Open-access Commercial DURAClone panels for extending the repertoire of multicolour immunophenotypic panels in an academic flow cytometry laboratory in South Africa

Abstract

BACKGROUND: Commercial multicolour fixed immunophenotyping panels can improve flow cytometric diagnostic immunophenotyping repertoire. OBJECTIVE: This study validated the commercially available, standardised Beckman Coulter lyophilised DURAClone RE panels to discriminate specific haematolymphoid subtypes. METHODS: We compared the diagnostic capability of the DURAClone acute leukaemia B (ALB), chronic leukaemia B (CLB), and plasma cells (PC) panels to the predicate second-line panels in Charlotte Maxeke Johannesburg Academic Hospital, Johannesburg, South Africa, from April to August 2020. Clinical diagnostic concordance between the in-house second-line immunophenotyping (the predicate method) and DURAClone was established. The ALB panels tested for precursor B-cell acute lymphoblastic leukaemia (n = 11) or normal bone marrow haematogones (n = 9); CLB panels established haematolymphoid subtypes of mature B-cell lymphoproliferative disorders (B-LPD) (n = 20), while PC panels detected plasma cell dyscrasias (PCD) (n = 17). Flow cytometer setup and data interpretation to discriminate normal and aberrant immunophenotypes were per manufacturer's instructions. RESULTS: There was 100% clinical diagnostic concordance between the predicate and the test panels for second-line diagnostic investigation of B-ALL (with additional CD56), mature B-LPD (with additional discernment of CD81, ROR-1, CD79b and CD43) and PCD CONCLUSION: The DURAClone CLB exceeded the predicate second-line performance, offering extended second-line diagnostic discernment of mature B-LPD subtypes and discernment of CD5+ B-LPD from other non-CD5+ (or CD5-) B-LPD; likewise, the PC panels enabled discovery of PCD. While ALB testing offered no additional diagnostic advantage over existing predicate investigation, CD58 did offer additional information to discern haematogones from B-ALL.

flow cytometry; multicolour; immunophenotyping; lymphoma; leukaemia; plasma cell dyscrasia; acute lymphoblastic leukaemia; B-cell lymphoproliferative disorder; lyophilised reagent; fixed panels


Introduction

Flow cytometry immunophenotypic analysis is a powerful diagnostic and research tool for investigating haemato-lymphoid malignancies, such as leukaemia, lymphoma, and plasma cell dyscrasias (PCD). Recently, multicolour flow cytometric immunophenotyping has become invaluable for improving detection and detailed immunophenotyping during diagnosis and detection of low frequencies of abnormal or aberrant cell populations during minimal residual disease assessment.1,2234,54,64,7

Commercially available, standardised, lyophilised, fixed and pre-titrated antibody panels,8,9,10,11,12,1314159914,15

Our laboratory previously established the positive impact on quality, laboratory workflow and diagnostic usefulness of the commercially available, fixed and standardised BC ClearLLab 10C169,161,217,1819

Existing predicate second-line investigation utilises locally established, in-house, non-standardised and manually assembled liquid fluorochrome-conjugated antibody panels. The primary aim of this study was to evaluate the fixed and standardised, pre-titrated commercially available DURAClone RE ALB, RE CLB and RE PC panels (Beckman Coulter, Mumbai, India) as a comprehensive and compact alternative to our existing second-line immunophenotypic investigation of haematological neoplasms. There were two parts to this study. First, we compared the DURAClone CLB and PC tubes to the laboratory's in-house predicate 2-4-colour method for the following markers: combinations of CD10, CD23, FMC7 and CD22 to discern mature B-cell lymphoproliferative disorders, or CD19, CD38, CD56, CD200 and CD138 to characterise plasma cells. Secondly, we verified manufacturer-described expression for those additional markers that were not included in the older predicate 2-4-colour method, such as verifying the expression of CD43, CD79b, CD81 and ROR-1 in the CLB tube and CD38 and CD138 with CD27, CD81 and CD56 included in the DURAClone PC tube against expected expression in normal leucocyte population counterparts. We also validated the overall diagnostic outcome of the ALB tube; here, we asked whether the CD58 included in the ALB tubes offered any additional diagnostic advantage in identifying a B-ALL that had been identified in the first-line ClearLLab 10C investigation.

