Open-access Radiology blues: Comparing occupational blue-light exposure to recommended safety standards

Abstract

BACKGROUND: The blue-light hazard is a well-documented entity addressing the detrimental health effects of high-energy visible light photons in the range of 305 nm - 450 nm. Radiologists spend long hours in front of multiple light-emitting diode (LED)-based diagnostic monitors emitting blue light, predisposing them to potentially higher blue-light dosages than other health professionals. OBJECTIVES: The authors aimed to quantify the blue light that radiology registrars are exposed to in daily viewing of diagnostic monitors and compared this with international occupational safety standards. METHOD: A limited cross-sectional observational study was conducted. Four radiology registrars at two academic hospitals in Bloemfontein from 01 October 2021 to 30 November 2021 participated. Diagnostic monitor viewing times on a standard workday were determined. Different image modalities obtained from 01 June 2019 to 30 November 2019 were assessed, and blue-light radiance was determined using a spectroscope and image analysis software. Blue-light radiance values were compared with international safety standards RESULTS: Radiology registrars spent on average 380 min in front of a diagnostic display unit daily. Blue-light radiance from diagnostic monitors was elevated in higher-intensity images such as chest radiographs and lower for darker images like MRI brain studies. The total blue-light radiance from diagnostic display units was more than 10 000 times below the recommended threshold value for blue-light exposure. CONCLUSION: Blue-light radiance from diagnostic displays measured well below the recommended values for occupational safety. Hence, blue-light exposure from diagnostic monitors does not significantly add to the occupational health burden of radiologists. CONTRIBUTION: Despite spending long hours in front of diagnostic monitors, radiologists' exposure to effective blue-light radiance from monitors was far below hazardous values. This suggests that blue-light exposure from diagnostic monitors does not increase the occupational health burden of radiologists.

blue-light hazard; ocular health radiology; blue-light radiance; occupational blue-light hazard; blue-light exposure; safe occupational exposure


Introduction

Blue light and its relation to health swiftly became one of the buzzwords in a myriad of discussions on physical condition and well-being.12,3,4,5

As a result of the fundamental nature of their work, radiologists spend long hours scrutinising images on visual display units (VDUs) - often up to five brightly illuminated monitors at a time. In addition to this, smartphones, tablets, personal computers and televisions for purposes of communication, referencing, administrative tasks at work, studying and leisure also add a significant component to the total daily screen time.6

An important contributing factor to the unfavourable effect of prolonged screen time is greater exposure to the short wavelength photons at the ultraviolet end of the visible light spectrum or blue light.678910

Chronic eye strain and the subsequent symptoms that follow and affect ocular health and vision, as well as mental health illness and burnout, especially during residency or registrar periods, were identified as occupational hazards radiologists face that are unrelated to radiation.1112,13

The concept of potential phototoxic effects to the retina related to blue-light exposure is referred to as the blue-light hazard. Blue light imparts retinal phototoxicity via energy deposition with direct subsequent damage to photoreceptor cells.1415

In this study, the blue-light exposure from VDUs of radiology registrars working in a training institution was quantified and compared with recommended international safety standards. By analysing this parameter, insight was gained into one of the main occupational health challenges in radiology.

Quantifying blue light

The American Conference of Governmental Industrial Hygienists (ACGIH®) have developed a series of Threshold Limit Values (TLVs®) and Biological Exposure Indices (BEIs®) to serve as guidelines for safe occupational exposure to chemical substances and physical agents, including non-ionising radiation such as blue light.1617

The radiant luminance of blue light, LB, from the source in question is used to assess the amount of light incident on the retina. Radiant luminance is a radiometric concept deduced from radiant flux, flux density and intensity. Radiant flux, Φe, is an indication of the amount of energy transmitted for a specific time period and is measured as energy units transferred per unit of time, or Watts (J/s). Flux density, €, is obtained when radiant flux is measured in a specific area unit. In the case of ocular photometry, it is the radiant flux at the retina (W/cm2

