Abstract
The aim was to develop and validate a coherent and sustainable reverse-phase high-performance liquid chromatographic method for the quantification of Ethosuximide in oral solution. The partitioning was achieved using a stationary phase Shimpak solar C18 column (4.6 × 250 mm, 5 μm) with the mobile phase (60% acetonitrile and 40% water), flow rate (1 mL/min) and wavelength (210 nm). The developed method was validated as per ICH Q2R1 guidelines. The optimized method was evaluated for greenness and whiteness using analytical tools: AES, AGREE, MoGAPI, BAGI, RAPI, RGB fast models and compared to the reported methods. Ethosuximide was eluted at 3.445 min. The method was linear at 10-70 μg/mL with r2 = 0.9995. All the validation parameters were within the accepted values. The projected HPLC method met the requirements for effectiveness, greenness, efficiency, analytical performance, and productivity in terms of green and white analytical chemistry principles. Moreover, the optimized and validated analytical method was significantly better compared to the two reported methods on the basis of analytical performance, eco-friendliness and throughput. The results of the developed analytical method are reproducible, efficient with excellent chromatographic properties and effectively employed for regular quality assessment of Ethosuximide in oral solution.
Green and White analytical chemistry; Ethosuximide; Validation; Sustainability; ICH Q2R1
INTRODUCTION
Epilepsy is associated with serious physical, mental health and cognitive comorbidities. Anti-epileptic drugs (AEDs) are used as single or combination drug therapy in the management of several forms of epilepsy.1Figure 1) is the leading choice for the handling of petit mal and myoclonic seizures due to its efficiency and a typically acceptable side effect profile.11 2
RP-HPLC is a well-developed and widely used analytical technique for separating and analysing various compounds, including pharmaceuticals, biomolecules and environmental samples.345
Several analytical methods were reported for the determination of Ethosuximide either alone or with other antiepileptics such as UV-spectrophotometry, HPLC, LC/MS, and GC/MS.6-9 The literature study revealed that only two RP-HPLC methods have been published and are mainly concentrated on forced degradation, profiling of related substances, and impurities.6-7 These reports possess some drawbacks such as, the use of more than two solvents, higher retention time, and higher consumption of solvents. Therefore, a simple, quick, efficient, eco-friendly and validated RP-HPLC method for estimating Ethosuximide in pure and oral solution to improve the drawbacks of previously published methods. The Greenness and Whiteness of the optimized analytical method were evaluated using AES, AGREE, MoGAPI, BAGI, RAPI and RGB models. Further, comparison of proposed and two reported analytical methods was also carried out in these lines. To the best of our knowledge, this is the first report to combine green and white analytical assessment tools for the quantification of Ethosuximide, offering an environmentally responsible and practically feasible alternative to previously published methods.
EXPERIMENTALInstrumentation
The study was carried out using a Shimadzu prominence LC-20 AD with PDA connected to Lab solution software, spectrophotometer (LAB India- 3000+), ultrasonic water bath (Power Sonic 405), digital weighing balance (Shimadzu ATX224R) and micro-pipette (10-1000 μL from Eppendorf) based on the requirements of analytical work.
Chemicals and dosage form
Ethosuximide was obtained as a gift sample for research purposes from industry. Absenz (Ethosuximide oral solution IP) was purchased from Sun Pharma Laboratories Pvt. Ltd. High performance liquid chromatography-grade water and acetonitrile (purchased from Thermo Fisher Scientific, Mumbai, India), methanol (purchased from S D Fine-Chem Limited, Mumbai, India) were used throughout this study.
Standard solution preparation
Ethosuximide (10 mg) was weighed and transferred into a 10 mL volumetric flask, dissolved using diluent and sonicated for 10 min, and the volume was made up to the mark with 60% acetonitrile: 40% water (1000 μg/mL). The sub-stock was prepared by pipetting 1 mL of the above solution into a 10 mL volumetric flask with diluent. The final concentration of 30 μg/mL was prepared by taking 3 mL from the sub-stock into a 10 mL volumetric flask, sonicated for 10 min and made up to the mark using diluent.