Methods

Ethical considerations

The University of the Witwatersrand Health Research Ethics Committee approved the study (Ethics Clearance number M1704129). The study's objective was to validate the commercially available Beckman Coulter DURAClone PC, CLB and PC fixed-tube, pre-titrated panels as an alternative second-line diagnostic workup to our existing in-house second-line immunophenotypic investigation. The interpretation of flow cytometric data and consequent clinical diagnostic outcomes of the existing in-house, second-line, non-standardised panels were compared to the overall clinical diagnostic outcomes of the DURAClone analyses.

Study design and site

This study was a prospective observational cohort study conducted according to the Standards for reporting Diagnostics Accuracy.2021

Testing approach

Immunophenotypic testing of all samples at the Charlotte Maxeke Johannesburg Academic Hospital flow cytometry unit is two-tiered. Firstly, an initial first-line workup is performed with BC ClearLLab 10-C (Beckman Coulter, Miami, Florida, United States). The first workup identifies most acute myeloid and lymphoid leukaemia subtypes, distinguishes early and mature T-cell or B-cell lymphoproliferative disorders, and hints at the presence of a PCD (if a population of brightly expressing CD38 cells is noted). A limitation of this ClearLLab 10C system is that further immunophenotypic characterisation of mature B-cell lymphoproliferative disorders1719,2217,18

Samples

After all routine testing with ClearLLab 10C and in-house second-line testing (Table 1), samples having sufficient remnant prepared-cell-concentrate or at least 1 mL remaining of the whole sample were re-tested using one of the test methods (either DURAClone RE ALB, RE CLB or RE PC). The aim was to collect at least 20 clinical specimens for each validation (estimated to be 60 samples). All peripheral blood and bone marrow samples were collected by attending physicians into ethylenediaminetetraacetic acid vacutainer tubes; the single body fluid sample was not collected into an anticoagulant but submitted in a vacutainer tube without anticoagulant. At the end of the study period, a total of 57 anonymised samples were identified for second-line DURAClone comparison, including peripheral blood (n = 13), bone marrow (n = 43), and a pleural fluid sample (n = 1) collected from paediatric and adult patients. For DURAClone ALB testing, 20 bone marrow aspirate samples were tested. Eleven of these patients had a diagnosis of B-ALL (n = 11), while nine had immunophenotypically normal haematogones. One of the nine was a follow-up of a B-ALL with normal haematogones and no evidence of residual disease. Twenty samples diagnosed with a B-cell lymphoproliferative disorder by predicate ClearLLab 10C and in-house second-line were referred for DURAClone CLB testing. The CLB testing set comprised 12 bone marrow aspirate samples, seven peripheral blood samples, and a single pleural fluid sample. Due to the decommissioning of the older FACSCalibur flow cytometer during the period of study (the instrument used to undertake the predicate in-house second-line testing), only 17 samples with a diagnosis of a PCD by ClearLLab 10-C and in-house second-line investigation were tested by DURAClone PC panels.

Table 1

Instruments

The NaviosTM (BC, Miami, Florida, United States) and the FACSCalibur flow cytometer (Becton Dickinson Biosciences, San Jose, California, United States) were used during this study. The NaviosTM flow cytometer was used to analyse the predicate ClearLLab 10C as well as the test DURAClone panels. Internal quality control on the NaviosTM included daily assessment of background contamination, cellular events carryover between tubes, and fluorospheres acquisition to verify the flow cytometer optical alignment and fluidics (Flow-Check Pro, BC, Lismeehan, Ireland). In addition, ClearLLab™ Normal and Abnormal process control cells (BC, Lismeehan, Ireland) were used to verify sample processing, acquisition, and analysis. ClearLLab 10C panel acquisition setup was achieved by applying target values set in the manufacturer manual and using BC FlowSet Pro beads to adjust the voltages that enabled optimal detection and separation of dim and bright antigens. After pilot testing (data not shown), the NaviosTM instrument ClearLLab 10C settings were deemed appropriate for the DURAClone multicolour data acquisition and analysis.

