Abstract
BACKGROUND: The use of point-of-care INR (international normalised ratio) testing is on the rise because of its ability to facilitate quick clinical decisions as it allows for rapid turnaround time of INR results. Ensuring the quality of these tests is crucial when implementing them in new settings. OBJECTIVE: This study aimed to assess the sensitivity, specificity, and clinical agreement between the CoaguChek® XS INR meter and the Beckman Coulter ACL 7000 automated coagulometer in outpatients undergoing warfarin therapy. METHODS: A prospective cohort design was employed for this comparison. The analysis included Spearman's correlation, a Bland-Altman plot, and a Cohen's Kappa analysis of paired INR values. Key outcome measures included the percentage sensitivity, specificity, and clinical agreement of the two methods. A p-value < 0.05 was considered statistically significant. RESULTS: The study involved 89 patients, predominantly female (59.4%), with a mean age of 46 years. Deep vein thrombosis was the most common indication for warfarin therapy (51.9%). The mean difference between INR values determined by the two methods was -0.15 (± 0.27, N = 89), and the Spearman rank correlation coefficient was 0.97 (p < 0.001, N = 89). A clinical agreement of 83.8% (p < 0.05) was obtained. The sensitivity was 94.4%, with specificities of 82.5% for subtherapeutic and 100% for supratherapeutic INRs CONCLUSION: The CoaguChek® XS INR meter was proven to be a reliable alternative to the Beckman Coulter ACL 7000 automated coagulometer for monitoring INR results and adjusting warfarin therapy in well-trained patients with stable INR values. WHAT THIS STUDY ADDS: The CoaguChek® XS INR meter was proven to be a reliable alternative to the Beckman Coulter ACL 7000 automated coagulometer for monitoring INR results..
CoaguChek® XS INR meter; Beckman Coulter ACL 7000 automated coagulometer; sensitivity; specificity; clinical concordance
Introduction
Tadesse et al. highlighted that a key factor affecting the quality of anticoagulation control in healthcare facilities is the high patient load.12
Point-of-care tests are performed at or near the patient's site of care, eliminating the need to transfer samples to a laboratory for testing. In the field of haematology, various POC tests such as activated clotting time, thromboelastography, platelet function, and D-dimer are available. In particular, the measurement of INR for monitoring warfarin therapy is crucial in this domain.345,66
The CoaguChek® XS INR meter from Roche Diagnostics has been shown in several studies to provide accurate and precise results comparable to laboratory-based INR determination.7,8® XS INR meter and conventional laboratory methods, differences in results were noted at higher INR values. Studies have highlighted the importance of considering factors like serum fibrinogen levels in interpreting results from INR-POC meters.9® XS INR meter, are crucial metrics that need to be considered.
It is important to remain vigilant about the potential for errors in INR results when using POC-INR devices, as highlighted by a case report involving a CoaguChek® S INR meter.101112® XS INR meter, these comparisons have not been carried out using the Beckman Coulter ACL 7000 automated coagulometer as the comparator. Finally, the CoaguChek® XS INR meter has not been validated in the Namibian setting where some of the factors exist, such as malnutrition, which could affect its accuracy.13® XS INR meter in comparison to the Beckman Coulter ACL 7000 automated coagulometer of the Namibia Institute of Pathology for monitoring patients on warfarin therapy at the Warfarin Outpatient Clinic of Windhoek Central Hospital.
Methods
Ethical considerations
This research was part of a larger intervention study in which some of the patients had their INR checked using both the CoaguChek® XS INR meter and the Beckman Coulter ACL 7000 automated coagulometer.12
Validation procedure
Starting from 02 June 2021 to 28 July 2021, the pharmacist carried out POC INR testing using the CoaguChek® XS INR meter on 89 patients who also had their INRs checked by the Namibia Institute of Pathology laboratory. The test strips of the CoaguChek® XS INR meter were costly to procure; this limited the number of patients tested with the CoaguChek® XS INR meter to 89. Convenience sampling was employed as the 89 patients had to have given consent to participate in the main study and visited the clinic from 02 June 2021 to 28 July 2021. Permission was sought from the patients to test their INR using the POC meter. International normalised ratio-POC testing involved setting up the POC meter (by configuring the meter to INR measurement and inserting the INR test strip into the meter), cleaning the surface of the finger with an alcohol swab, pricking the finger, collecting the drop of blood using a sample collector, applying the sample onto the test strips and waiting for the meter to display the INR results. Both the POC meter INR results and Beckman Coulter ACL 7000 automated coagulometer INR results were recorded on the data collection tool.