Monitors of VDUs consist of multiple arrays of light-emitting diode (LED) light sources emitting light in a Lambertian distribution. When looking at a monitor of a VDU, the sum of all the small sources combined is emitted in a diffuse manner, as illustrated. Light emitted from VDUs is considered to be originating from an extended source. Emission occurs from the surface in all directions, hence the solid angle subtending the source must be considered to be accurate and include all light emissions (Figure 1).18

Figure 1

The effective blue-light radiance, LB, is the unit of significance when evaluating radiance dose to the retina. It is attained by summating or integrating the product of the radiance of the source for every wavelength and the blue-light hazard weighting function B(λ):16,17

The blue-light hazard weighting function represents the relative sensitivity of the human eye and the potential of specific wavelengths in the blue-light range to induce photochemical injury. In the case of an aphakic eye, a hazard function, A(λ), with heavier weighting is used to compensate for the absence of the lens.17

When assessing typical workday exposures, viewing times will exceed 167 min (10 000 s), and the exposure limit of the blue light-weighted effective radiance LB is defined by the ICNIRP as:13

Methodology

A cross-sectional observational study was conducted to establish blue-light exposure. Participants consisted of four radiology registrars working at two academic hospitals in central South Africa from 01 October 2021 to 30 November 2021. Measurements were performed for a total duration of 20 working days.

Only the registrars working at CT and MR imaging stations were included. Reporting of daily scheduled CT and MR scans involves the prolonged evaluation of images on a minimum of three brightly lit computer monitors, which include two dedicated diagnostic display (DDD) units and one standard monitor. Other workstations such as ultrasound, mammography, fluoroscopy and interventional radiology are less predictable than CT and MR, as they involve more active and procedural patient interaction. These stations automatically yield less screen time and subsequent blue-light exposure and were excluded. During conduction of this study, the department did not have a registrar allocated to full-time plain film reporting; hence, it was not possible to include a participant reporting only chest radiography (CXR).

This study encompassed two pathways that were ultimately combined to obtain effective blue-light radiation (Figure 2). Current photometric systems used to evaluate blue light have restricted practical capabilities outside laboratory settings because of factors such as weight, bulky size, expensive components and difficulty in providing reproducible accuracy.19Figure 2).

Registrar viewing times

The TLVs® as determined by the ACGIH® are provided as maximum values for different viewing time ranges (Table 1).16® to be applied.

Participants logged the time they spent viewing images on DDD units by means of a stopwatch application. They activated the stopwatch whenever viewing was started and stopped it when they left the workstation or stopped viewing images on the DDD unit. Viewing times were recorded over the duration of 20 workdays to obtain an estimate of the total viewing time registrars spend in front of the DDD units per day.

Instrumentation

The authors assessed a 3-MP diagnostic reporting monitor that was calibrated as per requirements for licence holders by the Department of Health in South Africa.2021

The AvaSpec-ULS2048CL Spectrometer, with optimal efficiency for non-ionising radiation in the visible range, was used for measurements of blue-light emission from DDD units. This spectrometer uses a fibre optic system to measure light emission from a source by counting the number of photons for each wavelength in a specific range. The fibre optic detector was mounted in a fixed plastic envelope to ensure reproducibility of all measurements in terms of area of measurement, influence of ambient light and distance of detector from monitor. The measured data were delivered in spectral form for different wavelengths of light, displayed as the number of photons of each wavelength on a graph (Figure 3). The spectrometer was calibrated to a standard source of white light using the principle of blackbody radiation and an incandescent tungsten lamp, emitting white light at a specific known temperature.22

Figure 3

Establishing average intensities of different imaging modalities

A radiological image displayed on a DDD unit comprises a matrix of different pixels. Each pixel is allocated a number that gives it an 'address' in the matrix and displays a certain value of the greyscale to make up the final image. The greyscale is a range of different intensities varying from black to white. Black is defined as no intensity and white is the highest intensity. A value of '0' is assigned to black, and '255' to white. A computer interprets an image as a range of different values allocated to each pixel in the matrix. Each pixel displays a specific value of the greyscale to form the image (Figure 4). These pixel values that make up an image can be displayed in a greyscale histogram, which is a graph that displays the number of pixels as a function for each intensity value of the greyscale.22