The FACSCalibur flow cytometer was used to acquire the predicate in-house second-line 2-4-colour fluorescence panels using CellQuest software (BD, San Jose, California, United States). Quality control performed on the BD FACSCalibur included daily assessment of background contamination, carryover, acquisition, and analysis of manufacturer-recommended 3-colour and APC Calibrite beads (Becton Dickinson Biosciences, San Jose, California, United States). Further acquisition and analysis of ImmunotrolTM process control (BC, Brea, California, United States) using four monoclonal antibodies, CD14 FITC, CD13 PE, CD45 PERCP and CD3 APC (all Becton Dickinson Biosciences, San Jose, California, United States), was done.

Sample preparation

Sample preparation included lysing two or more 0.5 mL sample aliquots (dependent on the initial white cell count) with 14.5 mL isotonic ammonium chloride pH 7.1-7.4 (8.99% NH4Cl, 0.84% NaHCO3 and 0.0372% ethylenediaminetetraacetic acid; Merck, Darmstadt, Germany) in a 15 mL conical centrifuge tube for 15 min at room temperature. After incubation, samples were spun at 3000 g for 3 min, the supernatant was decanted, while the pellet was washed four times with 14 mL phosphate-buffered saline at pH 7.3 ± 2 (Oxoid Ltd, Basingstoke, United Kingdom) containing 0.09% sodium azide (NaN3) (Merck, Darmstadt, Germany) and 0.2% bovine serum albumin (Biowest, Nuaille, France). Following washing, samples were reconstituted to 0.5 mL with phosphate-buffered saline, and the white cell count was determined. Cells were then diluted (or concentrated) to achieve the recommended cellular concentration of ≤ 10 000 cells/µL for monoclonal antibody incubation. Subsequently, 100 µL of this cell concentrate, which contained approximately 106Table 1). If the sample was deemed to be suitable for inclusion in the study (i.e. had an established diagnosis by predicate ClearLLab 10C), and if there was sufficient remaining material, 100 µL of the remaining cell concentrate was tested with either DURAClone ALB, CLB or PC panels (Table 1). Details of the markers included in these DURAClone panels are outlined in Table 2. Each DURAClone panel has a 'spare' capacity for additional markers per the investigator's need. In our study, in two cases of precursor B-ALL, the marker CD22 (APC) (BC, Miami, Florida, United States) was added in liquid format to the ALB panel to occupy the free APC channel (see Table 2, DURAClone RE ALB panel) to verify CD22 expression in the context of a B-ALL. In a single case of a suspected PCD, liquid CD117 ECD (BC, Miami, Florida, United States) was added to the DURAClone RE PC tube in the 'free' ECD channel to demonstrate expression of CD117 in a case of multiple myeloma. After adding the cell aliquot into panels, and, where applicable, the addition of liquid monoclonal reagent, all samples were incubated at room temperature (average 20 °C - 22 °C) in the dark for 15 min. After incubation, all samples were washed once with phosphate-buffered saline and reconstituted to 0.5 mL.

Table 2

Data acquisition and interpretation of flow cytometric case data

All samples were acquired on a flow cytometer to acquire raw flow cytometric listmode data. For the predicate in-house panels acquired on the FACSCalibur, 5000 (.fcs data) events were collected. Paint-A-Gate software (Becton Dickinson Biosciences, San Jose, California, United States) was used for analysis of raw FACSCalibur flow cytometric data with a primary gating focus using CD45 and side scatter.

For both the ClearLLab 10C and DURAClone panels, at least 50 000 (listmode data) events were acquired on the Navios. Kaluza C™ version 1.1 (BC, Miami, Florida, United States) software was used to analyse all raw listmode data to facilitate clinical interpretation. Kaluza C™ ClearLLab B, T, M1 and M2, and DURAClone panel analysis protocols were developed according to manufacturer specifications.239,16 Finally, simultaneous identification of normal populations in the background was based on local in-house developed gating strategies to identify granulocytes, monocytes and mature lymphocytes using CD45, side scatter characteristics and specific regular expression of the markers studied (Figure 1 and Figure 2).

