Open-access Clinical diagnostic reference levels and image quality metrics for CT in oncology patients

Abstract

BACKGROUND: In CT, the volumetric CT dose index (CTDIvol), dose-length product (DLP) and patient's size-specific dose estimates (SSDE) are used as diagnostic reference level (DRL) metrics. OBJECTIVES: To develop clinical local DRL values for CT chest-abdomen-pelvis (CAP) examinations using the CTDIvol, DLP and SSDE, and to determine the image quality achieved. METHOD: In total, 201 cancer patients were included in the study. The scanning parameters, dose metrics from the CT unit and participants' body mass index (BMI) were documented. The local CT DRL values for CAP examinations were defined as the median and 75th percentiles of the dose distribution. RESULTS: The local DRL values given in terms of median CTDIvol ranged between 8.4 mGy and 12.7 mGy for the different types of cancers. The median DLPs ranged from 848 to 1173.4 mGy.cm for the various cancers. Generally, the radiation dose was directly proportional to the BMI and number of scan phases. Significant differences were observed between the DRLs for the various size-related parameters, number of scan phases and BMI classifications. The image quality was clinically satisfactory. CONCLUSION: No baseline data for clinical DRL values were available for this medical oncology department. The achieved DRLs were similar to published size-specific DRLs. The image quality was maintained during CT imaging. Dose optimisation and image quality assessment should be implemented to ensure optimal scanning parameters for different cancers in CT CAP examinations. CONTRIBUTION: The first size-specific local DRL values for CT CAP investigations carried out on oncology patients in South Africa have been established.

cancer; computed tomography; local diagnostic reference level; image quality index; size-specific dose estimates


Introduction

Justification and optimisation are two important principles when ionising radiation is used for medical purposes.11

In planned diagnostic medical imaging such as computed tomography (CT), calculating the diagnostic reference level (DRL) values contributes to the optimisation of patient doses delivered for this imaging modality.1vol) and/or dose-length product (DLP) metrics.2,3,4,5,6,7,8 Size-specific dose estimates (SSDE), based on the patient's size, are another metric to define DRL values, introduced by the American Association of Physicists in Medicine (AAPM)910,11,12,13 the SSDE is not widely used as it is not routinely generated on the patient radiation dose structured report produced by CT units, contrary to the CTDIvol and DLP.14,15

Clinical DRL values have also been published based on the CTDIvol and DLP,16,17,18 and only two studies have been published presenting clinical DRL values (CTDIvol and DLP) that were patient size specific.17,19 No oncology-specific clinical DRL values for SSDE based on the size of the patient could be traced in the literature.

The clinical indication protocol further influences the number of scan phases and in cases of multiphase examinations, the dose to the patient. Some multiphase CT examinations can consist of two to four scan phases depending on the clinical indication and protocol. These phases may include the non-contrast phase, arterial phase, portal venous or venous phase and delayed or late phase. During the arterial phase, obtained at 35 s - 40 s after contrast medium injection, pathologies such as hepatocellular carcinoma and adenoma in the liver can be identified.2020 2122222123,24 Therefore, when DRL values are calculated, multiphase examinations should be calculated separately from single-phase CT examinations.25

The South African Health Products Regulatory Authority (SAHPRA) also recommends DRL values for the most commonly performed examinations in medical imaging.262728

Research methods and design

Two hundred and one patients aged 18 years or older, who were scheduled for CT CAP examinations and provided informed consent to have their weight and height measured for calculation of their body mass index (BMI), were included in the study. The clinical indication for all the patients was cancer-related. The CT CAP examinations that formed part of the study were for diagnostic purposes and not for radiation therapy planning. The participants were not traced for other dose-related examinations or treatments apart from the CT CAP examinations for cancer evaluation, oncologic follow-up and/or staging. The different types of cancer were grouped according to anatomical systems. Cervix, uterus, vulva, ovarian and endometrial cancer were grouped together as part of the female reproductive system. Oesophageal, stomach, colon or rectosigmoid, anal, rectal and caecal cancer formed part of the gastrointestinal tract (GIT) of the digestive system. Breast cancer was the most common clinical indication for the CT CAP examination in this study and was not grouped with another anatomical system.

