Open-access In-vitro comparison of bonding time and strength of adhesive pre-coated and standard metal orthodontic brackets

Abstract

OBJECTIVE: To compare the shear bond strength and bond time of 3M Unitek's APC (Adehesive Pre-Coated) Flash-Free™ system applied to metal brackets. Materials and Methods: An in vitro study was performed on 40 extracted sound human premolar teeth randomly divided into two groups (20 per group) bonded either with Adhesive Pre-Coated Flash-Free metal brackets or metal brackets coated manually with Transbond XT™ light-cure adhesive. Bonding time was measured using a stopwatch. Thermoycling was performed on the samples (500 cycles) to simulate the oral environment between baths of 5°C and 55°C distilled water. Debonding shear bond strength measurements were performed in an Instron universal testing machine. RESULTS: The APC Flash-Free group bonded in significantly (p<0.001) less time (mean 34.06s/tooth) than the manually coated group (mean 55.14s/tooth). Shear bond strength of the manually coated group was significantly (p<0.001) higher (mean 13.32 MPa) than the APC Flash-Free group (mean 10.95 MPa). CONCLUSION: The APC Flash-Free free system is efficient and allows for reduced chair time during the bonding appointment while attaining a mean shear bond strength of 10.95MPa, which is higher than the minimum shear bond strength of between 4MPa and 7MPa1.


INTRODUCTION

Since the inception of the acid etch technique described by Buonocore23-5 When a bracket is bonded to a tooth, it is either bonded by manually coating brackets with adhesive or with brackets pre-coated with adhesive.6, 7 8, 9

The second problem with manually coated brackets is that clinically, a common site of enamel demineralization is at the enamel-adhesive interface of the tooth surrounding an orthodontic bracket.8-11 When the excess adhesive flash is not removed adequately, the rough adhesive surface remaining provides a site for rapid attachment and growth of oral microorganisms.12-14 Patients undergoing orthodontic treatment, thus face a high risk of developing enamel demineralization (white spot lesions) and caries at the bracket-enamel interface.14

The APC Flash-Free Adhesive Coated Appliance system was developed by 3M Unitek (Monrovia, Calif) in 2013 in an attempt to eliminate flash removal.9, 159,15 The bracket base comprises a nonwoven polypropylene mesh infused with a low viscosity resin.9,15 When applying pressure to the compressible mat, the resin is expressed in sufficient quantity to spread out and conform to the tooth surface, making uniform and consistent contact with no flash to clean-up.15,16 The low viscosity resin is achieved by reducing the filler content of the adhesive,1718

In addition, Foersch et al.99

Research performed by 3M Unitek reported a reduced bonding time, adequate bond strength, and no adhesive flash clean-up when the APC system was applied to ceramic brackets.15

The objective of this study was to determine (1) the shear bond strength (SBS) between APC Flash-Free™ adhesive coated and manually coated metal Victory Series™ (3M Unitek, Monrovia, Calif) brackets and (2) to determine whether there is a significant difference in the bond time between the two systems.

MATERIALS AND METHODS

Ethics Approval

The study protocol was approved by the Faculty of Health Sciences Research Ethics Committee, University of Pretoria, South Africa. Ethics Reference No.: 499/2017.

Sample

Forty extracted human premolar teeth were collected from the Maxillofacial and Oral Surgery clinic. The sample size was guided by previously published in-vitro studies evaluating bonding time and bracket bond strength on extracted human teeth.6,19-22 Any deposits on the teeth were removed with a brush or a dental scaler and rinsed under running water. The inclusion criteria for the extracted teeth were sound enamel surfaces with no carious lesions, fluorosis or cracks, no demineralisation of the enamel, and no restorations present. The teeth were randomly divided into two groups (20 per group) and stored in a 0.1% Thymol solution to inhibit any microbial growth.23

Brackets

Adhesive precoated brackets were introduced by 3M Unitek in 1991.24242515

METHODOLOGY

Bonding procedure: The bonding surfaces of each tooth were polished with non-fluoridated pumice and water for 15 seconds. The surfaces were then rinsed, air dried and etched with 37% phosphoric acid (Ultraetch™, Ultradent, St Louis, MO, USA) for 30 seconds. After rinsing and drying, Transbond XT™ primer (3M Unitek, Monrovia, Calif) was applied and thinned out using a burst of air. The bonding procedure was performed by the principal investigator as follows:

• Group 1: Transbond XT was applied to the bracket base directly. The bracket was then placed on the tooth with Arm constant pressure. The excess flash was then removed, and the bracket was corrected to the ideal occluso-gingival and mesio-distal position on the tooth.