Figure 1

Figure 2

A similar approach was used for each of the comparative second-line DURAClone RE ALB, RE CLB, and RE PC listmode data analysis but additionally incorporated manufacturer-recommended gating strategies for clinical interpretation of data.2324

Statistical analysis

Clinical diagnostic outcomes were collated into Microsoft Excel (Redmond, Washington, United States) spreadsheets. Marker expressions and specific clinical diagnoses noted for the DURAClone RE ALB, RE CLB, and RE PC panels were compared to the clinical diagnostic outcomes from the existing in-house 2-4-colour antibody panels described (Table 2). The agreement between methods (comparing the final diagnostic outcome) was assessed using contingency tables to determine sensitivity and specificity. True positives were regarded as clinical diagnostic outcomes matching the predicate ClearLLab 10-C with in-house panels, and test system outcomes using ALB, CLB or PC. For example, when a B-cell CLL was diagnosed on the initial ClearLLab investigation with CD23 expression by predicate second-line testing, a diagnosis of a B-cell CLL was confirmed in the CLB tube with positive ROR-1 and CD43 expression. True negatives were those cases where there was no disease noted, either by predicate ClearLLab and second-line testing or by test method with DURAClone panel testing. False positives were defined as those cases where there was no disease noted on predicate ClearLLab with in-house second-line testing, but the disease was noted by respective DURAClone analysis. Likewise, false negatives were defined as disease on predicate first-line ClearLLab and second-line testing, but no disease by respective DURAClone panel testing.

Results

Normal and abnormal B progenitor cells panel

Twenty samples were tested with the DURAClone ALB panel. The DURAClone RE ALB evaluation on both routine diagnostic precursor B-ALL samples, and patient samples who were being followed up after therapy for precursor B-ALL, revealed 100% positive and negative agreement to ClearLLab 10C reported outcomes (Table 3). The estimated sensitivity and specificity rates were both 100%. CD58 expression (an additional marker included in the DURAClone ALB tube, but not ClearLLab tubes) enabled further confirmation of the diagnostic outcome reported; here, there was consistent under-expression of CD58 noted on haematogones and consistent over-expression seen on abnormal precursor B-cells (blasts), verifying manufacturer-described CD58 expression. CD22 APC expression (that was added as an additional marker panel in two precursor B-ALL cases to demonstrate that the addition was possible) was verified against the expected disease outcome (positive in two cases of B-ALL) and met the manufacturer's specifications in the DURAClone RE ALB (Table 3).

Chronic B-cell lymphoproliferative disorder panel

The DURAClone RE CLB panel identified the B-cell target population using a combination of CD19 and CD5 or CD20, as well as surface CD81, ROR1, CD79b, and CD43 expression. Twenty samples were tested with the DURAClone CLB panel. The calculated sensitivity and specificity were 100%, with full diagnostic concordance noted across all cases evaluated (n = 20). All CLL cases (n = 16/20) showed under-expression (weak) of CD81, CD79b and CD20 (Table 4). Further, surface ROR1 expression was detected in 15 of the 16 (93.75%) CD5-expressing CLL cases; all CLL showed expression of CD43. A single case of follicular lymphoma (n = 1) showed a bright expression of CD79b. Three cases (15%) evaluated could not be definitively sub-classified with the in-house second-line panel analysis or the DURAClone RE CLB reagents. An additional control sample (not included in Table 2), with confirmed 20% of B-cell precursors, revealed and confirmed bright CD81 on the normal precursor B-cells (Figure 1).

Normal and abnormal plasma cells panel

Seventeen samples were tested with the DURAClone PC panel. A plasma cell population was identified by backbone markers CD138 and CD38 in the DURAClone RE PC panel, with subsequent determination of expression of CD19, CD27, CD45, CD56 and CD200. There was cross-panel marker equivalency to the existing in-house panels (Table 5) and clinical outcomes, with 100% agreement achieved, and estimated sensitivity and specificity rates of 100%. In addition, the assessment of CD27 and CD81 on target plasma cells revealed consistent under-expression of CD81 or CD27 (n = 14/14) (Figure 2). A single (n = 1) PCD had positive surface CD117 expression (Figure 2) on the aberrant identified plasma cell population, in agreement with CD117 expression noted in the matching ClearLLab 10C M2 analysis.

Discussion

The recently evaluated,9,161125,26,2716

Complementary and supplementary multicolour panel options are published3,4,10,19,281692422,29,30

Normal and abnormal B progenitor cells panel

The expression of CD10, CD19, CD34, CD38, CD20 and CD45 in the DURAClone RE ALB panel showed excellent cross-panel marker expression equivalency and concordant clinical diagnostic outcome when compared to ClearLLab 10C B-cell panel expression. CD58 together with CD38, CD10, CD19, CD34, CD20 and CD45 has potential for sensitive minimal residual disease assessment31,32313132Table 2).