Sample solution preparation
The Absenz oral solution contains 250 mg of Ethosuximide in 5 mL. For the preparation of stock solution, 2 mL of Ethosuximide oral solution was pipetted into a 10 mL volumetric flask, and diluted with mobile phase, sonicated and filtered with a PES filter membrane media syringe. For sub-stock, 1 mL of the above stock solution was transferred to a 10 mL volumetric flask and made up with diluent and sonicated. Then, 3 mL of sub-stock solution was transferred to a 10 mL volumetric flask, diluted and made up to the mark with mobile phase to give a concentration of 30 μg/mL.
Method development
The selection of the mobile phase for Ethosuximide depended on its physical and chemical properties. Various solvents such as methanol, acetonitrile, water and glacial acetic acid were used with varying ratios to get a suitable, stable, mobile phase composition. To develop and optimize the chromatographic conditions, fine-tuning of flow rate and mobile phase composition was attempted.
Method validation
HPLC method validation involved finding and demonstrating that a method consistently produces expected, reliable and accurate results under specified conditions, while meeting all of the validation parameters outlined in the ICH Q2R1 guidelines including system suitability, specificity, linearity, instrumental precision, method precision, accuracy, robustness, LOQ, LOD and ruggedness.
Greenness and whiteness evaluation of the analytical method
Evaluating the greenness and whiteness of the analytical method was done by assessing the impact of the optimized analytical method on environmental, efficiency and practical performances. These evaluations are critical in modern analytical chemistry to ensure sustainability, safety and effectiveness. Greenness and Whiteness of the developed and reported analytical methods were evaluated using the following tools: Analytical Eco-Scale, AGREE, MoGAPI, BAGI, RAPI and RGB fast model. Analytical Eco-Scale is a semi-measurable tool used to assess the greenness of the analytical method in accordance with the twelve principles of green analytical chemistry. It begins with a perfect score of 100 points, which reflects an ideal green method. Penalty points are distributed based on parameters such as, type and amount of solvents/reagents used and the quantity consumed in proposed and reported methods, energy utilization, occupational hazard, waste production and treatment, and finally safety. The amount of waste generated was calculated by multiplying the run time against the flow rate of a single analysis. Penalty points for waste treatment were given based on the type of treatment adopted for waste management. An eco-scale score above 75 was considered as an excellent green method.1011 12
Blue Applicability Grade Index (BAGI) is considered a well-established tool for green metrics and predominantly concentrates on the practical side of white analytical chemistry. It is a star-shaped pictogram with ten main attributes and a score displayed in the center having a scale of 25-100, that illustrates the functionality and applicability of the optimized method. The inner portion of the pictogram (1 to 5) reflects either the number of steps for sample preparation or the stages for analytical determination, such as the kind of analysis, analytical apparatus, number of analytes analyzed concurrently, and sample preparation steps. The outer portion of the pictogram, from 6 to 10, included the total number of samples investigated per hour, the nature of the materials and reagents used in the analytical method, the level of automation, and the quantity of sample used. Each part of the star-shaped pictogram is assigned a distinct colour based on a score: for high compliance the score is 10 (dark blue), 7.5 for medium compliance (blue), 5.0 for poor compliance (light blue), and 2.5 for no compliance (white). The overall score for analytical method performance is: 100 for outstanding and 25 for the worst. At least 60 points are recommended for any analytical method to be considered as Green and White.13
The Red Analytical Performance Index tool is used to deliver multifaceted information about wide-ranging analytical parameters. This to ol considers ten parameters for score distribution, which includes intermediate precision, repeatability, reproducibility, recovery and matrix effect, LOQ, working range, trueness, ruggedness/robustness and selectivity. The scoring scale has five levels, which are as follows: a score of 0 is white, scores of 2.5, 5.0, and 7.5 are an intermediate colour, and 10 is dark red. Furthermore, a score of zero means the result is unacceptable. The total of all the scores falls which should be between 0 and 100, is displayed in the central portion of the pictogram.141415
RESULTS AND DISCUSSION
Chromatographic trails
The total of eight chromatographic trails were carried out. In the first five trails, various solvent combinations consisting of methanol: water (40:60 v/v), methanol: water (60:40 v/v), methanol: water (65:35 v/v), methanol: water (70:30 v/v) and methanol: water with 1% v/v acetic acid (65:35 v/v) were done. All data from these trails were rejected based on higher retention time of Ethosuximide, reproducibility, appearance of extra peaks, interference with the blank and long stabilization time with high baseline noise. Moreover, using acetonitrile: water in the ratio of 40:60 and 50:50 v/v were also not considered due to high retention time and reproducibility. However, trial 8 demonstrated good chromatographic conditions for the selective and specific analysis of Ethosuximide in formulation and demonstrated excellent adherence to the guidelines. Therefore, trial 8 was chosen for the further experiments.