• Group 2: The APC Flash-Free coated brackets were removed from their individually packaged containers and placed onto the tooth with Arm constant pressure. The bracket was then corrected to the ideal occluso-gingival and mesio-distal position on the tooth.

An LED light curing unit (Woodpecker™, Model: LED.B, Woodpecker Medical Instrument Co, Guilin) was used to cure the adhesive. The light was applied for ten seconds (five seconds from the mesial and 5 seconds from the distal of the bracket) at an intensity of between 1070 mw/ cm22

Measurement of bond time: The bonding procedure was timed to the nearest hundredth of a second by an independent observer using a stopwatch. Timing began when the operator secured the bracket onto the bracket holding forceps and was stopped when the operator deemed the bracket to be in the correct occluso-gingival and mesio-distal position.

Thermocycling: Thermocycling tests were developed after it was noted that oral temperature changes resulted in stresses at the restorative-substrate interface.26in-vitro by immersing specimens in circulating baths set at predetermined temperatures for predetermined time durations.2728, 293030

Debonding procedure: The bonded teeth were secured into copper rings which were then mounted onto the clamp of an Instron™ universal testing machine (Model 3366, Instron Corp, Norwood, MA, USA). The mounted teeth were adjusted until the occlusal portion of the bracket was parallel to the shearing blade of the testing machine. The shearing blade of the Instron™ machine applied an occluso-gingival load to the bracket at a speed of 1mm per minute and SBS was recorded in Mega Pascals (MPa).

Statistical Analysis

Based on previous studies6, 9,16 31

Descriptive statistics, including the mean, standard deviation, minimum, and maximum values, were calculated for each group tested. The two-sample t-test was used to test for differences between groups and any statistical interaction between the different adhesives used. The Shapiro-Wilks test was used to test the normality of the distribution of Shear Bond Strength and Bond Time. All statistical calculations were performed at the 95% confidence interval.

RESULTS

Bond Time

The mean bond time of brackets in Group 1 (Transbond XT™) was 55.14s (min: 49.12, max: 59.46, SD: 3.14). Brackets in Group 2 (Flash-Free) were bonded in a mean time of 34.06s (Table 1). According to the two-sample t-test, the time-saving effect of Group 2 (FlashFree) (21.01s per bracket) was found to be statistically significant (p<0.001).

Shear Bond Strength

The mean SBS of the brackets in Group 1 (Transbond XT™) was 13.32 MPa (min: 9.23, max: 15.67, SD: 1.72). The mean SBS of the brackets in Group 2 (Flash-Free) was 10.95 MPa (min: 6.19 MPa, max: 16.7 MPa, SD: 2.29) (Table 2). According to the two-sample t-test, Group 1 (Transbond XT™) brackets bonded with a significantly higher shear bond strength than brackets in Group 2 (Flash-Free) (p<0.001).

DISCUSSION

The objective of this study was to evaluate the shear bond strength (SBS) and bond time of a flash free pre-coated bracket adhesive system (APC Flash-Free, 3M Unitek, Monrovia, California).

This in vitro study demonstrated a statistically significant decrease in bonding time when using the pre-coated APC Flash-Free™ system [Figure 1]. This was expected since two steps (adhesive application and excess adhesive flash removal) are eliminated from the bonding procedure. The flash-free system took, on average, 21 seconds less to bond per tooth. This agrees with the results of Lee and Kanavakis6917 96173232

In this study, the ideal bracket position was marked on each tooth. This step was performed to eliminate the time taken to position each bracket at the correct height and mesio-distal position. Due to variations in tooth morphology, positioning the bracket ideally without prior marking could result in unreliable recorded bond times. It should be considered that the brackets of the flash-free system are sealed individually in a light resistant package and the handling of this packaging during the bonding procedure could possibly affect the overall bond time. In this study the time taken to open the package was not considered when measuring bond time.

