CORRECTION
Addendum: Reliability of nurse-administered infant hearing screening using otoacoustic emissions
Mukovhe Phanguphangu; Andrew J. Ross
Department of Family Medicine, College of Health Sciences, University of KwaZulu-Natal, Durban, South Africa
Correspondence
References
-
In the published article Phanguphangu, M., & Ross, A.J. (2025). Reliability of nurse-administered infant hearing screening using otoacoustic emissions. South African Journal of Communication Disorders, 72(1), a1092. https://doi.org/10.4102/sajcd.v72i1.1092,
Updated referral rate (26% vs. 23%)
The corrected referral rate uses the appropriate denominator (number of ears screened). This adjustment does not change the interpretation of screening outcomes or the reliability estimates. The study's discussion and conclusions remain supported, as the primary objective was to evaluate inter-tester reliability, not screening yield.
Updated unit of analysis (infants vs. ears)
Terminology in the published version alternates between 'infants', 'babies' and 'ears'. The consistent unit of analysis for this study is ears, as inter-rater reliability (Kappa statistics) was conducted at the ear level. The following sections should therefore use the term 'ears' for clarity and consistency:
Sections requiring correction:
1. Research methods and design (study design and statistical analysis): Any reference to reliability being calculated per 'infant' should be amended to specify that agreement was calculated per 'ear'.
2. Results (reliability outcomes): All agreement, referral and screening outcomes should refer to 'ears screened' rather than 'infants screened'.
3. Table 1: The denominator and all outcome descriptors should reflect ears as the analytic unit.
Sections not requiring substantive revision:
1. Background and rationale sections that refer generally to 'infants' as the target population may remain unchanged, as these describe the population screened rather than the analytic unit.
2. The discussion and conclusions remain valid because the reliability estimates were always calculated correctly at the ear level; the issue was terminological clarification rather than analytic error.
Updated statistical reporting
Adding 95% confidence intervals (CI) and p-values enhances transparency and aligns the article with best-practice statistical conventions. The magnitude and interpretation of agreement remain unchanged: Audiologist vs Nurse 1, k = 0.80 (95% CI: 0.62-0.98; p < 0.001); Audiologist vs Nurse 2, k = 0.89 (95% CI: 0.69-1.00; p < 0.001); overall agreement, k = 0.81 (95% CI: 0.64-0.98; p < 0.001). These values continue to indicate substantial to near-perfect agreement and do not alter the study's conclusions.
Table 1 has been updated to include corrected and additional data. These changes align with statistical reporting guidelines, which recommend presenting the Kappa statistic (Cohen's Kappa) with the 95% CI to accurately reflect the distribution of the effect estimate.
Overall scientific integrity
The statistical analyses were conducted correctly at the ear level in the original submission. The addendum, therefore, clarifies terminology to align the written description with the analysis. Thus, none of the updates:
changes the direction or magnitude of the reliability findings
alters the interpretation of agreement levels
modifies the study's conclusions
affects the validity of the recommendation that nurse-administered otoacoustic emissions screening demonstrates strong reliability when appropriately trained and supervised.
References
Phanguphangu, M., & Ross, A.J. (2025). Reliability of nurse-administered infant hearing screening using otoacoustic emissions.
South African Journal of Communication Disorders, 72(1), a1092.
https://doi.org/10.4102/sajcd.v72i1.1092
Correspondence:
Mukovhe Phanguphangu
Published: 15 Apr. 2026
Publication Dates
-
Publication in this collection
05 May 2026
-
Date of issue