Chronic B-cell lymphoproliferative disorder panel

The expression of CD5, CD19, CD20 and CD45 in the DURAClone RE CLB reagent showed excellent cross-panel marker expression, equivalent to ClearLLab 10C B panel expression, with 100% (n = 20/20) and clinical diagnostic outcomes agreement. The additional markers in the DURAClone RE CLB panel, including CD81, ROR1, CD79b and CD43 which are not included in the laboratory predicate ClearLLab system, specifically the B-cell tube, met the requirements for intended use and the manufacturer performance specifications. The DURAClone RE CLB panel proved to be an excellent fixed supplementary panel to replace our outdated second-line investigation for the workup of mature B-cell lymphoid proliferations and proved to be especially useful for discerning CD19/CD5 positive haematolymphoid disease subtypes. The surface CD81, CD79b and CD20 showed consistent under-expression in CLL, while the presence of ROR1 was also helpful in discerning CLL over mantle cell lymphoma as described in previous studies,17,18,3334,3517,18,33

Normal and abnormal plasma cells panel

The CD38, CD138, CD45, CD19, CD56, and CD200 expressions in the DURAClone RE PC panel showed excellent cross-panel marker expression and equivalent diagnostic outcome and FACSCalibur/PAG analysis outcomes, with 100% (n = 17/17) clinical diagnostic concordance (agreement). In addition, as previously described,29303014,30.

Limitations

Firstly, the sample size per panel evaluated is small, and further studies are needed to confirm the outcomes reported here. Secondly, the fluorochromes used in the panels are specifically designed for use on a BC Navios instrument. Therefore, the products could be used on alternative instruments only if the respective filter setups23

Conclusion

This study confirms that the BC DURAClone ALB and CLB panels are suitable to provide additional second-line immunophenotypic workup of precursor B-ALL and mature B-cell lymphoproliferative disorders resepectively, at disease presentation and are suitable to supplement first-line ClearLLab 10C laboratory predicate method testing. The under-expression of CD81, CD79b, and positive CD43 and ROR1 expression, noted in the DURAClone CLB panel specifically assists in distinguishing B-cell CLL from mantle cell lymphoma and other CD5 negative B-cell LPDs.

The DURAClone RE PC panel is suitable for the second-line investigation of PCD. CD27, CD56, CD81 and CD200, together with CD19, CD38 and CD138 in a single analysis, were efficient in identifying aberrant plasma cells.

Although there was diagnostic and individual marker concordance, we did not find the ALB more useful for diagnosing B-ALL over our existing ClearLLab system (utilising the B-cell and M2 tubes). The inclusion of CD5831

Lastly, the potential benefit of using commercialised multicolour lyophilised fixed panel preparations having stable, standardised antibody reagents reduces the risks of technical error, improves laboratory efficiency, and simplifies reagent inventory. In addition, the DURAClone 'free fluorescence channels' provide some additional flexibility for a laboratory to use their preferred markers to establish diseases (or not) of their choice.

Competing interests

The authors declare that they have no financial or personal relationships that may have inappropriately influenced them in writing this article.

Authors' contributions

D.K.G. conceptualised, led and funded the study, provided project administration and project supervision as well as supervision of all laboratory testing and data analysis. D.L. and D.K.G. oversaw technical aspects and the setup of protocols for flow cytometric data acquisition and data analysis. L.S. and M.P. analysed flow cytometric data and recorded the outcomes in spreadsheets for analysis comparison. D.K.G. undertook final checking of data and compiled the final laboratory validation report. L.S. and D.K.G. translated the validation report into a first draft for publication format. All subsequent drafts were written by D.K.G.

Sources of support

Funding for this study was made possible with research incentive funds accrued to D.K.G. through the University of the Witwatersrand.

Data availability

All outcomes are anonymised and published in the maunscript. Individal raw flow cytometric listmode data files are available on request from the corresponding author, D.K.G.

Disclaimer

The views and opinions expressed in this article are those of the authors and do not necessarily reflect the official policy or position of any of the affiliated agencies of the authors.

Acknowledgements

D.K.G. and D.L. thank Mrs Merriam Machaba at Charlotte Maxeke Johannesburg Academic Hospital flow cytometry.