Before the commencement of data collection, the MDW 300 L scale (Adam Equipment SA [Pty] Ltd; Johannesburg, South Africa), used to measure the BMI, was calibrated. Patient data were collected from 01 March 2021 to 10 June 2021. The medical oncology department and the patients for the study were selected using a convenience sampling technique. The sample (201 patients) was divided into four different groups according to their BMI classifications (underweight: < 18.5 kg/m222222

Data were obtained from the GE BrightSpeed 8 CT unit (General Electric Healthcare; Chicago, IL, USA) that was installed in 2007 at the medical oncology department as a demonstration unit. The scan was performed with a body filter, using a standard convolution kernel in helical acquisition mode. The department used an image reconstruction algorithm and automatic tube current modulation (ATCM). The ATCM used was SmartmA and AutomA, while the reconstruction algorithm used was the filtered back projection (FBP). One protocol was used to scan all 201 patients scheduled for CT CAP examinations. The CT acquisition parameters are displayed in Table 1. At the start of data collection, the pitch was 1.675, but the pitch was later changed to 1.35 for the protocol. Only the number of scan phases, the scan length and scan time varied among the patients. The CT images included in the study were reported by radiologists at the medical oncology department.

Research instrument

A patient-dose document that was quantitative in nature was compiled by the researcher and benchmarked with the literature29,30 for use in this study. The patient-dose document descriptively collected specific information from the CT unit and measurements from the patients that were used to develop DRL values. This included information pertaining to the patient such as the BMI, weight and age (obtained prior to the examination after informed consent had been provided). Numerical data such as the rotation time, pitch, CTDIvol and tube current from the CT unit were captured on the patient-dose document for each patient in the study. Information such as the clinical indication and use of contrast medium was recorded.

All the CT images were stored on an external hard drive in Digital Imaging and Communications in Medicine (DICOM) format so that anterior-posterior (AP) and lateral (LAT) measurements on the CT images could be determined, measured using electronic callipers on a MicroDicom (mDicom) viewer (MicroDicom Ltd.; Sofia, Bulgaria),28

Dose metric - Size-specific dose estimates

The five patient size-related parameters are the sum of AP and LAT dimensions, AP dimension, effective diameter, LAT dimension and BMI (non-specific) for SSDE calculation.28931BMI (non-specific) refers to the SSDE that was calculated using the patient's BMI. Thus, the patient size (BMI) was taken into consideration to calculate the SSDE:

Data analysis

The data from the patient-dose document were transferred to a Microsoft Excel Version 365 (Microsoft Corporation; Redmond, WA, USA) spreadsheet and minimum, maximum and mean values were determined. The medians (50th percentile) and 75th percentiles were established for the CTDIvol, DLP and the five size-related parameters for SSDE28p < 0.05 was used. Furthermore, the Spearman correlation coefficient was used to investigate whether a correlation between the size-related parameters and DRL quantities was noted. A significant correlation between the size-related parameters and DRL quantities was assumed when p < 0.05.

Image quality index

For each patient CT image sequence, adjacent images were subtracted from each other using the MATrix LABoratory (MATLAB) R2017a (The Math Works, Inc., Natick, MA, USA).28Figure 1. In a study by Moghadam et al.,1132

Figure 1

Ethical considerations

The Health Sciences Research Ethics Committee (HSREC) of the University of the Free State in Bloemfontein, South Africa (Ethics clearance number: UFS-HSD2019/0612/011002) provided ethical approval for the research. Permission was granted by the Free State Province Department of Health to acquire patient information from the CT unit because data were collected at a public health institution. Written informed consent was provided by all the patients who participated in the study.

Results

A total of 201 adult patients were referred to the medical oncology department for CT CAP examinations during the period 01 March 2021 to 10 June 2021. Three-quarters (n = 151; 75.0%) of the patients were females. The patients' age ranged between 22 years and 87 years (median age 52 years). All the patients referred for a CT CAP examination had cancer-related clinical indications such as breast, penile, lung, rectal and vulva cancer. One-third (n = 65; 32.3%) of the patients displayed a normal BMI (18.5 kg/m22

The findings in this study are the first set of clinical local DRL (LDRL) values (median and 75th percentile) at this particular medical oncology department for CT CAP clinical size-specific DRL values for specific types of cancer. The three main cancers recorded in this study were, (1) breast cancer (two scan phases) with 82 patients, (2) cervical, uterus, vulva, ovarian and endometrial cancer (two scan phases) with 30 patients and (3) oesophageal, stomach, colon and/or rectosigmoid, anal, rectal and caecum cancer (two scan phases) with 32 patients (Figure 2 and Figure 3).