Optimized analytical method
The RP-HPLC method was developed using the Shimpak Solar C18 column (4.6 × 250 mm, 5 um) with 60 % ACN and 40 % water; 1 mL/min flow rate and 210 nm analytical wavelength. The Ethosuximide peak was observed at 3.445 min having a tailing factor of 1.28 and theoretical plates of 9727 (Figure 2B).
Method validation - System suitability
Six injections of standard Ethosuximide (30 μg/mL) exhibited consistent results (Table 1). The percentage RSD values for all the critical chromatographic parameters are within the acceptable range (% RSD < 2.0).
Specificity
The chromatogram of the standard and sample showed identical retention times which is an indication of the absence of interference (Figure 2A-C). The peak purity index of Ethosuximide oral solution was 0.999 (Figure 2D). This further confirmed that the components of the oral solution did not interfere with Ethosuximide peak.
Linearity
The standard calibration curve was plotted as mean peak area vs the different concentrations of Ethosuximide in the range of 10-70 μg/mL with correlation co-efficient (r2Figure 3) showing the results were within the acceptable standards.
Accuracy
The accuracy was confirmed by spiking the standard solution at three levels (80%, 100%, 120%) in the diluted sample solution. The results of % recovery (Table 2) for the three levels are within the acceptable range (98%-102%).
Instrument precision and method precision
Repeatability was confirmed by consistent results for all the investigated chromatographical parameters with six injections. Intermediate precision was established by evaluation of intra- and inter-day experiments. Similarly, the method precision was measured by performing an assay on both the intra- and inter-day experiments.
All the critical chromatographic parameters were within the acceptance criteria. Thus, all of the precision data revealed that the proposed analytical method is reproducible (Table 3).
Robustness
The robustness was done by introducing minor variations in the experimental settings such as a change in the flow rate (± 0.1 mL/min), mobile phase composition (± 5 mL), and wavelength (± 2 nm). The % RSD for all three factors were inside the acceptance limit of less than 2 (Table 4).
Ruggedness
This test required two different analysts to carry out the experiments which involved six replicate injections of Ethosuximide into the HPLC. The ruggedness results (Table 5) showed that the % RSD for all the parameters, performed by both analysts, are within the acceptable range of less than 2.
Limit of detection (LOD) and limit of quantification (LOQ)
The LOD and LOQ were determined to be 1.74 μg/mL and 5.81 μg/mL, respectively, based on the signal to noise ratio.
Statistical comparison of the proposed and reported method
A statistical comparison of the assay results from the proposed and reported methods was carried out using the Student t-test (One tail) and the variance ratio F test (Table 6). The statistical comparison reveals that there was a significant difference, using the Student t-test (One tail), between the proposed method and the reported method. However, a comparison using the variance ratio F test, exhibited an insignificant difference.
