Open-access Simultaneous multi-element electrothermal atomic absorption determination using a low resolution CCD spectrometer and continuum light source: The concept and methodology

A low resolution CCD spectrometer with continuum light source and fast-heated graphite tube atomizer was employed for simultaneous multi-element determination by electrothermal atomic absorption spectrometry (SMET AAS). The sample vaporization pulse was monitored by fast scanning of vapour spectra within the 190-410 nm wavelength range; absorption was measured at the CCD pixels corresponding to atomic resonance lines; function absorbance vs. concentration of atomic vapour was automatically linearized, and the modified signals integrated. The setup consisted of a D2 or Xe arc lamp, a spectral instrument with a half-width of transmittance profile 120 pm, a linear CCD array attached to a PC and a tube atomizer furnished with a carbon fibre collector. In the experiments simultaneous determination of 18 elements was performed in the mixed solutions at the mg L-1 to L-1 level, within 4-4.5 orders of magnitude linear concentration range. About 1-2 min was needed for the measurement and calculation. Limits of detection (LOD) for individual elements were 1.5-2 orders of magnitude higher than those in the single element ET AAS, but similar or below those in flame AAS. Further reduction of LODs and correction of possible spectral and chemical interferences are associated with optimization of the light source and atomization programme and modification of the calculation algorithm.

Electrothermal atomic absorption spectrometry; simultaneous multi-element determination; CCD spectrometer; fast-heated graphite tube atomizer


RESEARCH ARTICLE

Simultaneous multi-element electrothermal atomic absorption determination using a low resolution CCD spectrometer and continuum light source: The concept and methodology

Dimitri A. Katskov*; G. Eunice Khanye

Department of Chemistry and Physics, Tshwane University of Technology, Pretoria, 0001 South Africa

ABSTRACT

A low resolution CCD spectrometer with continuum light source and fast-heated graphite tube atomizer was employed for simultaneous multi-element determination by electrothermal atomic absorption spectrometry (SMET AAS). The sample vaporization pulse was monitored by fast scanning of vapour spectra within the 190-410 nm wavelength range; absorption was measured at the CCD pixels corresponding to atomic resonance lines; function absorbance vs. concentration of atomic vapour was automatically linearized, and the modified signals integrated. The setup consisted of a D2 or Xe arc lamp, a spectral instrument with a half-width of transmittance profile 120 pm, a linear CCD array attached to a PC and a tube atomizer furnished with a carbon fibre collector. In the experiments simultaneous determination of 18 elements was performed in the mixed solutions at the mg L-1 to L-1 level, within 4-4.5 orders of magnitude linear concentration range. About 1-2 min was needed for the measurement and calculation. Limits of detection (LOD) for individual elements were 1.5-2 orders of magnitude higher than those in the single element ET AAS, but similar or below those in flame AAS. Further reduction of LODs and correction of possible spectral and chemical interferences are associated with optimization of the light source and atomization programme and modification of the calculation algorithm.

Keywords: Electrothermal atomic absorption spectrometry, simultaneous multi-element determination, CCD spectrometer, fast-heated graphite tube atomizer

Full text available only in PDF format.

Acknowledgements

The authors would like to acknowledge the National Research Foundation, South Africa, and Tshwane University of Technology for financial support of this research.

References

  • 1 J.M. Harnly, J. Anal. At. Spectrom., 1999,14, 137-146.
  • 2 B. Welz, H. Becker-Ross, S. Florek and U. Heitmann, High-resolution Continuum Source AAS: The Better Way to do Atomic Absorption Spectrometry, Wiley-VCH, Weinheim, Germany, 2005.
  • 3 http://www.analytik-jena.de (accessed 1 March 2010).
    » http://www.analytik-jena.de
  • 4 U. Heitmann, H. Becker-Ross and D. Katskov, Spectrochim. Acta, Part B, 2006, 61, 351-360.
  • 5 B.V Lvov, Atomic Absorption Spectrochemical Analysis, Adam Hilger, Elsevier, London, UK, 1970.
  • 6 K.I. Tarasov, The Spectroscope, Adam Hilger, Bristol, UK, 1974.
  • 7 A.C.G. Mitchell and M.W. Zymansky, Resonance Radiation and Excited Atoms, Cambridge University Press, Cambridge, UK, 1961.
  • 8 B. Welz and M. Sperling, Atomic Absorption Spectrometry, 3rd edn., Wiley-VCH, Weinheim, Germany, 1999.
  • 9 http://www.OceanOptics.com (accessed 1 March 2010).
    » http://www.OceanOptics.com
  • 10 http://www.cortec.ru (accessed 1 March 2010).
    » http://www.cortec.ru
  • 11 RSA Provisional Patent Application 2008/06901.
  • 12 D.A. Katskov, J. Anal. At. Spectrom., 2005, 20, 220-226.
  • 13 D.A. Katskov, Yu. M. Sadagov and M. Banda, J. Anal. At. Spectrom., 2005, 20, 227-232.
  • 14 SIMAA 6000 Simultaneous Multi-element Atomic Absorption Spectrometer, Perkin-Elmer Bodenseewerk, Überlingen, Germany, 1995.
  • 15 http://en.wikipedia.org (accessed 1 March 2010).
    » http://en.wikipedia.org
  • 16 D.A. Katskov, A.M. Shtepan, I.L. Grinshtein and A.A. Pupyshev, Spectrochim. Acta, Part B, 1992, 47, 1023-1041.
  • 17 http://physics.nist.gov (accessed 1 March 2010).
    » http://physics.nist.gov
  • 18 J.G. Jackson, R.W. Fonseca and J.A. Holcombe, Spectrochim. Acta, Part B, 1995, 50, 1837-1846.
  • 19 G. Tyler, VARIAN Instruments at Work. Application Note. AA or ICU: Which Do You Choose? ICP-3, Varian Inc., Mulgrave, Victoria, Australia, 3 September 1991.
  • 20 P. Heitland, Spectro ICP Report No. ICP-32, Spectro Analytical Instruments, Kleve, Germany, 2001.


Accepted 3 May 2010

*To whom correspondence should be addressed. E-mail:

Correspondence

katskovda@tut.ac.za

Publication Dates

  • Publication in this collection
    16 July 2015
  • Date of issue

History

  • Received
  • Accepted
    03 May 2010
location_on
The South African Chemical Institute The South African Journal of Chemistry, University of the Witwatersrand, The Secretary - Laila Smith, School of Chemistry, Private Bag X3, Braamfontein, Johannesburg, ZA, 2050, Tel:+27 11 717 6705, 061 282 3477 - E-mail: editorsajchem@saci.co.za
rss_feed Acompañe los números de esta revista en su lector de RSS
Ir para arriba