Thermocycling is a method of simulating changes in oral temperature which occurs during routine function. This is achieved using circulating baths set at predetermined temperatures. Previous studies have demonstrated that after thermocycling, a significant decrease in shear bond strength was observed.28, 296, 9, 16

In this study the mean shear bond strength of Group 2 (Flash-Free) (10.95 MPa) was significantly lower than that of Group 1 (Transbond XT™) (13.32 MPa) (p<0.001) [Figure 2]. These results conflict with the results of Lee and Kanavakis66

In addition, the bond strength achieved by Lee and Kanavakis63334633, 346

According to Reynolds, the minimum shear bond strength required for clinical success should be in the range of 6 to 8 MPa.1632

CONCLUSION

• This study demonstrated that the APC Flash-Free system applied to metal brackets produced shear bond strengths adequate for clinical use.

• The APC Flash-Free™ system is a convenient bonding method which could potentially save significant chair time.

Conflict of interest

None declared.

REFERENCES

  • 1. Reynolds IR. A Review of Direct Orthodontic Bonding. Br J Orthod. 1975;2(3):171-8.
  • 2. Buonocore MG. A simple method of increasing the adhesion of acrylic Alling materials to enamel surfaces. J Dent Res. 1955;34(6):849-53.
  • 3. Anusavice KJ, Phillips RW, Shen C, Rawls HR. Phillips' science of dental materials. St. Louis, Mo.: Saunders; 2013.
  • 4. Green JIJ. The origins and evolution of fixed orthodontic appliances. J Dent Nurs. 2014;10:524-8.
  • 5. Ewoldsen N, Demke RS. A review of orthodontic cements and adhesives. Am J Orthod Dentofacial Orthop. 2001;120(1):45-8.
  • 6. Lee M, Kanavakis G. Comparison of shear bond strength and bonding time of a novel flash-free bonding system. Angle Orthod. 2016;86(2):265-70.
  • 7. Grünheid T, Sudit GN, Larson BE. Debonding and adhesive remnant cleanup: an in vitro comparison of bond quality, adhesive remnant cleanup, and orthodontic acceptance of a flash-free product. Eur J Orthod. 2015;37(5):497-502.
  • 8. Kim J, Kanavakis G, Finkelman MD, Lee M. Microleakage under ceramic flash-free orthodontic brackets after thermal cycling. Angle Orthod. 2016;86(6):905-8.
  • 9. Foersch M, Schuster C, Rahimi RK, Wehrbein H, Jacobs C. A new flash-free orthodontic adhesive system: A first clinical and stereomicroscopic study. Angle Orthod. 2016;86(2):260-4.
  • 10. Gwinnett AJ, Ceen RF. Plaque distribution on bonded brackets: a scanning microscope study. Am J Orthod Dentofacial Orthop. 1979;75(6):667-77.
  • 11. Ho CS, Ming Y, Foong KW, Rosa V, Thuyen T, Seneviratne CJ. Streptococcus mutans forms xylitol-resistant biofilm on excess adhesive flash in novel ex-vivo orthodontic bracket model. Am J Orthod Dentofacial Orthop. 2017;151(4):669-77.
  • 12. Mei L, Busscher HJ, van der Mei HC, Ren Y. Influence of surface roughness on streptococcal adhesion forces to composite resins. Dent Mater J. 2011;27(8):770-8.
  • 13. Årtun J, Brobakken BO. Prevalence of carious white spots after orthodontic treatment with multibonded appliances. Eur J Orthod. 1986;8(4):229-34.
  • 14. Rosenbloom RG, Tinanoff N. Salivary Streptococcus mutans levels in patients before, during, and after orthodontic treatment. Am J Orthod Dentofacial Orthop. 1991;100(1):35-7.
  • 15. Cinader DK, Aung M, Ugai R, Conley A. APC™ Flash-free adhesive: a technical overview. Orthodontic Perspectives. 2013;20(1):7-9.
  • 16. Grunheid T, Sudit GN, Larson BE. Debonding and adhesive remnant cleanup: an in vitro comparison of bond quality, adhesive remnant cleanup, and orthodontic acceptance of a flash-free product. Eur J Orthod. 2015;37(5):497-502.
  • 17. Grünheid T, Larson BE. Comparative assessment of bonding time and 1-year bracket survival using flash-free and conventional adhesives for orthodontic bracket bonding: A split-mouth randomized controlled clinical trial. Am J Orthod Dentofacial Orthop. 2018;154(5):621-8.