Green evaluation of the optimized analytical method
Analytical Eco-Scale
Analytical Eco-Scale data for the developed and reported methods are presented in Tables 7 and 8. The penalty points for reagents/solvents were calculated by multiplying the number of hazard pictograms listed in the material safety data sheet against the score assigned to the signal word (warning =1, danger =2). Here, acetonitrile has two pictograms and the signal word 'danger', hence the proposed method was assigned four penalty points in the reagent utilization section. The energy consumption for a single analysis is considered to be the total power consumption. The developed method consumed less than 1.5 kWh per sample, so 1 penalty point was assigned. For waste production, the penalty points are calculated by multiplying run time multiplied against the flow rate. The proposed method received 3 penalty points, whereas reported methods 1 and 2 received 5 penalty points each. The proposed method involved an oxidative degradation of acetonitrile into non-toxic/safer by-products, hence it received 2 penalty points. However, no waste treatment was specified in the reported methods, therefore each got 3 penalty points. The AEC scores collectively demonstrated that the proposed method had a higher eco-score and better greenness than the reported methods.
AGREE model
The optimized approach has an AGREE value of 0.58, which is greater than the other two reported methods (Table 8). In the developed method, segments 2, 4, 6, and 9 were dark green, indicating a small sample size, few steps in sample preparation (such as sonication and filtering), and no derivatization process that consumed less than 1.5 kWh of energy, respectively. Red for the 3rd segment represented an analytical device was offline. The 8th segment has a light green shade because a single analyte was used per run. Segments 11 and 12 were orange, which was due to the use ofhazardous reagents like acetonitrile. Yellow appeared for the 1st, 5th, 7th, 10th segments representing the sampling method (off-line), manual handling of the instrument, waste generation (5 mL), and use of some bio-based reagent.
Modified GAPI
Modified GAPI provided the quantification of greenness of an analytical method using a scoring system for each segment as well as for the overall system (Table 8). In the proposed method, greener segments were due to the lack of sample preservation, transportation, storage and additional treatment. Further, the method used less than 10 mL of solvent per run and a hermetic sealing process. The yellow segments were related to simple procedures (filtration, sonication), selected solvent (acetonitrile) and a health-safety NFPA hazard score of 2 or 3. In addition, energy of less than 1.5 kWh was used and less than 10 mL of waste was generated. The remaining red segments were due to offline sample collection, use of non-green solvent (ACN). The developed method displayed a total score of 79 which is considered as green compared to the two reported methods having scores of 70 and 66, respectively.
BAGI model
The proposed method had a comparatively higher Blue Applicability Grade Index (BAGI) score of 77.5 than reported methods 1 and 2 which had scores of 67.5 and 70.0, respectively (Table 8). The BAGI pictogram has ten parts or sections. Sections 1, 5, 6, 7, 8, and 10 are dark blue, revealing that the proposed method was very much compliant for: quantitative analysis; on-site sample preparation; rate of analysis of more than ten samples per hour; use of common commercially available reagents such as ACN; no pre-concentration required and small sample size. The 3rd section had a moderate blue shade since the use of the HPLC-UV instrument implies reasonable compliance. Light blue for the 9th section implies semi-automatic instrumentation. White for sections 2 and 4 was due to single compound analysis and no simultaneous sample preparation (Table 8).
RAPI Tool
The Red Analytical Performance Index (RAPI) focuses primarily on the evaluation of the analytical performance by considering the ten parameters as per the ICH Q2R1 guidelines. The 2nd, 3rd, and 8th sections were dark red for the proposed method since the % RSD for intermediate precision and reproducibility was less than 10 % and standard linearity was greater than the 0.99 correlation coefficient (Table 8). Sections 1, 4, and 9 were red. Section 1 represented the repeatability which was less than 10% RSD; section 4 showed the trueness (<1.0), and section 9 showed the robustness which consisted of three factors (change in the mobile phase ratio, flow rate, and wavelength). Sections 5, 6 and 7 were a lighter shade of red. Section 10 was white which demonstrated that there was no selectivity. Overall, the proposed method was analytically more efficient (RAPI score 67.5) than the two reported methods (RAPI scores of 57.5 and 50.0, respectively) (Table 8).