Figure 2

Figure 3

For penile, testicular and prostate cancer (male reproductive system), only 13 patients underwent two scan phases, and one patient underwent three scan phases. In relation to renal and bladder cancer (urinary system), only three patients underwent two scan phases. The same trend of less than 20 cases was noted for (1) pancreas, liver and bile duct cancer (two scan phases), (2) breast cancer (three scan phases), (3) cervical, uterus, vulva, ovarian and endometrial cancer (three scan phases), (4) oesophageal, stomach, colon and/or rectosigmoid, anal, rectal and caecum cancer (three scan phases), (5) sarcoma, scalp squamous cell carcinoma and melanoma (two scan phases), (6) lung cancer (two scan phases) and (7) the exclusion or query of cancer (two scan phases).

Figure 2 and Figure 3 illustrate the clinical LDRL values for specific cancers based on various dose metrics. Table 2 gives the summary of the clinical LDRL values for non-specific cancers and a number of scan phases as the median and 75th percentile for the different dose metrics based on the non-specific BMI and BMI classifications. Significant differences ranging from < 0.0001 to 0.0321 were noted between the use of two and three scan phases based on the radiation dose the patient received. A strong significant positive correlation (0.82-0.86) was noted between the size-related parameters (SSDEBMI [non-specific], SSDESUM Dimension, SSDEAP Dimension, SSDEEffective diameter and SSDELAT Dimension). The clinical DRL values based on the DLP and CTDIvol of this study were compared with the DRL values published by Habib Geryes et al.19 Table 3.

Table 2

Table 3

The noise index (NI) at the medical oncology department was kept constant for all the CT CAP examinations, as displayed in Table 1. The liver displayed in the axial portal venous phase of the CT examination was used to measure the image quality objectively in this study for all patients and the different BMI classifications, as seen in Table 2.

Discussion

The CT unit is not equipped with the latest available technology, such as adaptive statistical iterative reconstruction (ASIR), to promote further radiation dose reduction to the patient, which is a common scenario in resource-restricted public healthcare facilities in South Africa. Financial constraints in government institutions could be one of the reasons why the latest technology was not used at the participating department. However, the ATCM was activated at the medical oncology department to reduce the dose by adjusting the exposure parameters according to patient sizes, while maintaining the image quality of the CT images.33,34,35 Objective image quality assessment of the CT images was done using MathWorks® imtool3D_td in MATLAB (https://www.mathworks.com/matlabcentral/fileexchange/74761-imtool3d_td), which was satisfactory for all the patients included in the study.

In prior studies,12,34 the image quality was measured over various regions of the body for CT CAP and abdomen-pelvis, such as the aorta, liver and spleen. However, in a study by Moghadam et al.1136,37 Furthermore, a radiology report was provided for all the examinations included in the study at the participating hospital, which also indicated that the images were of acceptable diagnostic value.

The axial portal venous phase was used to assess the image quality in this study, as described by Ahmad et al.,373737

Clinical DRL values based on the median value of the dose metrics for the patients included in the study and the 75th percentile for those dose metrics to give an indication of the range of the dose metric values could only be calculated for the most common cancers recorded in this study because 20 or more patients are usually considered to be sufficient to establish DRL values, as indicated by the ICRP.14

Although the same anatomical region was scanned for the same main clinical indication (cancer), the dose differed among the various cancers. There were no significant differences (p = 0.12) between the median values of the SSDEBMI (non-specific) for the specific types of cancer, which was expected, because the same clinical protocol was used, also described by Pema et al.3BMI [non-specific]) among the specific cancers, calculating clinical DRL values will enable radiology departments to easily identify areas to review and/or adjust their respective protocols, while ensuring that optimal image quality is maintained to provide a diagnosis.