References

  • 1.Borowitz MJ, Wood BL, Devidas M, et al. Prognostic significance of minimal residual disease in high risk B-ALL: A report from Children's Oncology Group study AALL0232. Blood. 2015;126(8):964-971. https://doi.org/10.1182/blood-2015-03-633685
    » https://doi.org/10.1182/blood-2015-03-633685
  • 2.Theunissen P, Mejstrikova E, Sedek L, et al. Standardized flow cytometry for highly sensitive MRD measurements in B-cell acute lymphoblastic leukemia. Blood. 2017;129(3):347-357. https://doi.org/10.1182/blood-2016-07-726307
    » https://doi.org/10.1182/blood-2016-07-726307
  • 3.Van Dongen JJ, Lhermitte L, Bottcher S, et al. EuroFlow antibody panels for standardized n-dimensional flow cytometric immunophenotyping of normal, reactive and malignant leukocytes. Leukemia. 2012;26(9):1908-1975. https://doi.org/10.1038/leu.2012.120
    » https://doi.org/10.1038/leu.2012.120
  • 4.Kalina T, Flores-Montero J, Van der Velden VH, et al. EuroFlow standardization of flow cytometer instrument settings and immunophenotyping protocols. Leukemia. 2012;26(9):1986-2010. https://doi.org/10.1038/leu.2012.122
    » https://doi.org/10.1038/leu.2012.122
  • 5.Glier H, Heijnen I, Hauwel M, et al. Standardization of 8-color flow cytometry across different flow cytometer instruments: A feasibility study in clinical laboratories in Switzerland. J Immunol Methods. 2019;475:112348. https://doi.org/10.1016/j.jim.2017.07.013
    » https://doi.org/10.1016/j.jim.2017.07.013
  • 6.Kalina T, Brdickova N, Glier H, et al. Frequent issues and lessons learned from EuroFlow QA. J Immunol Methods. 2019;475:112520. https://doi.org/10.1016/j.jim.2018.09.008
    » https://doi.org/10.1016/j.jim.2018.09.008
  • 7.Davis BH, Holden JT, Bene MC, et al. 2006 Bethesda International Consensus recommendations on the flow cytometric immunophenotypic analysis of hematolymphoid neoplasia: Medical indications. Cytometry B Clin Cytom. 2007;72 Suppl 1:S5-S13. https://doi.org/10.1002/cyto.b.20365
    » https://doi.org/10.1002/cyto.b.20365
  • 8.Hedley BD, Cheng G, Luider J, et al. Initial flow cytometric evaluation of the Clearllab lymphoid screen. Cytometry B Clin Cytom. 2018;94(5):707-713. https://doi.org/10.1002/cyto.b.21603
    » https://doi.org/10.1002/cyto.b.21603
  • 9.Hedley BD, Cheng G, Keeney M, et al. A multicenter study evaluation of the ClearLLab 10C panels. Cytometry B Clin Cytom. 2020.
  • 10.Flores-Montero J, Grigore G, Fluxa R, et al. EuroFlow Lymphoid Screening Tube (LST) data base for automated identification of blood lymphocyte subsets. J Immunol Methods. 2019;475:112662. https://doi.org/10.1016/j.jim.2019.112662
    » https://doi.org/10.1016/j.jim.2019.112662
  • 11.Hedley BD, Keeney M, Popma J, Chin-Yee I. Novel lymphocyte screening tube using dried monoclonal antibody reagents. Cytometry B Clin Cytom. 2015;88(6):361-370. https://doi.org/10.1002/cyto.b.21251
    » https://doi.org/10.1002/cyto.b.21251
  • 12.Van der Velden VHJ, Flores-Montero J, Perez-Andres M, et al. Optimization and testing of dried antibody tube: The EuroFlow LST and PIDOT tubes as examples. J Immunol Methods. 2019;475:112287. https://doi.org/10.1016/j.jim.2017.03.011
    » https://doi.org/10.1016/j.jim.2017.03.011
  • 13.Correia RP, Rajab A, Bento LC, et al. A ten-color tube with dried antibody reagents for the screening of hematological malignancies. Int J Lab Hematol. 2018;40(2):136-143. https://doi.org/10.1111/ijlh.12753
    » https://doi.org/10.1111/ijlh.12753