Figure 4

An example of the histogram for the chest radiograph in Figure 4 is demonstrated in Figure 5. On the x-axis, greyscale values from 0 to 255 are presented, with the y-axis representing the number of pixels displaying each greyscale value. Also demonstrated in Figure 5 is a histogram of a single slice of an MR image. Intuitively, the MR image demonstrates larger amounts of darker image elements, explaining the relative shift of the histogram towards the left-sided, darker end of the greyscale. This observation is based on the standard presentation presets of the PACS system in the department under study and without adjusting any window settings. The bright chest radiograph, with greyscale values in a wide range, demonstrates a broad histogram with large amounts of different intensity values. If the intensity values of all the pixels in the histogram are summated, an average number, Iavg, indicating the intensity of the image as a whole, can be determined.

Viewed content will play an important part in the dosage of blue light, as images with increased intensity will yield more bright light from the screen and, as a result, increased associated blue-range photons. For the purpose of this study, standard CXR, 5 mm axial slices of pre-contrasted CT images of the brain and 4 mm axial T1 and T2 pre-contrasted MR images of the brain were utilised. Commonly used modalities were selected to provide an indication of how looking at different imaging modalities may influence blue-light exposure.

The sample of assessed images was randomly selected and included 100 PA view chest radiographs, 30 pre-contrasted CT brain studies consisting of 5 mm slices and 25 pre-contrasted MRI brain studies, T1 and T2 sequences, 4 mm slices. Images were selected from all the studies performed at Universitas Academic Hospital from June 2019 to November 2019. The authors purposely chose a 6-month period that fell outside the South African national lockdown (announced at the end of March 2020) as the lockdown regulations and practices significantly reduced patient numbers in the months following their implementation. Arbitrary holiday periods of December and April were excluded to give a realistic representation of a standard workday. A list of total studies for the respective modalities during said time periods was obtained and entered into Microsoft Excel. Every study was allocated a number, and from this, random samples were selected.

Images from the selected studies were assessed using ImageJ, a public domain Java image processing software package.23

As blue-light radiance for different greyscale values was known from spectral measurements, it was possible to deduce blue-light radiance for the different average image intensities of selected modalities (Figure 6).

Ethical considerations

An application for full ethical approval was made to the Health Sciences Research Ethics Committee of the University of the Free State and ethical approval was received on 10 June 2021 (ref. no. UFS-HSD2021/0153/2906). All procedures performed in studies involving human participants were in accordance with the ethical standards of the institutional and national research committee and with the 1964 Helsinki Declaration and its later amendments or comparable ethical standards. Written informed consent was obtained from all individual participants involved in the study.

Results

The average viewing time that registrars spent in front of DDDs on a standard workday was determined as 6 h 20 min (380 min) with a median of 06 h15 min (375 min). This included time spent reporting and reviewing images on workstation monitors and excluded times when registrars were physically away from their stations. The average viewing time of 380 min was used for further calculations (Table 2). All participants were considered senior registrars and were within 14 months from sitting for their final exit exams.

Effective blue-light radiance, LB, was determined for a spectrum of different greyscale values according to Equation 1. Summation of measured spectral values and weighting with the blue-light hazard function to encompass sensitivity of the normal and aphakic human eye was performed. A paired-sample t-test was conducted to compare effective blue-light radiance, LB, with hazard function weighting for normal and aphakic eyes applied. There was no significant difference in the results of effective blue-light radiance, LB, between using normal eye hazard function (mean = 0.137, standard deviation [s.d.] = 0.117) and aphakic hazard function (mean = 0.137, s.d. = 0.118); t(13) = −0.081, p = 0.936. Only normal hazard function weighting was thus used. Using the relation of average image intensity (Iavg) to a specific greyscale value, the effective blue-light radiance, LB, was determined for CXR, CT brain and MRI brain (Table 3).

Average image intensities were determined as 56 for CXR, 38 for CT brain and 32 for MRI brain. These values correspond to 56, 38 and 32 on the greyscale, respectively. For visual clarity, the authors represent this on a logarithmic scale, as the measured results were very small and the difference between the threshold and measured values was substantial (Figure 7).