Clinical LDRL values were also calculated for non-specific cancers (all cancers) and the number of scan phases (201 patients) (Table 2). Underweight patients were excluded because there were less than 20 patients in that category. Furthermore, the image quality index based on these two variables is also highlighted in Table 2. All the patients were injected with contrast medium intravenously. The majority of patients received two scan phases (arterial and portal venous) for CT CAP examinations, while 20 (10.0%) patients received three scan phases (arterial, portal venous and delay). An increase in the number of scan phases resulted in a dose increase for all dose metrics, which was also evident in previous studies.23,24

A significant difference between the median values for SSDEBMI (non-specific) (p = 0.0013), DLP (p < 0.0001) and CTDIvol (p = 0.0321) was observed for the number of scan phases (two or three). Radiologists decide on the number of scan phases (protocol) a patient should receive based on the pathology and clinical history indicated on the X-ray request form. It is important that the protocol applied for cancer diagnosis examinations is justified and that the information provided by the scan phases contributes to image diagnosis.24

The radiation dose increased for all the dose metrics when the BMI increased from normal weight to obese (Table 2). However, overweight patients received a higher dose than obese patients for the specific size-related SSDE dose metrics SSDEBMI (non-specific), SSDELAT Dimension, SSDESUM Dimension and SSDEEffective diameter, as displayed in Table 2. A possible reason for this could perhaps be attributed to the different patient habitus3839AP Dimension (p = 0.1094), SSDELAT Dimension (p = 0.7621), SSDESUM Dimension (p = 0.5036) and SSDEEffective diameter (p = 0.5172). However, a significant difference was noted between the overweight and obese patients for SSDEBMI (p = 0.0015). A possible reason why a significant difference occurred between overweight and obese patients for SSDEBMI could be because there are little to no differences among researchers in measuring the BMI, whereas researchers measure the AP and LAT dimensions differently to calculate the SSDE.31

The comparison of the DRL values between this study and that of Habib Geryes et al.19119191919 Table 3. The DRL values for CAP cancer were higher in this study compared to those reported by Habib Geryes et al.,1919

The overweight patients with breast cancer had a DLP of 1077.8 mGy.cm, while the DLP was 1263.0 mGy.cm for obese patients based on the median. Clear differences for the DLP were noted among the BMI classifications for breast cancer. A significant difference (p < 0.0001) was also observed between the median values for SSDEBMI for the different BMI classifications, which emphasises the importance of calculating clinical DRL values for specific BMI groups as well.

The DLP differed among the specific cancers, which might be attributed to the longer scan lengths that were obtained. Among the three cancer groups, GIT organ-related cancer had a longer mean scan length of 394 mm compared to breast cancer and cancer of the reproductive system, which had a mean scan length of 379 mm and 388 mm, respectively. A feasible explanation could be that a longer scan length was used for patients with oesophageal, stomach, colon and/or rectosigmoid, anal, rectal and caecum cancer not to miss any possible metastases, and therefore, the DLP increased, which was also evident in previous studies.5,7

A strong positive correlation (0.82-0.86) was noted between the size-related parameters; therefore, it can be concluded that any size-related parameter can be used to determine the SSDE for cancer patients. However, using the BMI instead of the AP and LAT dimensions to calculate the SSDE will result in minimal measurement inconsistency.31p < 0.0001) was also observed between the size-related parameters. A significant difference (p < 0.0001) between the CTDIvol, DLP and the SSDE for all the BMI classifications was noted, indicating a significant effect of these quantities on the DRL. Deciding on which DRL quantity to use in CT imaging to develop DRL values is important. Research has shown that it is better to use the SSDE to calculate patient doses because the DRL quantity is based on the size of the patient.40vol and the SSDE for all the BMI classifications. As the BMI increases, the dose the patient receives also increases.

A convenience sampling technique was used to select the participating medical oncology department because the department was in close proximity to the researcher. The patients who met the inclusion criteria at the participating medical oncology department were also selected using the convenience sampling technique, because these patients were already booked for CT CAP examinations, their clinical indication was cancer-related and the patients were easily accessible. However, a good representation of the underweight patients could not be achieved using the convenience sampling technique. As a result, the main limitation noted for the study was that the clinical DRL values could not be developed for underweight patients because there were less than 20 patients in this BMI classification. Another limitation was that only one medical oncology department was included in the study to evaluate the patient doses and image quality metrics. Additionally, an image quality scoring system was not utilised to assess subjective image quality. Because the radiologist reported on the CT images without commenting negatively about the image quality, the authors presumed that the image quality was of diagnostic value. Furthermore, the patient doses and image quality metrics calculated only focussed on CT CAP examinations for adult patients. Moreover, the image quality was only determined on the liver in this study and not on the other internal organs such as the spleen and intestines.