  • 14.Bouriche L, Bernot D, Nivaggioni V, Arnoux I, Loosveld M. Detection of minimal residual disease in B cell acute lymphoblastic leukemia using an eight-color tube with dried antibody reagents. Cytometry B Clin Cytom. 2019;96(2):158-163. https://doi.org/10.1002/cyto.b.21766
    » https://doi.org/10.1002/cyto.b.21766
  • 15.Bento L, Correia R, De Sousa F, et al. Performance of eight-color dry antibody reagent in the detection of minimal residual disease in chronic lymphocytic leukemia samples. Cytometry B Clin Cytom. 2020;98(6):529-535. https://doi.org/10.1002/cyto.b.21875
    » https://doi.org/10.1002/cyto.b.21875
  • 16.Glencross DK, Pretorius M, Swart L, Lawrie D. Evaluation of the Beckman Coulter (BC) ClearLLab 10C Leukaemia/Lymphoma Immunophenotyping Panel system in a busy, high-volume academic laboratory in Johannesburg. Afr J Lab Med. 2022;11(1):a1458. https://doi.org/10.4102/ajlm.v11i1.1458
    » https://doi.org/10.4102/ajlm.v11i1.1458
  • 17.Rawstron AC, Fazi C, Agathangelidis A, et al. A complementary role of multiparameter flow cytometry and high-throughput sequencing for minimal residual disease detection in chronic lymphocytic leukemia: An European Research Initiative on CLL study. Leukemia. 2016;30(4):929-936. https://doi.org/10.1038/leu.2015.313
    » https://doi.org/10.1038/leu.2015.313
  • 18.Rawstron AC, Kreuzer KA, Soosapilla A, et al. Reproducible diagnosis of chronic lymphocytic leukemia by flow cytometry: An European Research Initiative on CLL (ERIC) & European Society for Clinical Cell Analysis (ESCCA) Harmonisation project. Cytometry B Clin Cytom. 2018;94(1):121-128. https://doi.org/10.1002/cyto.b.21595
    » https://doi.org/10.1002/cyto.b.21595
  • 19.Roshal M, Flores-Montero JA, Gao Q, et al. MRD detection in multiple myeloma: Comparison between MSKCC 10-color single-tube and EuroFlow 8-color 2-tube methods. Blood Adv. 2017;1(12):728-732. https://doi.org/10.1182/bloodadvances.2016003715
    » https://doi.org/10.1182/bloodadvances.2016003715
  • 20.Bossuyt PM, Reitsma JB, Bruns DE, et al. Towards complete and accurate reporting of studies of diagnostic accuracy: The STARD initiative. Clin Chem Lab Med. 2003;41(1):68-73. https://doi.org/10.1515/CCLM.2003.012
    » https://doi.org/10.1515/CCLM.2003.012
  • 21.United Kingdom National External Quality Assessment Scheme (U.K. NEQAS) U. Leukaemia immunophenotyping part 1 and part 2 [homepage on the Internet]. [Cited 2021 Aug]. Available from: http://www.ukneqasli.co.uk/eqa-pt-programmes/flow-cytometry-programmes/leukaemia-immunophenotyping-part-1-accredited/
    » http://www.ukneqasli.co.uk/eqa-pt-programmes/flow-cytometry-programmes/leukaemia-immunophenotyping-part-1-accredited/
  • 22.Wang HW, Lin P. Flow cytometric immunophenotypic analysis in the diagnosis and prognostication of plasma cell neoplasms. Cytometry B Clin Cytom. 2019;96(5):338-350. https://doi.org/10.1002/cyto.b.21844
    » https://doi.org/10.1002/cyto.b.21844
  • 23.Beckman Coulter SA. DURAClone panels 2021 [homepage on the Internet]. [cited 2021 Aug]. Available from: https://www.beckman.co.za/reagents/coulter-flow-cytometry/antibodies-and-kits/duraclone-panels
    » https://www.beckman.co.za/reagents/coulter-flow-cytometry/antibodies-and-kits/duraclone-panels
  • 24.Campo E, Harris NL, Jaffe ES, et al. World Health Organization (WHO) classification of tumours of haematopoietic and lymphoid tissues (medicine). Geneva: WHO; 2017.