According to ICNIRP and ACGIH® guidelines, as the average daily viewing time of registrars exceeds 167 min, recommended acceptable exposure falls in category 3 in Table 1. According to Equation 2, effective blue-light radiance in this group should be below 0.01 W/cm216,17t-test was performed to compare the mean effective blue-light radiance, LB, to the threshold limit of acceptable blue-light exposure at viewing times exceeding 167 min per day. The mean value of effective blue-light radiance, LB, (mean = 0.137, s.d. = 0.118) was significantly different from the threshold limit of acceptable blue-light exposure at viewing times exceeding 167 min per day; t(13) = −318216.031, p < 0.001. It was clear that the determined blue-light radiance values were drastically lower than the recommended threshold limit value represented by the dotted line (Figure 7). Daily blue-light exposure in terms of blue-light radiance was concluded to be more than 10 000 times below the maximum limit.

Discussion

According to the CIE, light sources with luminance of less than 10 000 cd/m224225

Unlike the well-defined deterministic dose ranges for ionising radiation dose effects, specific phototoxic retinal dosages have, according to the authors' knowledge, mainly been described in animal models and in vitro studies thus far.15,26,2728® and ICNIRP recommends exposure values in terms of ranges of acceptable effective blue-light radiance and maximum dose approximations, but exact dosages remain vague.

The retinal dosage of blue light is determined by many factors, including properties of the eye itself, physiological and health status of the viewer and external influences. Pupil diameter, focal length, eye movements and characteristics of the structures comprising the eye regulate the area of retina that is exposed.17,29302929,31,3233

Effectively, eye movement will enlarge the area of the retina upon which blue light photons are incident. In case of prolonged viewing with decreased eye movements, such as intense evaluation of an image, the blue light dosage to the retina will thus be higher.17343536

This assessment of blue-light exposure in registrars yielded effective blue-light radiance values of more than 10 000 times below the recommended safe value of 0.1 W/cm262237

Limitations of this study include the small study population, self-reported viewing times and the exclusion of the multitude of variables that may influence retinal blue-light deposition as described.The viewing time determined by the authors was carried out using a basic method of activating a stopwatch while reporting. Independent verification of viewing times by means of an objective observer measuring time or assessment of login time data from PACS was not performed. Although all registrars were dedicated and actively participating in the process, it should be considered that viewing times may have been either over- or underestimated because of human error.

The study's evaluation was carried out using three common imaging modalities, that is, CXR, CT brain and MRI brain. These modalities are often the workhorses in radiology, but a wide variety of other types of images are also read. The brightness of the viewed content will determine blue-light exposure; hence, brighter-appearing images such as fluoroscopic studies will most likely yield more blue light than a darker mammography study. Other modalities were omitted for the sake of simplicity and reproducibility of the study.

The DDD units at the institution under study are compliant with local regulatory standards, but each device remains unique. As a result of technical variances, different display units may, within limits, vary in their emission of blue light. It is also important to consider that DDD units are not the only devices emitting blue light that radiology practitioners are exposed to. Often, practitioners will review images at machine terminals immediately after acquisition to decide whether further images or sequences are required. These monitors also emit blue light but were not included in our measurements. Furthermore, in the current era of electronic device usage for a multitude professional and personal duties, a large component of blue-light exposure that will add to retinal dosage is assumed to come from these devices. Although beyond the scope of this study, this method can be used in future research to quantify blue light from, for instance, cellular phones, tablets, personal computers and televisions. Based on the current study's results and the CIE position on sources with a luminance of less than 10 000 cd/m224

Although measured values were well below threshold limits, a clear relation between the viewed content and blue-light radiance was determined. For darker images with lower average intensities such as MRI, less blue-light radiance was determined than for brighter chest radiographs. Thus, if brighter images with increased content in the upper ranges of the greyscale are viewed, more effective blue-light radiance will be measured. However, as evident from the higher greyscale values in Figure 7, if an individual was viewing even maximal greyscale value content (i.e. a white screen) for the duration of one working day, blue light radiance would still be well within the ranges considered safe.