Future studies are necessary to determine DRL values based on the clinical indication for underweight patients, and to investigate a larger variety of different types of cancer, evaluating them in terms of BMI classifications and the same number of scan phases. In addition, an image quality scoring system should be utilised to assess subjective image quality when clinical DRL values are established. Furthermore, image quality metrics and clinical DRL values should be assessed for paediatric CT examinations in future studies.

Conclusion

At the time of the study, no baseline DRL values were available at this particular medical oncology department. For future diagnostic CT examinations, the results of this study could be applied as a starting point. Dose differences occurred among the specific cancers and the different BMI classifications. Hence, it is essential to calculate DRL values based on patients' sizes. Furthermore, significant differences were also noted among the number of scan phases, which clearly indicated that the number of scan phases influences the dose. The quality of these images was diagnostically and clinically adequate for drawing a conclusion on the patients' cancer status.

Competing interests

The authors declare that they have no financial or personal relationships that may have inappropriately influenced them in writing this article.

Authors' contributions

I.-K.S. was the principal investigator of the research and wrote the original draft of the manuscript. I.-K.S and S.A. contributed towards the formal analysis. I.-K.S., S.A. and W.I.D.R. conceptualised the study, while S.A. wrote the programme for image quality analysis. S.A. and W.I.D.R. supervised and contributed towards the final drafting of the manuscript. I.-K.S., S.A. and W.I.D.R. approved the final manuscript for publication.

Funding information

Funding was received from the joint collaboration project Radiation Protection Improvement in South Africa (RAPISA), the Central University of Technology (CUT) Free State, the National Research Foundation (NRF, grant no.: 121950) and the Department of Higher Education and Training (DHET).

Data availability

The data that support the findings of this study are available from the corresponding author, I.-K.S., upon request.

Disclaimer

The views and opinions expressed in this article are those of the authors and are the product of professional research. They do not necessarily reflect the official policy or position of any affiliated institution, funder, agency or that of the publisher. The authors are responsible for this article's results, findings and content.

Acknowledgements

The authors acknowledge Dr Belinda van der Merwe for being a critical reader of the article, Mrs Maryn Viljoen, biostatistician, for assistance with the statistical analysis of data and Dr Daleen Struwig, medical writer/editor, for technical and editorial preparation of the article.