Research instruments
Two research instruments were used: a data collection tool. CoaguChek® XS INR meter, and the Beckman Coulter ACL 7000 automated coagulometer. The data collection tool was adopted from the information/data collection system used by a free clinic in the United States (Birmingham Free Clinic, Pittsburgh, Pennsylvania). This tool was slightly modified to fit the purpose of this study.14® XS INR meter (Roche Diagnostics, Basel, Switzerland) was used to obtain the POC-INR values.15® XS INR testing involves three stages: the pre-analytical, analytical, and post-analytical.1516
Data analysis
Statistical analyses were conducted using IBM Statistical Package for Social Sciences for Windows version 29.0 (Statistical Package for Social Sciences Inc, Chicago, Illinois, United States), with statistical significance set at p ≤ 0.05. The sensitivity and specificity of the CoaguChek® XS INR meter were evaluated by comparing paired INR results from the POC device and the Beckman Coulter ACL 7000 automated coagulometer in each patient. This comparison involved Spearman's correlation analysis, a Bland-Altman plot, and a Cohen's Kappa analysis of the paired INR values. Spearman's coefficient interpretation was based on a scale ranging from very weak to very strong.1718
Results
Demographics of the participants
Of the 89 patients for whom INR values determined by two methods were available, 55 (61.8%) were women and 34 (38.2%) were men. Their median age was 42 years and 50% of the participants were aged between 18 and 66 years. The top three most frequent clinical indications for warfarin use were: deep vein thrombosis (n = 45, 50.6%), double valve replacement (mitral and aortic) (n = 18, 20.2%), and atrial fibrillation (n = 9, 10.1%).
Correlation analysis
International normalised ratio results from the CoaguChek® XS meter were positively and highly correlated to the Beckman Coulter ACL 7000 automated coagulometer results (rs: 0.97, p < 0.001, N = 89) (Figure 1).
The Bland-Altman plot
The Bland-Altman plot, inclusive of outliers, showed that the mean INR difference between the CoaguChek® XS INR meter and the Beckman Coulter ACL 7000 automated coagulometer was -0.15 (± 0.27, N = 89) (Figure 2). The limits of agreement were between -0.68 and 0.38, a range in which 95% of the differences fell in the paired INR values between the two methods. Five INR differences fell outside the aforementioned limits of agreement. The first INR difference outside the limits of agreement had a mean INR of about 2.5. A negative trend was evident as shown by the regression line (y = 5.94 - (0.09 * X). The Bland-Altman plot essentially showed that the difference in the paired INR values between the two methods increased with the increase in mean paired INR values, meaning that the higher the INR the more likely it was to be a discrepancy.
Cohen's Kappa analysis
The agreement of INR measurements between the CoaguChek® XS and Beckman Coulter ACL 7000 automated coagulometer was further assessed based on three categories, i.e. subtherapeutic, therapeutic, and supratherapeutic, using Cohen's Kappa analysis. The results showed a statistically significant and high clinical agreement between CoaguChek® XS INR and Beckman Coulter ACL 7000 automated coagulometer of 83.8%, p < 0.05.
In comparison to the Beckman Coulter ACL 7000 automated coagulometer, the sensitivity of the CoaguChek® XS INR meter in detecting therapeutic INRs accurately was assessed to be 94.4%. The POC device had a specificity of 82.5% in detecting subtherapeutic INR values correctly and 100% for detecting supratherapeutic INR values correctly (Table 1).
Discussion
In anticoagulation services, patient education, along with the safe use of POC devices for INR monitoring and warfarin therapy adjustments, is crucial for introducing patient self-management and potentially reducing the workload on healthcare workers. The accuracy and precision of POC devices in measuring INR values must be assessed. This study showed a very high positive correlation (r = 0.973, p < 0.001) between the Roche CoaguChek® XS INR meter and the ACL INR analyser of the Namibia Institute of Pathology, highlighting its potential role in expanding access to anticoagulation services in Namibia.