Recent studies were not able to establish clear evidence to advocate for the use of glasses with blue light-blocking lenses. Available research on blue-blocking lenses is mainly based on animal data and laboratory-centred experiments rather than extensive human clinical trials, leaving potential paucities in clinical applications.38,3940

Future research in a study population consisting of a wider variety of practising radiologists, including the public and private sectors, may yield sensible information regarding the role that different work circumstances play. Other imaging modalities can be assessed to provide a more exhaustive overview of blue-light exposure. Independent observation of viewing time by means of computer login data or motion sensors can be considered to reduce human error that may occur with self-measured viewing times. With minor modifications, the methods and results of this study can essentially be extrapolated to any industry requiring employees to work on a VDU and where there is concern regarding occupational blue-light exposure.

Conclusion

The dangers of blue light and its effect on ocular and general health are well documented. The scope of practice requires healthcare practitioners in diagnostic radiology to spend long hours in front of DDD units that use LED technology, which emit high amounts of blue light. The authors' concern was that this may predispose radiology practitioners to an occupational hazard related to blue light and its sequelae. It was found that the average time a radiology registrar spends in front of a monitor per workday is 380 min. The content that was being viewed played a part in the blue light exposure. Brighter chest radiographs demonstrated more blue-light radiance than darker MRI brain images. Nonetheless, exposure to effective blue-light radiance from monitors during a standard working day was far below the ranges that are recognised as hazardous.

Competing interests

The authors declare that they have no personal or financial incentives which may have influenced them in the writing of this article.

Authors' contributions

M.W. was the principal investigator and prepared the manuscript. J.J.T. contributed with conceptualising of measurements and the mathematical model. J.J.v.R. provided critical revision and the final draft overview of the article.

Funding information

This research received no specific grant from any funding agency in the public, commercial or not-for-profit sectors.

Data availability

Data pertaining to this research are available from the corresponding author, M.W.

Disclaimer

The views and opinions expressed in this article are those of the authors and do not necessarily reflect the official policy or position of any affiliated agency of the authors.

Acknowledgements

The authors would like to thank Ms C. Stofile, Department of Medical Physics, University of the Free State, for helping with measurements and providing input regarding quality control standards; Mr C. van Rooyen, Department of Biostatistics, University of the Free State, for his help with statistical analysis of the data; registrars from the Department of Clinical Imaging Sciences who participated in the study, namely Dr Lilanie Biddulph, Dr Philip Zeelie, Dr Francois van der Merwe and Dr Jandus de Villiers.