References

  • 1. International Commission on Radiological Protection (ICRP). ICRP publication 103. The 2007 recommendations of the International Commission on Radiological Protection. Ann ICRP [serial online]. 2013 [cited 2024 Jun 03];53:1689-1699. Available from: https://journals.sagepub.com/doi/pdf/10.1177/ANIB_37_2-4
    » https://journals.sagepub.com/doi/pdf/10.1177/ANIB_37_2-4
  • 2. Ekpo EU, Adejoh T, Akwo JD, et al. Diagnostic reference levels for common computed tomography (CT) examinations: Results from the first Nigerian nationwide dose survey. J Radiol Prot. 2018;38(2):525-235. https://doi.org/10.1088/1361-6498/aaaaf8
    » https://doi.org/10.1088/1361-6498/aaaaf8
  • 3. Pema D, Kritsaneepaiboon S. Radiation dose from computed tomography scanning in patients at songklanagarind hospital: Diagnostic reference levels. J Health Sci Med Res. 2020;38(2):135-143. https://doi.org/10.31584/jhsmr.2020732
    » https://doi.org/10.31584/jhsmr.2020732
  • 4. Asadinezhad M, Toossi MTB, Nouri M. Diagnostic reference levels for computed tomography examinations in Iran: A nationwide radiation dose survey. Iran J Med Phys [serial online]. 2019 [cited 2024 Jun 03];16(1):19-26. Available form: https://ijmp.mums.ac.ir/article_11500_33ed7c2911ea710161cf0ff9be551351.pdf
    » https://ijmp.mums.ac.ir/article_11500_33ed7c2911ea710161cf0ff9be551351.pdf
  • 5. Nam S, Park H, Kwon S, et al. Updated national diagnostic reference levels and achievable doses for CT protocols: A national survey of Korean hospitals. Tomography. 2022;8(5):2450-2459. https://doi.org/10.3390/tomography8050203
    » https://doi.org/10.3390/tomography8050203
  • 6. Kumsa MJ, Nguse TM, Ambessa HB, Gele TT, Fantaye WG, Dellie ST. Establishment of local diagnostic reference levels for common adult CT examinations: A multicenter survey in Addis Ababa. BMC Med Imaging. 2023;23(1):6. https://doi.org/10.1186/s12880-023-00963-1
    » https://doi.org/10.1186/s12880-023-00963-1
  • 7. Erem G, Ameda F, Otike C, et al. Adult computed tomography examinations in Uganda: Towards determining the national diagnostic reference levels. BMC Med Imaging. 2022;22(1):112. https://doi.org/10.1186/s12880-022-00838-x
    » https://doi.org/10.1186/s12880-022-00838-x
  • 8. Zalokar N, Žager Marciuš V, Mekiš N. Establishment of national diagnostic reference levels for radiotherapy computed tomography simulation procedures in Slovenia. Eur J Radiol. 2020;127:108979. https://doi.org/10.1016/j.ejrad.2020.108979
    » https://doi.org/10.1016/j.ejrad.2020.108979
  • 9. American Association of Physicists in Medicine (AAPM). Size-specific dose estimates (SSDE) in pediatric and adult body ct examinations [homepage on the Internte]. AAPM Report No. 204. College Park, MD: AAMP; 2011 [cited 2024 Jun 03]. Available from: https://www.aapm.org/pubs/reports/rpt_204.pdf
    » https://www.aapm.org/pubs/reports/rpt_204.pdf
  • 10. Mehdipour A, Parsi M. Patient dose survey based on size-specific dose estimate and acceptable quality dose in chest and abdomen/pelvis CT examinations. Radiat Prot Dosimetry. 2019;185(2):176-182. https://doi.org/10.1093/rpd/ncy288
    » https://doi.org/10.1093/rpd/ncy288
  • 11. Moghadam N, Lecomte R, Mercure S, Rehani MM, Nassiri MA. Simplified size adjusted dose reference levels for adult CT examinations: A regional study. Eur J Radiol. 2021;142:109861. https://doi.org/10.1016/j.ejrad.2021.109861
    » https://doi.org/10.1016/j.ejrad.2021.109861
  • 12. Beeres M, Römer M, Bodelle B, et al. Chest-abdomen-pelvis CT for staging in cancer patients: Dose effectiveness and image quality using automated attenuation-based tube potential selection. Cancer Imaging. 2014;14(1):28. https://doi.org/10.1186/s40644-014-0028-7
    » https://doi.org/10.1186/s40644-014-0028-7
  • 13. Xu J, Wang X, Yang P, Luo K, He X. Size-specific dose estimates of radiation based on body weight and body mass index for chest and abdomen-pelvic CTs. Biomed Res Int. 2020;2020:6046501. https://doi.org/10.1155/2020/6046501
    » https://doi.org/10.1155/2020/6046501
