vol.14 número1Herpes simplex virus-2 infections in pregnant women from South Africa: Evaluation of the ImmunoFLOW rapid testUsing Systematized Nomenclature of Medicine clinical term codes to assign histological findings for prostate biopsies in the Gauteng province, South Africa: Lessons learnt índice de autoresíndice de materiabúsqueda de artículos
Home Pagelista alfabética de revistas  

Servicios Personalizados

Revista

Articulo

Indicadores

    Links relacionados

    • En proceso de indezaciónCitado por Google
    • En proceso de indezaciónSimilares en Google

    Compartir


    African Journal of Laboratory Medicine

    versión On-line ISSN 2225-2010versión impresa ISSN 2225-2002

    Resumen

    MASEMOLA, Kagiso M.; KONE, Ngalulawa; CHIKOMBA, Chemedzai  y  KHOZA, Siyabonga. Validation of Nova Stat Profile Prime Plus point-of-care testing for dialysis in South Africa. Afr. J. Lab. Med. [online]. 2025, vol.14, n.1, pp.1-9. ISSN 2225-2010.  https://doi.org/10.4102/ajlm.v14i1.2663.

    BACKGROUND: Chronic kidney disease is a global health crisis, and delays in laboratory testing worsen outcomes. Point-of-care testing (POCT) has shown utility in various settings, but its performance at high creatinine levels seen in advanced chronic kidney disease is variable OBJECTIVE: We evaluated the analytical and clinical performance of the Nova Stat Profile Prime Plus (SPP+) point-of-care analyser among patients on maintenance haemodialysis METHODS: A prospective study was conducted at Chris Hani Baragwanath Hospital, Johannesburg, South Africa. In phase one (01 June 2023 - 31 July 2023), precision, linearity, and accuracy of SPP+ were assessed using remnant patient samples. Blood gases, electrolytes, and metabolic parameters were compared with the GEM Premier 5000, while urea and creatinine were compared with the Roche cobas c702. In phase two (01 November 2023 - 31 December 2023), SPP+ was clinically validated among adults undergoing haemodialysis. Whole blood was collected pre- and post-dialysis to assess dialysis adequacy and the creatinine index RESULTS: In phase one, SPP+ showed acceptable precision (coefficients of variation 0% - 2.7%), linearity, and correlation (r > 0.90). Creatinine showed proportional bias (4.56% [0.33 - 8.80]) at higher concentrations. Among 51 haemodialysis patients (22 women, 29 men; aged 32-51 years), SPP+ showed 88.6% agreement for single pool Kt/V and urea reduction ratio. However, creatinine index agreement was low (34.3%, Cohen's kappa = 0.24, p < 0.0001 CONCLUSION: Nova SPP+ was comparable to central laboratory analysis, though caution is needed at high creatinine levels, where central laboratory analysis remains the preferred choice WHAT THIS STUDY ADDS: This study provides the first South African data on POCT in haemodialysis. The analyser has demonstrated potential for monitoring, but performance concerns remain at high creatinine levels. The findings offer practical guidance for integrating POCT into advanced chronic kidney disease care in a resource-limited setting.

    Palabras clave : creatinine index; critical care analysers; haemodialysis adequacy; Nova Stat Profile Prime Plus; point-of-care testing.