References

  • 1.Ouyang X, Yang J, Hong Z, Wu Y, Xie Y, Wang G. Mechanisms of blue light-induced eye hazard and protective measures: A review. Biomed Pharmacother. 2020;130(June):110577. https://doi.org/10.1016/j.biopha.2020.110577
    » https://doi.org/10.1016/j.biopha.2020.110577
  • 2.Coles-Brennan C, Sulley A, Young G. Management of digital eye strain. Clin Exp Optom. 2019;102(1):18-29. https://doi.org/10.1111/cxo.127983.
    » https://doi.org/10.1111/cxo.127983
  • 3.Cheng X, Song M, Kong J, et al. Influence of prolonged visual display terminal Use und exercise on physical and mental conditions of internet staff in Hangzhou, China. Int J Environ Res Public Health. 2019;16(1829):1-11. https://doi.org/10.3390/ijerph16101829
    » https://doi.org/10.3390/ijerph16101829
  • 4.Arnetz BB, Berg M. Melatonin and adrenocorticotropic hormone levels in video display unit workers during work and leisure. J Occup Environ Med. 1996;38(11):1108-1110. https://doi.org/10.1097/00043764-199611000-00010
    » https://doi.org/10.1097/00043764-199611000-00010
  • 5.Kluger N, Pecquet C. Dermatoses associated with high technology (cell phones, computers and video games). Ann Dermatol Venereol. 2012;139(3):230-238. https://doi.org/10.1016/j.annder.2011.09.192
    » https://doi.org/10.1016/j.annder.2011.09.192
  • 6.Chawla A, Lim TC, Shikhare SN, Munk PL, Peh WCG. Computer vision syndrome: Darkness under the shadow of light. Can Assoc Radiol J. 2019;70(1):5-9. https://doi.org/10.1016/j.carj.2018.10.005
    » https://doi.org/10.1016/j.carj.2018.10.005
  • 7.Aronsson K, Teär Fahnhjelm K, Nylén P, Eklund J. Visual ergonomics and eye strain in eye care professionals [homepage on the Internet]. NES conference in Saltsjöbaden. 2012 [cited 2022 July 05]. Available from: http://kth.diva-portal.org/smash/record.jsf?pid=diva2%3A601718&dswid=9560
    » http://kth.diva-portal.org/smash/record.jsf?pid=diva2%3A601718& dswid=9560
  • 8.Al Dandan O, Hassan A, Al Shammari M, Al Jawad M, Alsaif HS, Alarfaj K. Digital eye strain among radiologists: A survey-based cross-sectional study. Acad Radiol. 2021;28(8):1142-1148. https://doi.org/10.1016/j.acra.2020.05.006
    » https://doi.org/10.1016/j.acra.2020.05.006
  • 9.Atwal SS, Prasad A, Deepak D, Agarwal K. Health issues among radiologists: Toll they pay to their profession. J Clin Diagn Res. 2017;11(4):TJ01-TJ02. https://doi.org/10.7860/JCDR/2017/17023.9537
    » https://doi.org/10.7860/JCDR/2017/17023.9537
  • 10.Halpenny D, O'Driscoll D, Torreggiani WC. Ocular health among radiologists in the age of PACS: Is it time for our profession to open its eyes to this issue in light of existing European legislation? Br J Radiol. 2012;85(1020):1309-1311. https://doi.org/10.1259/bjr/27200799
    » https://doi.org/10.1259/bjr/27200799
  • 11.Naveen B. Evaluation of non-radiation occupational hazards faced by radiologists. Int J Radiol Diagn Imaging. 2020;3(2):18-20. https://doi.org/10.33545/26644436.2020.v3.i2a.91
    » https://doi.org/10.33545/26644436.2020.v3.i2a.91
  • 12.Bedrosian TA, Nelson RJ. Timing of light exposure affects mood and brain circuits. Transl Psychiatry. 2017;7(1):e1017. https://doi.org/10.1038/tp.2016.262
    » https://doi.org/10.1038/tp.2016.262
  • 13.Sirsawy U, Steinberg WJ, Raubenheimer JE. Levels of burnout among registrars and medical officers working at Bloemfontein public healthcare facilities in 2013. S Afr Fam Pract. 2016;58(6):213-218. https://doi.org/10.1080/20786190.2016.1198088
    » https://doi.org/10.1080/20786190.2016.1198088
  • 14.Vicente-Tejedor J, Marchena M, Ramírez L, et al. Removal of the blue component of light significantly decreases retinal damage after high intensity exposure. PLoS One. 2018;13(3):e0194218. https://doi.org/10.1371/journal.pone.0194218
    » https://doi.org/10.1371/journal.pone.0194218