  • 14. Vañó E, Miller DL, Martin CJ, et al. ICRP Publication 135: Diagnostic reference levels in medical imaging. Ann ICRP. 2017;46(1):1-144. https://doi.org/10.1177/0146645317717209
    » https://doi.org/10.1177/0146645317717209
  • 15. WG-28 Physics. Digital Imaging and Communications in Medicine (DICOM) Supplement 191: Patient Radiation Dose Structured Report (P-RDSR). 2017 [cited 2024 Jun 03]. Available from: https://www.dicomstandard.org/docs/librariesprovider2/dicomdocuments/news/ftsup/docs/sups/sup191.pdf?sfvrsn=4cccbffa_2
    » https://www.dicomstandard.org/docs/librariesprovider2/dicomdocuments/news/ftsup/docs/sups/sup191.pdf?sfvrsn=4cccbffa_2
  • 16. Botwe BO, Schandorf C, Inkoom S, Faanu A, Rolstadaas L, Goa PE. National indication-based diagnostic reference level values in computed tomography: Preliminary results from Ghana. Phys Med. 2021;84:274-284. https://doi.org/10.1016/j.ejmp.2021.03.012
    » https://doi.org/10.1016/j.ejmp.2021.03.012
  • 17. Brat H, Zanca F, Montandon S, et al. Local clinical diagnostic reference levels for chest and abdomen CT examinations in adults as a function of body mass index and clinical indication: A prospective multicenter study. Eur Radiol. 2019;29(12):6794-6804. https://doi.org/10.1007/s00330-019-06257-x
    » https://doi.org/10.1007/s00330-019-06257-x
  • 18. Bos D, Yu S, Luong J, et al. Diagnostic reference levels and median doses for common clinical indications of CT: Findings from an international registry. Eur Radiol. 2022;32(3):1971-1982. https://doi.org/10.1007/s00330-021-08266-1
    » https://doi.org/10.1007/s00330-021-08266-1
  • 19. Habib Geryes B, Hornbeck A, Jarrige V, Pierrat N, Ducou Le Pointe H, Dreuil S. Patient dose evaluation in computed tomography: A French national study based on clinical indications. Phys Med. 2019;61:18-27. https://doi.org/10.1016/j.ejmp.2019.04.004
    » https://doi.org/10.1016/j.ejmp.2019.04.004
  • 20. Dao BT, Nguyen TV, Pham HH, Nguyen HQ. Phase recognition in contrast-enhanced CT scans based on deep learning and random sampling. Med Phys. 2022;49(7):4518-4528. https://doi.org/10.1002/mp.15551
    » https://doi.org/10.1002/mp.15551
  • 21. Vasanawala SS, Desser T, Jeffrey RB. Value of delayed imaging in MDCT of the abdomen and pelvis. Am J Roentgenol. 2006;187(1):154-163. https://doi.org/10.2214/AJR.05.0148
    » https://doi.org/10.2214/AJR.05.0148
  • 22. Chan MG, Cassidy FH, Andre MP, Chu P, Aganovic L. Delayed imaging in routine CT examinations of the abdomen and pelvis: Is it worth the additional cost of radiation and time? Am J Roentgenol. 2014;202(2):329-335. https://doi.org/10.2214/AJR.12.10468
    » https://doi.org/10.2214/AJR.12.10468
  • 23. Palorini F, Origgi D, Granata C, Matranga D, Salerno S. Adult exposures from MDCT including multiphase studies: First Italian nationwide survey. Eur Radiol. 2014;24(2):469-483. https://doi.org/10.1007/s00330-013-3031-7
    » https://doi.org/10.1007/s00330-013-3031-7
  • 24. Tsapaki V, Rehani M, Saini S. Radiation safety in abdominal computed tomography. Semin Ultrasound, CT MRI. 2010;31(1):29-38. https://doi.org/10.1053/j.sult.2009.09.004
    » https://doi.org/10.1053/j.sult.2009.09.004
  • 25. Varghese B, Kandanga I, Puthussery P, et al. Radiation dose metrics in multidetector computed tomography examinations: A multicentre retrospective study from seven tertiary care hospitals in Kerala, South India. Indian J Radiol Imaging. 2018;28(2):250-257. https://doi.org/10.4103/ijri.IJRI_394_17
    » https://doi.org/10.4103/ijri.IJRI_394_17
  • 26. South African Health Products Regulatory Authority (SAHPRA). Guideline for code of practice for users of medical x-ray equipment. SAHPGL-RDN-XR-02 Version 1. Pretoria: SAHPRA; 2022 [cited 2024 Jun 03]. Available from: https://www.sahpra.org.za/document/guideline-for-code-of-practice-for-users-of-medical-xray-equipment/
    » https://www.sahpra.org.za/document/guideline-for-code-of-practice-for-users-of-medical-xray-equipment/
  • 27. Trauernicht CJ, Pitcher RD. An audit of published South African diagnostic reference level data. J Radiol Prot. 2021;41(2):291-304. https://doi.org/10.1088/1361-6498/abfc98