Comparing these results with previous studies, our findings align with the high correlation reported by Kalçlk et al. between Roche's CoaguChek® XS INR meter and the laboratory method (STAGO STAR).7® XS INR meter and STAGO STAR analyser increased with higher mean INR values, consistent with findings from other studies.7,194
The Bland-Altman plot in our study showed a narrower limit of agreement compared to that reported by Moore et al., with 95% of the differences falling within these limits.48,2021
Moreover, Gardiner et al. explained that out-of-range INRs from POC devices were mostly because of the use and preparation of INR testing rather than faulty instruments or test strips.203® INR testing did not receive formal training on the use of the device. The relative inaccuracy of POC devices in measuring INR at higher mean INR values led authors like Sharma et al. to recommend that laboratory INR testing should be conducted in patients with an INR result of > 4.0 from the POC device.22
It is important to note that almost 100% of the POC results matched with the overall assessment of INR control: subtherapeutic, therapeutic, and supratherapeutic. The Cohen's Kappa analysis between Roche's CoaguChek® XS INR meter and Beckman Coulter ACL 7000 automated coagulometer yielded a Kappa value of 0.838 (p < 0.001). The Kappa analysis is used to assess clinical concordance between two methods based on a grading system (k value of < 0.4, poor; 0.4-0.75, fair to good; > 0.75, excellent).18® XS INR meter had a sensitivity of 94.4% in detecting therapeutic INR values correctly. It also revealed a specificity of 82.5% (subtherapeutic) and 100% (supratherapeutic) in detecting non-therapeutic INR values correctly. Although most studies do not report on the sensitivity and specificity of POC devices, Weyrauch et al. similarly reported a sensitivity of 91.7% and specificity of 99.2%.23
This study confirmed that the CoaguChek® XS INR meters are safe for use in patients attending our warfarin clinic. Their major drawback is that, without reimbursement, they can be costly for health systems.2024® XS INR meter results and Beckman Coulter ACL 7000 automated coagulometer INR results. With proper training for patients and healthcare workers, POC testing can be implemented safely at the Warfarin Outpatient Clinic of Windhoek Central Hospital. With reimbursement, the use of POC devices can be expanded to include supervised patient self-management of warfarin therapy.
Limitations
A limitation of the study was that the INR results using the CoaguChek® XS INR meter could not be obtained in all patients because it was too costly to procure the test strips for all patients in the intervention study. This was the sole determinant of the sample size, with no power determinations carried out. Second, the test was carried out by an early-career pharmacist who had no formal training on the CoaguChek® XS INR meter.
Conclusion
In conclusion, the CoaguChek® XS INR meter is a safe alternative to Beckman Coulter ACL 7000 automated coagulometer for monitoring INR results and adjusting warfarin therapy in well-trained patients within the tested range. Therefore, we recommend that Roche's CoaguChek® XS INR monitoring can be carried out safely among outpatients in Namibia.
Competing interest
The authors, Moses M. Thikukutu, Lauren J. Jonkman, Bonifasius S. Singu, Mubita Mwangana, and Roger K. Verbeeck, declare that they have no financial or personal relationships that may have influenced them inappropriately in writing this article.
CRediT authorship contribution
Moses M. Thikukutu: Conceptualisation, Formal Analysis, Investigation, Methodology, Project administration, Visualisation, Writing - original draft. Lauren J. Jonkman: Conceptualisation, Methodology, Supervision, Writing - review & editing. Bonifasius S. Singu: Conceptualisation, Methodology, Supervision, Writing - review & editing. Mubita Mwangana: Conceptualisation, Formal, analysis, Methodology, Supervision, Writing - review & editing. Roger K. Verbeeck: Conceptualisation, Formal analysis, Methodology, Supervision, Writing - review & editing. All authors reviewed the article, contributed to the discussion of results, approved the final version for submission and publication, and take responsibility for the integrity of its findings.
Sources of support
This research received no specific grant from any funding agency in the public, commercial or not-for-profit sectors.
Data availability
The data that support the findings of this study are not available openly because of sensitivity and confidentiality and are available from the corresponding author, Moses M. Thikukutu, upon reasonable request.
Disclaimer
The views and opinions expressed in this article are those of the authors and are the product of professional research. They do not necessarily reflect the official policy or position of any affiliated institution, funder, agency, or that of the publisher. The authors are responsible for the results, findings, and content of this article.