  • » https://doi.org/10.1002/cyto.b.20384
  • 25.Glencross DK, Coetzee LM, Cassim N. An integrated tiered service delivery model (ITSDM) based on local CD4 testing demands can improve turn-around times and save costs whilst ensuring accessible and scalable CD4 services across a national programme. PLoS One. 2014;9(12):e114727. https://doi.org/10.1371/journal.pone.0114727
    » https://doi.org/10.1371/journal.pone.0114727
  • 27.Coetzee LM, Glencross DK. Performance verification of the new fully automated Aquios flow cytometer PanLeucogate (PLG) platform for CD4-T-lymphocyte enumeration in South Africa. PLoS One. 2017;12(11):e0187456. https://doi.org/10.1371/journal.pone.0187456
    » https://doi.org/10.1371/journal.pone.0187456
  • 28.Novakova M, Glier H, Brdickova N, et al. How to make usage of the standardized EuroFlow 8-color protocols possible for instruments of different manufacturers. J Immunol Methods. 2019;475:112388. https://doi.org/10.1016/j.jim.2017.11.007
    » https://doi.org/10.1016/j.jim.2017.11.007
  • 29.Flores-Montero J, De Tute R, Paiva B, et al. Immunophenotype of normal vs. myeloma plasma cells: Toward antibody panel specifications for MRD detection in multiple myeloma. Cytometry B Clin Cytom. 2016;90(1):61-72. https://doi.org/10.1002/cyto.b.21265
    » https://doi.org/10.1002/cyto.b.21265
  • 30.Chen F, Hu Y, Wang X, Fu S, Liu Z, Zhang J. Expression of CD81 and CD117 in plasma cell myeloma and the relationship to prognosis. Cancer Med. 2018;7(12):5920-5927. https://doi.org/10.1002/cam4.1840
    » https://doi.org/10.1002/cam4.1840
  • 31.Tsitsikov E, Harris MH, Silverman LB, Sallan SE, Weinberg OK. Role of CD81 and CD58 in minimal residual disease detection in pediatric B lymphoblastic leukemia. Int J Lab Hematol. 2018;40(3):343-351. https://doi.org/10.1111/ijlh.12795
    » https://doi.org/10.1111/ijlh.12795
  • 32.Zeidan MA, Kamal HM, El Shabrawy DA, Esh AM, Sattar RH. Significance of CD34/CD123 expression in detection of minimal residual disease in B-ACUTE lymphoblastic leukemia in children. Blood Cells Mol Dis. 2016;59:113-118. https://doi.org/10.1016/j.bcmd.2016.05.005
    » https://doi.org/10.1016/j.bcmd.2016.05.005
  • 33.Broome HE, Rassenti LZ, Wang HY, Meyer LM, Kipps TJ. ROR1 is expressed on hematogones (non-neoplastic human B-lymphocyte precursors) and a minority of precursor-B acute lymphoblastic leukemia. Leuk Res. 2011;35(10):1390-1394. https://doi.org/10.1016/j.leukres.2011.06.021
    » https://doi.org/10.1016/j.leukres.2011.06.021
  • 34.Kohnke T, Wittmann VK, Bucklein VL, et al. Diagnosis of CLL revisited: Increased specificity by a modified five-marker scoring system including CD200. Br J Haematol. 2017;179(3):480-487. https://doi.org/10.1111/bjh.14901
    » https://doi.org/10.1111/bjh.14901
  • 35.Mora A, Bosch R, Cuellar C, et al. CD200 is a useful marker in the diagnosis of chronic lymphocytic leukemia. Cytometry B Clin Cytom. 2019;96(2):143-148. https://doi.org/10.1002/cyto.b.21722
    » https://doi.org/10.1002/cyto.b.21722

Correspondence:
Deborah Glencross

Received: 31 Aug. 2021
Accepted: 28 July 2022
Published: 29 Nov. 2022

Publication Dates

  • Publication in this collection
    17 Jan 2023
  • Date of issue

History

  • Received
    31 Aug 2021
  • Accepted
    28 July 2022
location_on
African Society for Laboratory Medicine (Ethiopia) AOSIS Publishing, Postnet Suite #110, Private Bag X19, Durbanville, Western Cape, ZA, 7551, Tel: +27 21 975 2602, Tel: 086 1000 381 - E-mail: publishing@aosis.co.za
rss_feed Acompañe los números de esta revista en su lector de RSS
Ir para arriba