 12.Hopkins MJ, Ashton LJ, Alloba F, Alawattegama A, Hart IJ. Validation of a laboratory-developed real-time PCR protocol for detection of Chlamydia trachomatis and Neisseria gonorrhoeae in urine. Sex Transm Infect. 2010;86(3):207-211. https://doi.org/10.1136/sti.2009.040634
    » https://doi.org/10.1136/sti.2009.040634
  • 13.Tabrizi SN, Unemo M, Limnios AE, et al. Evaluation of six commercial nucleic acid amplification tests for detection of Neisseria gonorrhoeae and other Neisseria species. J Clin Microbiol. 2011;49(10):3610-3615. https://doi.org/10.1128/JCM.01217-11
    » https://doi.org/10.1128/JCM.01217-11
  • 14.World Health Organization. WHO Guidelines for the Treatment ofNeisseria gonorrhoeae. Geneva: WHO; 2016.
  • 15.Bignell C, Unemo M, et al. 2012 European guideline on the diagnosis and treatment of gonorrhoea in adults. Int J Std Aids. 2013;24(2):85-92.
  • 16.World Health Organization. Guidelines for the management of sexually transmitted infections. Geneva: World Health Organization Library; 2003.
  • 17.Van Der Eem L, Dubbink JH, Struthers HE, et al. Evaluation of syndromic management guidelines for treatment of sexually transmitted infections in South African women. Trop Med Int Health. 2016 Sep;21(9):1138-1146. https://doi.org/10.1111/tmi.12742
    » https://doi.org/10.1111/tmi.12742
  • 18.Pettifor A, Walsh J, Wilkins V, Raghunathan P. How effective is syndromic management of STDs? A review of current studies. Sex Transm Dis. 2000;27(7):371-385. https://doi.org/10.1097/00007435-200008000-00002
    » https://doi.org/10.1097/00007435-200008000-00002
  • 19.Reyburn H. WHO Guidelines for the Treatment of Chlamydia trachomatis. Who. 2016;340(May 28 1):c2637-c2637.
  • 20.Sonkar SC, Wasnik K, Kumar A, Mittal P, Saluja D. Comparative analysis of syndromic and PCR-based diagnostic assay reveals misdiagnosis/ overtreatment for trichomoniasis based on subjective judgment in symptomatic patients. Infect Dis Poverty. 2016;5(1):42. https://doi.org/10.1186/s40249-016-0133-x
    » https://doi.org/10.1186/s40249-016-0133-x
  • 21.Gaydos C, Hardick J. Point of care diagnostics for sexually transmitted infections: Perspectives and advances. Expert Rev Anti Infect Ther. 2014;12(6):657-672. https://doi.org/10.1586/14787210.2014.880651
    » https://doi.org/10.1586/14787210.2014.880651
  • 22.Gaydos CA, Van Der Pol B, Jett-Goheen M, et al. Performance of the cepheid CT/NG Xpert rapid PCR test for detection of Chlamydia trachomatis and Neisseria gonorrhoeae. J Clin Microbiol. 2013;51:1666-1672. https://doi.org/10.1128/JCM.03461-12
    » https://doi.org/10.1128/JCM.03461-12

Correspondence:
Vicky Cuylaerts

Received: 19 Dec. 2017
Accepted: 05 Oct. 2018
Published: 18 Apr. 2019

Correspondence

vcuylaerts@itg.be

Publication Dates

  • Publication in this collection
    04 Feb 2020
  • Date of issue

History

  • Received
    19 Dec 2017
  • Accepted
    05 Oct 2018
location_on
African Society for Laboratory Medicine (Ethiopia) AOSIS Publishing, Postnet Suite #110, Private Bag X19, Durbanville, Western Cape, ZA, 7551, Tel: +27 21 975 2602, Tel: 086 1000 381 - E-mail: publishing@aosis.co.za
rss_feed Acompanhe os números deste periódico no seu leitor de RSS
Ir para o topo