  • 18. Faltermeier A, Rosentritt M, Faltermeier R, Reicheneder C, Müßig D. Influence of filler level on the bond strength of orthodontic adhesives. Angle Orthod. 2007;77(3):494-8.
  • 19. Bishara SE, Gordan VV, VonWald L, Jakobsen JR. Shear bond strength of composite, glass ionomer, and acidic primer adhesive systems. Am J Orthod Dentofacial Orthop. 1999;115(1):24-8.
  • 20. Bishara SE, VonWald L, Laffoon JF, Warren JJ. Effect of a self-etch primer/adhesive on the shear bond strength of orthodontic brackets. Am J Orthod Dentofacial Orthop. 2001;119(6):621-4.
  • 21. Bishara SE, Olsen ME, Jakobsen JR. Evaluation of a new light-cured orthodontic bonding adhesive. Am J Orthod Dentofacial Orthop. 1998;114(1):80-7.
  • 22. Toledano M, Osorio R, Osorio E, Romeo A, de la Higuera B, Garcia-Godoy F. Bond strength of orthodontic brackets using different light and self-curing cements. Angle Orthod. 2003;73(1):56-63.
  • 23. Aydin B, Pamir T, Baltaci A, Orman MN, Turk T. Effect of storage solutions on microhardness of crown enamel and dentin. Eur J Dent. 2015;9(2):262-6.
  • 24. Hirani S, Sherriff M. Bonding characteristics of a self-etching primer and precoated brackets: an in vitro study. Eur J Orthod. 2006;28(4):400-4.
  • 25. Oliver B, Dama M. A retrospective six-months clinical trial of adhesive precoated brackets and bonding system. J Can Dent Assoc. 1997;63(2):101-3, 7.
  • 26. Nelsen RJ, Paffenbarger GC, Wolcott RB. Fluid exchange at the margins of dental restorations. JADA. 1952;44(3):288-95.
  • 27. Gale MS, Darvell BW. Thermal cycling procedures for laboratory testing of dental restorations. J Dent 1999;27(2):89-99.
  • 28. Summers A, Kao E, Gilmore J, Gunel E, Ngan P. Comparison of bond strength between a conventional resin adhesive and a resin-modified glass ionomer adhesive: an in vitro and in vivo study. Am J Orthod Dentofacial Orthop. 2004;126(2):200-6.
  • 29. Rix D, Foley TF, Mamandras A. Comparison of bond strength of three adhesives: composite resin, hybrid GIC, and glass-filled GIC. Am J Orthod Dentofacial Orthop. 2001;119(1):36-42.
  • 30. Bishara SE, Ostby AW, Laffoon JF, Warren J. Shear bond strength comparison of two adhesive systems following thermocycling: a new self-etch primer and a resin-modified glass ionomer. Angle Orthod. 2007;77(2):337-41.
  • 31. Bishara SE, Oonsombat C, Soliman MM, Warren JJ, Laffoon JF, Ajlouni R. Comparison of bonding time and shear bond strength between a conventional and a new integrated bonding system. Angle Orthod. 2005;75(2):237-42.
  • 32. Alakttash AM, Fawzi M, Bearn D. Adhesive precoated bracket systems and operator coated bracket systems: Is there any difference? A systematic review and meta-analysis. Angle Orthod. 2019;89(3):495-504.
  • 33. Reddy YG, Sharma R, Singh A, Agrawal V, Chaturvedi S. The Shear Bond Strengths of Metal and Ceramic Brackets: An in-Vitro Comparative Study. JCDR. 2013;7(7):1495-7.
  • 34. Uysal T, Ustdal A, Kurt G. Evaluation of shear bond strength of metallic and ceramic brackets bonded to enamel prepared with self-etching primer. Eur J Orthod. 2009;32(2):214-8.

Correspondence:
Dr Rashid Essop
Registrar, Department of Orthodontic, University of Pretoria
Email:
Tel: 0833388882

Author contributions:


1 . Dr Rashid Essop: 80%
2 . Dr Emad Ghabrial: 10%
3 . Prof Piet J Becker: 10%

Correspondence

essoprashid@gmail.com

Publication Dates

  • Publication in this collection
    14 Feb 2023
  • Date of issue
location_on
The South African Dental Association (SADA) The South African Dental Association (SADA), Private Bag 1, Houghton, Johannesburg, Gauteng, ZA, 2041, Tel: +27 11 530 6520 - E-mail: nmojela@sada.co.za
rss_feed Acompanhe os números deste periódico no seu leitor de RSS
Ir para o topo