    » https://doi.org/10.1088/1361-6498/abfc98
  • 28. Sebelego I, Acho S, Van der Merwe B, Rae WID. Size based dependence of patient dose metrics, and image quality metrics for clinical indicator-based imaging protocols in abdominal CT procedures. Radiography. 2023;29(6):961-974. https://doi.org/10.1016/j.radi.2023.07.011
    » https://doi.org/10.1016/j.radi.2023.07.011
  • 29. Public Health England. Doses in Computed Tomography (CT) examinations in the UK - 2011 Review. Oxfordshire: Crown; 2012.
  • 30. Foley SJ, McEntee MF, Rainford LA. Establishment of CT diagnostic reference levels in Ireland. Br J Radiol. 2012;85(1018):1390-1397. https://doi.org/10.1259/bjr/15839549
    » https://doi.org/10.1259/bjr/15839549
  • 31. O'Neill S, Kavanagh RG, Carey BW, Moore N, Maher M, O'Connor OJ. Using body mass index to estimate individualised patient radiation dose in abdominal computed tomography. Eur Radiol Exp. 2018;2(1):38. https://doi.org/10.1186/s41747-018-0070-5
    » https://doi.org/10.1186/s41747-018-0070-5
  • 32. Solomon JB, Li X, Samei E. Relating noise to image quality indicators in CT examinations with tube current modulation. Am J Roentgenol. 2013;200(3):592-600. https://doi.org/10.2214/AJR.12.8580
    » https://doi.org/10.2214/AJR.12.8580
  • 33. Chen JH, Jin EH, He W, Zhao LQ. Combining automatic tube current modulation with adaptive statistical iterative reconstruction for low-dose chest CT screening. PLoS One. 2014;9(4):e92414. https://doi.org/10.1371/journal.pone.0092414
    » https://doi.org/10.1371/journal.pone.0092414
  • 34. Yurt A, Özsoykal İ, Obuz F. Effects of the use of automatic tube current modulation on patient dose and image quality in computed tomography. Mol Imaging Radionucl Ther. 2019;28(3):96-103. https://doi.org/10.4274/mirt.galenos.2019.83723
    » https://doi.org/10.4274/mirt.galenos.2019.83723
  • 35. Martin CJ, Sookpeng S. Setting up computed tomography automatic tube current modulation systems. J Radiol Prot. 2016;36(3):R74-R95. https://doi.org/10.1088/0952-4746/36/3/R74
    » https://doi.org/10.1088/0952-4746/36/3/R74
  • 36. Ria F, Davis JT, Solomon JB, et al. Expanding the concept of diagnostic reference levels to noise and dose reference levels in CT. Am J Roentgenol. 2019;213(4):889-894. https://doi.org/10.2214/AJR.18.21030
    » https://doi.org/10.2214/AJR.18.21030
  • 37. Ahmad M, Liu X, Morani AC, et al. Oncology-specific radiation dose and image noise reference levels in adult abdominal-pelvic CT. Clin Imaging. 2023;93:52-59. https://doi.org/10.1016/j.clinimag.2022.10.016
    » https://doi.org/10.1016/j.clinimag.2022.10.016
  • 38. McLaughlin PD, Chawke L, Twomey M, et al. Body composition determinants of radiation dose during abdominopelvic CT. Insights Imaging 2018; 9:9-16. https://doi.org/10.1007/s13244-017-0577-y
    » https://doi.org/10.1007/s13244-017-0577-y
  • 39. Rawashdeh M, McEntee MF, Zaitoun M, et al. Knowledge and practice of computed tomography exposure parameters amongst radiographers in Jordan. Comput Biol Med. 2018;102:132-137. https://doi.org/10.1016/j.compbiomed.2018.09.020
    » https://doi.org/10.1016/j.compbiomed.2018.09.020
  • 40. Choudhary N, Bhupendra RS, Arvind S, Arun SO, Narinder PS, Sanjeev K. Patients dose estimation in CT examinations using size specific dose estimates. Radiat Prot Dosimetry. 2019;184(2):256-262. https://doi.org/10.1093/rpd/ncy207
    » https://doi.org/10.1093/rpd/ncy207

Correspondence:
Ida-Keshia Sebelego

Received: 13 June 2024
Accepted: 26 Aug. 2024
Published: 31 Oct. 2024

Correspondence

isebelego@cut.ac.za

Publication Dates

  • Publication in this collection
    03 Feb 2025
  • Date of issue

History

  • Received
    13 June 2024
  • Accepted
    26 Aug 2024
location_on
Radiological Society of South Africa AOSIS Publishing, Postnet Suite #110, Private Bag X19, Durbanville, Cape Town, Western Cape Province, ZA, 7551, Tel: +27 21 975 2602, Tel: 086 1000 381 - E-mail: publishing@aosis.co.za
rss_feed Stay informed of issues for this journal through your RSS reader
Go to top