Open-access Simultaneous determination of metals in coal with low-resolution continuum source atomic absorption spectrometer and filter furnace atomizer

The setup including low-resolution spectrometer with the charge-coupled device (CCD) detector, continuum radiation source and filter furnace (FF) atomizer was employed for direct simultaneous determination of Al, Fe, Mg, Cu and Mn in coal slurry. In the FF, sample vapour entered absorption volume by filtering through heated graphite. Absorption spectrum within 200-400 nm was repeatedly recorded during the atomization period with spectral resolution 0.3 nm. The output of the CCD elements was measured within each spectrum, atomic absorption at specific wavelengths measured and corrected with respect to the linearization algorithm, and integrated. Calibration was performed using carbon slurry impregnated by the analyte metals as well as with the analytes added to the slurries as multi-element solutions. The comparison showed preference of the second method, which provides for 60 % of measurements the results within 10 % deviation range from the certified reference data independent of concentration of the analyte. Low-resolution spectral instrument with fast CCD detection makes possible simultaneous detection of transient absorption signals for several elements. The use of continuum light source makes it possible to determine broad range of concentrations without slurry dilution.

Coal slurry; electrothermal atomic absorption spectrometry; simultaneous multi-element determination; filter furnace atomizer


RESEARCH ARTICLE

Simultaneous determination of metals in coal with low-resolution continuum source atomic absorption spectrometer and filter furnace atomizer

Gibson Jim; Dmitri A. Katskov*

Department of Chemistry and Physics, Tshwane University of Technology, Pretoria, 0001, South Africa

ABSTRACT

The setup including low-resolution spectrometer with the charge-coupled device (CCD) detector, continuum radiation source and filter furnace (FF) atomizer was employed for direct simultaneous determination of Al, Fe, Mg, Cu and Mn in coal slurry. In the FF, sample vapour entered absorption volume by filtering through heated graphite. Absorption spectrum within 200-400 nm was repeatedly recorded during the atomization period with spectral resolution 0.3 nm. The output of the CCD elements was measured within each spectrum, atomic absorption at specific wavelengths measured and corrected with respect to the linearization algorithm, and integrated. Calibration was performed using carbon slurry impregnated by the analyte metals as well as with the analytes added to the slurries as multi-element solutions. The comparison showed preference of the second method, which provides for 60 % of measurements the results within 10 % deviation range from the certified reference data independent of concentration of the analyte. Low-resolution spectral instrument with fast CCD detection makes possible simultaneous detection of transient absorption signals for several elements. The use of continuum light source makes it possible to determine broad range of concentrations without slurry dilution.

Keywords: Coal slurry, electrothermal atomic absorption spectrometry, simultaneous multi-element determination, filter furnace atomizer

Full text available only in pdf format.

Acknowledgements

The authors express their gratitude to National Research Foundation of South Africa and Tshwane University of Technology for the financial support of the research; and to Prof. Tittarelli for the advises rendered.

References

  • 1 F.E. Huggins, Int. J. Coal Geol, 2002, 50, 169-214.
  • 2 www.astm.org/Standards/D6349.htm (accessed 1 October 2010).
    » www.astm.org/Standards/D6349.htm
  • 3 I. Rodushkin, M.D.Axelsson and E. Burman, Talanta, 2000, 51, 743-759.
  • 4 L. Ebdon and J.R. Wilkinson, J. Anal. At. Spectrom., 1987, 2 325-328.
  • 5 M. Mujuru, R.I. McCrindle, B.M. Botha, P.P. Ndibewu, Fuel, 2009, 88, 719-724.
  • 6 B. Welz, M.G.R. Vale, M.M. Silva, H. Becker-Ross,M. Huang, S. Florek and U. Heitmann, Spectrochim. Acta part B, 2002, 57, 1043-1055.
  • 7 M.G.R. Vale, N. OleszczukandW.N.L. dos Santos, Appl. Spectros. Rev., 2006, 41, 377-400.
  • 8 M.M. Silva, M.G.R. Vale and E.B. Caramo, Talanta, 1999, 50, 1035-1043.
  • 9 B. Welz, H. Becker-Ross, S. Florek and U. Heitmann, High-Resolution Continuum Source AAS: the Better Way to do Atomic Absorption Spectrometry, Wiley-VCH, Weinheim, 2005.
  • 10 C. Bendicho and M.T.C. Loos-Vollebregt, J. Anal. At. Spectrom. 1991, 6, 353-374.
  • 11 http://lenntech.com (accessed 1 October 2010).
    » http://lenntech.com
  • 12 D.A. Katskov, Spectrochim. Acta Part B, 2007, 62, 897-917.
  • 13 D.A. Katskov and G.E. Khanye, S. Afr. J. Chem, 2010, 63, 45-57.
  • 14 D.A. Katskov, P.J.J.G. Marais, V.J. Katkovnik and P. Tittarelli, Spectrochim. Acta Part B, 1997, 52, 1377-1394.
  • 15 R.W.B. Pearse and A.G. Gaydon, The Identification of Molecular Spectra, Chapman & Hall, London, 1950 & 1976.
  • 16 D.A. Katskov, A.M. Shtepan, I.L. Grinshtein and A.A. Pupyshev, Spectrochim Acta Part B, 1992, 47, 1023-1041.

Received 18 October 2010
Revised 19 January 2011
Accepted 6 February 2011

Submitted by invitation to celebrate 2011 the 'International Year of Chemistry'.
* To whom correspondence should be addressed. E-mail:

Correspondence

katskovda@tut.ac.za

Publication Dates

  • Publication in this collection
    16 July 2015
  • Date of issue

History

  • Received
    18 Oct 2010
  • Reviewed
    19 Jan 2011
  • Accepted
    06 Feb 2011
location_on
The South African Chemical Institute The South African Journal of Chemistry, University of the Witwatersrand, The Secretary - Laila Smith, School of Chemistry, Private Bag X3, Braamfontein, Johannesburg, ZA, 2050, Tel:+27 11 717 6705, 061 282 3477 - E-mail: editorsajchem@saci.co.za
rss_feed Acompañe los números de esta revista en su lector de RSS
Ir para arriba