  • 15.Commission Internationale de l'Eclairage. Position statement on the Blue Light Hazard [homepage on the Internet]. Commission Internationale de l'Eclairage; 2019 [cited 2022 Mar 28]. Available from: https://cie.co.at/publications/positionstatement-blue-light-hazard-april-23-2019#:~:text=Whilst%20the%20CIE%20considers%20that,approach%20the%20blue%20light%20hazard
    » https://cie.co.at/publications/positionstatement-blue-light-hazard-april-23-2019#:~:text=Whilst%20the%20CIE%20considers%20that,approach%20the%20blue%20light%20hazard
  • 16.The American Conference of Governmental Industrial Hygienists. ACGIH® Threshold Limit Values (TLVs) and Biological Exposure Indices (BEIs). Cincinnati: Signature Publications, 2012; p. 871-1130.
  • 17.Ziegelberger G. ICNIRP guidelines on limits of exposure to incoherent visible and infrared radiation. Health Phys. 2013;105(1):74-96. https://doi.org/10.1097/HP.0b013e318289a611
    » https://doi.org/10.1097/HP.0b013e318289a611
  • 18.Bukshtab M. Photometry, radiometry, and measurements of optical losses [homepage on the Internet]. 2nd ed. Volume 209. Springer Series in Optical Sciences. Singapore: Springer Singapore, 2019; 15-25, 61 p. Available from: http://link.springer.com/10.1007/978-981-10-7745-6
    » http://link.springer.com/10.1007/978-981-10-7745-6
  • 19.Heo SY, Kim J, Gutruf P, et al. Wireless, battery-free, flexible, miniaturized dosimeters monitor exposure to solar radiation and to light for phototherapy. Sci Transl Med. 2018;10:eaau1643. https://doi.org/10.1126/scitranslmed.aau1643
    » https://doi.org/10.1126/scitranslmed.aau1643
  • 20.Department of Health Directorate. Requirements for licence holders with respect to quality control tests for diagnostic X-ray imaging systems [homepage on the Internet]. 2012 [cited 2022 Mar 24]. Available from: http://radiationsafe.co.za/wp-content/uploads/2017/08/1.4-Requirements-for-licenceholders-wrt-to-QC-tests-for-diag-x-ray-imaging-systems-2012.pdf
    » http://radiationsafe.co.za/wp-content/uploads/2017/08/1.4-Requirements-for-licenceholders-wrt-to-QC-tests-for-diag-x-ray-imaging-systems-2012.pdf
  • 21.Agfa South Africa (Pty) Ltd Inspection Body. Acceptance test protocol scope 1: Reporting monitors [homepage on the Internet]. 2021 [cited 2022 Mar 24]. Available from: http://INSPECTION%20BODY%20RESULTS/RM%20Universitas%20Academic%20%20Hospital/WS%2013/ Universitas%20Academic%20Hospital_Radiology_Scope%201_Dual%20Set_WS%2013_SR%23_14594.pdf
    » http://INSPECTION%20BODY%20RESULTS/RM%20Universitas%20Academic%20%20Hospital/WS%2013/ Universitas%20Academic%20Hospital_Radiology_Scope%201_Dual%20Set_WS%2013_SR%23_14594.pdf
  • 22.Gonzalez RC, Woods RE. Digital image processing. 2nd ed. Upper Saddle River, NJ: Prentice Hall, 2002; p. 88-95.
  • 23.Schneider CA, Rasband WS, Eliceiri KW. NIH Image to ImageJ: 25 years of image analysis. Nat Methods. 2012;9(7):671-675. https://doi.org/10.1038/nmeth.2089
    » https://doi.org/10.1038/nmeth.2089
  • 24.Commission Internationale de l'Eclairage. Photobiological safety of lamps and lamp systems. Standard No. CIE S 009. Vienna: CIE; 2006.
  • 25.Zhao ZC, Zhou Y, Tan G, Li J. Research progress about the effect and prevention of blue light on eyes. Int J Ophthalmol. 2018;11(12):1999-2003. https://doi.org/10.18240/ijo.2018.12.20
    » https://doi.org/10.18240/ijo.2018.12.20
  • 26.Ham WT, Mueller HA, Sliney DH. Retinal sensitivity to damage from short wavelength light. Nature. 1976;260:153. https://doi.org/10.1038/260153a0
    » https://doi.org/10.1038/260153a0
  • 27.Jaadane I, Villalpando Rodriguez G, Boulenguez P, et al. Retinal phototoxicity and the evaluation of the blue light hazard of a new solid-state lighting technology. Sci Rep. 2020;10(1):6733. https://doi.org/10.1038/s41598-020-63442-5
    » https://doi.org/10.1038/s41598-020-63442-5