RGB fast model
The proposed method had a score of 40.1 for analytical efficiency (Red). The ChlorTox scale, which is related to greenness, was 58.4. The last section, which was blue, exhibited economic efficiency with a score of 60.9 due to the sample throughput. The whiteness score for the proposed method was evaluated by considering all three aspects and was determined to be 52. The proposed analytical method performed significantly better in adhering to white principles than the two reported methods (white scores of 51 and 44, respectively) (Figure 4).
CONCLUSION
An improved method for estimating Ethosuximide in bulk and oral solution was developed and validated to meet all the pre-determined acceptance conditions of the ICH Q2R1 guidelines. The developed method is rapid, simple, precise and environmentally friendly, with good analytical performance in terms of green and white metric principles. Therefore, this method can be used for regular quality control analysis of Ethosuximide in pharmaceutical formulations.
SUPPLEMENTARY INFORMATIONThis manuscript does not have Supplementary Information. All the data is presented in the article.
AUTHOR CONTRIBUTIONSConcept: V.P.V.; Design: V.P.V., R.R., D.P.; Supervision: V.P.V.; Resources: V.P.V.; Materials: D.P., S.R.P.; Data collection and/or processing: R.R, S.R.P.; Analysis and/or interpretation: V.P.V., R.R.; Literature search: R.R., V.P.V.; Writing: R.R.; Critical reviews: V.P.V., S.R.P., D.P
DECLARATION OF COMPETING/FINANCIAL INTERESTS
The authors declared no conflict of interests.
DECLARATION OF GENERATIVE AI AND AI-ASSISTED TECHNOLOGIES
During the preparation of the manuscript, AI-assisted tool Grammarly was used for grammar checks or revisions.
AUTHORS ORCIDSReshma Rajanna: https://orcid.org/0009-0005-7483-3392
Veeresh Prabhakar Veerapur: https://orcid.org/0000-0002-6016-4766
Devireddy Prashanthi: https://orcid.org/0009-0001-6699-1447
Santhemavathuru Ramachandraiah Premkumara: https://orcid.org/0000-0002-8561-5885
ACKNOWLEDGEMENTS
The authors are thankful to the Management and Principal of Sree Siddaganga College of Pharmacy, Tumkur, for providing the required infrastructure to carry out the research work. The authors also extend our gratitude to Dr. Nirankar Nath Mishra, Professor at the Centre of Applied Research and Nanotechnology (CARN), Sree Siddaganga Institute of Technology, Tumkur, for providing the Millipore water for HPLC work.
REFERENCES
-
1. Gören MZ, Onat F. Ethosuximide from bench to beside. CNS Drug Rev. 2007;13(2):224-239. https://doi.org/10.1111/j.1527-3458.2007.00009.x.
» https://doi.org/10.1111/j.1527-3458.2007.00009.x -
2. Vaibhav NK, Sagar SY, Vrushali SD, Gita RG. Development and validation for UV spectrophotometric method of estimation of ethosuximide in bulk pharmaceutical dosage forms. World J. Pharm. Sci. 2020;9(1):868-873. https://doi.org/10.20959/wjpps20201-15197.
» https://doi.org/10.20959/wjpps20201-15197 -
3. Görög S. The changing face of pharmaceutical analysis. Trends Analyt Chem. 2007;26(1):12-17. https://doi.org/10.1016/j.trac.2006.07.011.
» https://doi.org/10.1016/j.trac.2006.07.011 -
4. Armenta S, Garrigues S, de la Guardia. Green analytical chemistry. Trends Analyt Chem. 2008;27(6):497-511. https://doi.org/10.1016/j.trac.2008.05.003.