  • 28.Krigel A, Berdugo M, Picard E, et al. Light-induced retinal damage using different light sources, protocols and rat strains reveals LED phototoxicity [homepage on the Internet]. Elsevier International Brain Research Organization; 2016. [cited 2020 July 04]. Available from: https://hal.sorbonne-universite.fr/hal-01383394
    » https://hal.sorbonne-universite.fr/hal-01383394
  • 29.Kelbsch C, Strasser T, Chen Y, et al. Standards in pupillography. Front Neurol. 2019;10:129. https://doi.org/10.3389/fneur.2019.00129
    » https://doi.org/10.3389/fneur.2019.00129
  • 30.Heo JY, Kim K, Fava M, et al. Effects of smartphone use with and without blue light at night in healthy adults: A randomized, double-blind, cross-over, placebo-controlled comparison. J Psychiatr Res. 2017;87(2017):61-70. https://doi.org/10.1016/j.jpsychires.2016.12.010
    » https://doi.org/10.1016/j.jpsychires.2016.12.010
  • 31.Karavanaki K, Davies AG, Hunt LP, Morgan MH, Baum JD. Pupil size in diabetes. Arch Dis Child. 1994;71(6):511-515. https://doi.org/10.1136/adc.71.6.511
    » https://doi.org/10.1136/adc.71.6.511
  • 32.Park JW, Kang BH, Kwon JW, Cho KJ. Analysis of various factors affecting pupil size in patients with glaucoma. BMC Ophthalmol. 2017;17(1):1-8. https://doi.org/10.1186/s12886-017-0564-6
    » https://doi.org/10.1186/s12886-017-0564-6
  • 33.Sharma S, Baskaran M, Rukmini AV, et al. Factors influencing the pupillary light reflex in healthy individuals. Graefe's Arch Clin Exp Ophthalmol. 2016;254:1353-1359. https://doi.org/10.1007/s00417-016-3311-4
    » https://doi.org/10.1007/s00417-016-3311-4
  • 34.O'Hagan JB, Khazova M, Price LLA. Low-energy light bulbs, computers, tablets and the blue light hazard. Eye. 2016;30:230-233. https://doi.org/10.1038/eye.2015.261
    » https://doi.org/10.1038/eye.2015.261
  • 35.Jaschinski W, Heuer H. Preferred position of visual displays relative to the eyes: A field study of visual strain and individual differences. Ergonomics. 1998;41(7):1034-1049. https://doi.org/10.1080/001401398186586
    » https://doi.org/10.1080/001401398186586
  • 36.Zekveld AA, Koelewijn T, Kramer SE. The pupil dilation response to auditory stimuli: Current state of knowledge. Trends Hear. 2018;22:1-25. https://doi.org/10.1177/2331216518777174
    » https://doi.org/10.1177/2331216518777174
  • 37.Bullough JD, Bierman A, Rea MS. Evaluating the blue-light hazard from solid state lighting. Int J Occup Saf Ergon. 2019;25(2):311-320. https://doi.org/10.1080/10803548.2017.1375172
    » https://doi.org/10.1080/10803548.2017.1375172
  • 38.Vagge A, Ferro Desideri L, Del Noce C, Di Mola I, Sindaco D, Traverso CE. Blue light filtering ophthalmic lenses: A systematic review. Semin Ophthalmol 36(7):541-548. https://doi.org/10.1080/08820538.2021.1900283
    » https://doi.org/10.1080/08820538.2021.1900283
  • 39.Singh S, Downie LE, Anderson AJ. Do Blue-blocking lenses reduce eye strain from extended screen time? A double-masked randomized controlled trial. Am J Ophthalmol. 2021;226:243-251. https://doi.org/10.1016/j.ajo.2021.02.010
    » https://doi.org/10.1016/j.ajo.2021.02.010
  • 40.Mainster MA, Findl O, Dick HB, et al. The blue-light hazard vs. blue-light-hype. Am J Ophthalmol. 2022;240:51-57. https://doi.org/10.1016/j.ajo.2022.02.016
    » https://doi.org/10.1016/j.ajo.2022.02.016

Correspondence:
Mari Wentzel

Received: 25 July 2022
Accepted: 28 Nov. 2022
Published: 31 Jan. 2023

Correspondence

wentzelmari@gmail.com

Publication Dates

  • Publication in this collection
    01 Feb 2024
  • Date of issue

History

  • Received
    25 July 2022
  • Accepted
    28 Nov 2022
location_on
Radiological Society of South Africa AOSIS Publishing, Postnet Suite #110, Private Bag X19, Durbanville, Cape Town, Western Cape Province, ZA, 7551, Tel: +27 21 975 2602, Tel: 086 1000 381 - E-mail: publishing@aosis.co.za
rss_feed Acompañe los números de esta revista en su lector de RSS
Ir para arriba