» https://doi.org/10.1016/j.trac.2008.05.003 -
5. Sajid M, Plotka-Wasylka J. Green analytical chemistry metrics: A review. Talanta. 2022;238(2):123046. https://doi.org/10.1016/j.talanta.2021.123046.
» https://doi.org/10.1016/j.talanta.2021.123046 -
6. El-Shabrawy Y, Walash M, El-Enany N, El-Shaheny R. Investigation of ethosuximide stability under certain ICH recommended stress conditions using a validated stability-indicating HPLC method. Anal Methods. 2018;10(12):1452-1458. https://doi.org/10.1039/C7AY02740K.
» https://doi.org/10.1039/C7AY02740K - 7. Karthikeyan G, Senthilnathan B, Pavitra S, Parameshwari M, Jayashree S. Analytical method development and method validation for the analysis of related substances in soft gelatin capsules of ethosuximide using RP-HPLC. Inventi Rapid: Pharm. Anal. Qual. Assur. 2023;4(3):1-10.
-
8. Bhatt M, Shah S, Shivprakash. Development of a high-throughput method for the determination of ethosuximide in human plasma by liquid chromatography mass spectrometry. J Chromatogr B Analyt Technol Biomed Life Sci. 2010;878(19):1605-1610. https://doi.org/10.1016/j.jchromb.2010.03.019.
» https://doi.org/10.1016/j.jchromb.2010.03.019 -
9. Khoschsorur GA, Frühwirth F, Halwachs-Baumann G. Simple and rapid HPLC method for simultaneous determination of multiple antiepileptic drugs in human serum. Chromatographia. 2001;54(5-6):345-349. https://doi.org/10.1007/BF02492681.
» https://doi.org/10.1007/BF02492681 -
10. Gahiszka A, Migaszewski ZM, Konieczka P, Namiesnik J. Analytical Eco-Scale for assessing the greenness of analytical procedures. TrAC. Trends Analyt Chem. 2012;37:61-72. https://doi.org/10.1016/j.trac.2012.03.013.
» https://doi.org/10.1016/j.trac.2012.03.013 -
11. Pena-Pereira F, Wojnowski W, Tobiszewski M. AGREE-Analytical GREEnness metric approach and software. Anal Chem. 2020;92(14):10076-10082. https://doi.org/10.1021/acs.analchem.0c01887.
» https://doi.org/10.1021/acs.analchem.0c01887 -
12. Mansour FR, Plotka-Wasylka J, Locatelli M. Modified GAPI (MoGAPI) tool and software for the assessment of method greenness: case studies and applications. Analytica. 2024;5(3):451-457. https://doi.org/10.3390/analytica5030030.
» https://doi.org/10.3390/analytica5030030 -
13. Manousi N, Wojnowski W, Plotka-Wasylka J, Samanidou V. Blue applicability grade index (BAGI) and software: a new tool for the evaluation of method practicality. Green Chem. 2023;25(19):7598-7604. https://doi.org/10.1039/D3GC02347H.
» https://doi.org/10.1039/D3GC02347H -
14. Nowak PM, Wietecha-Posluszny R, Pawliszyn J. White analytical chemistry: an approach to reconcile the principles of green analytical chemistry and functionality. Trends Analyt Chem. 2021;138:116223. https://doi.org/10.1016/j.trac.2021.116223.
» https://doi.org/10.1016/j.trac.2021.116223 -
15. Nowak PM, Arduini F. RGBfast- A user-friendly version of the Red-Green-Blue model for assessing greenness and whiteness of analytical methods. Green Anal Chem. 2024;10:100120. https://doi.org/10.1016/j.greeac.2024.100120.
» https://doi.org/10.1016/j.greeac.2024.100120
Revised: 16 June 2026
Accepted: 12 July 2026