Acknowledgements
This article is based on research originally conducted as part of the master's thesis of Moses Mukwipure Thikukutu, titled 'Standard-of-care versus pharmacist-supported interventions to improve anticoagulation control in Namibia', submitted to the Faculty of Health Sciences and Veterinary Medicine, Department of Pharmacology and Therapeutics, University of Namibia in 2024. The thesis was supervised by Lauren Jenifer Jonkman, Bonifasius Siyuka Singu, Mwangana Mubita, and Roger Karel Verbeeck. The manuscript has since been revised and adapted for journal publication. The original thesis is available at: https://link.springer.com/article/10.1007/s44250-025-00278-8. In addition, the authors would like to thank the cardiac team of Windhoek Central Hospital, who were fundamental in recruiting patients in the study.
References
-
1.Tadesse TA, Abiye AA, Endale S, Yadeta D, Chelkeba L, Fenta TG. Challenges of anticoagulation management service and need of establishing pharmacist-led anticoagulation clinic in tertiary care teaching hospital, Ethiopia: A qualitative study. J Multidiscip Healthc. 2022;15:743-754. https://doi.org/10.2147/JMDH.S359558
» https://doi.org/10.2147/JMDH.S359558 -
2.Mouton JP, Blockman M, Sekaggya-Wiltshire C, et al. Improving anticoagulation in sub-Saharan Africa: What are the challenges and how can we overcome them? Br J Clin Pharmacol. 2021;87(8):3056-3068. https://doi.org/10.1111/bcp.14768
» https://doi.org/10.1111/bcp.14768 -
3.Perry DJ, Fitzmaurice DA, Kitchen S, Mackie IJ, Mallett S. Point-of-care testing in haemostasis. Br J Haematol. 2010;150(5):501-514. https://doi.org/10.1111/j.1365-2141.2010.08223.x
» https://doi.org/10.1111/j.1365-2141.2010.08223.x -
4.Moore GW, Henley A, Cotton SS, Tugnait S, Rangarajan S. Clinically significant differences between point-of-care analysers and a standard analyser for monitoring the International Normalized Ratio in oral anticoagulant therapy: A multi-instrument evaluation in a hospital outpatient setting. Blood Coag. Fibrinolysis. 2007;18(3):287-292. https://doi.org/10.1097/MBC.0b013e328040c105
» https://doi.org/10.1097/MBC.0b013e328040c105 -
5.Moiz B, Rashid A, Hasan M, Jafri L, Raheem A. prospective comparison of point-of-care device and standard analyzer for monitoring of international normalized ratio in outpatient oral anticoagulant clinic. Clin Appl Thromb Hemost. 2018;24(7):1153-1158. https://doi.org/10.1177/1076029617752247
» https://doi.org/10.1177/1076029617752247 -
6.Christensen TD, Larsen TB. Precision and accuracy of point-of-care testing coagulometers used for self-testing and self-management of oral anticoagulation therapy. J Thromb Haemost. 2012;10(2):251-260. https://doi.org/10.1111/j.1538-7836.2011.04568.x
» https://doi.org/10.1111/j.1538-7836.2011.04568.x -
7.Kalçlk M, Yesin M, Gürsoy MO, et al. Comparison of the INR values measured by CoaguChek XS coagulometer and conventional laboratory methods in patients on VKA therapy. Clin Appl Thromb Hemost. 2017;23(2):187-194. https://doi.org/10.1177/1076029615595881
» https://doi.org/10.1177/1076029615595881 -
8.Benade EL, Jacobson BF, Louw S, Schapkaitz E. Validation of the CoaguChek XS international normalized ratio point-of-care analyser in patients at Charlotte Maxeke Johannesburg Academic Hospital, South Africa. S Afr Med J. 2016;106(3):280-283. https://doi.org/10.7196/SAMJ.2016.v106i3.9422
» https://doi.org/10.7196/SAMJ.2016.v106i3.9422 -
9.Kim SJ, Lee EY, Park R, Kim J, Song J. Comparison of prothrombin time derived from CoaguChek XS and laboratory test according to fibrinogen level. J Clin Lab Anal. 2015;29(1):28-31. https://doi.org/10.1002/jcla.21722
» https://doi.org/10.1002/jcla.21722 -
10.Van den Brule JMD, Romkes JH, Brouwer MA, Noyez L. An error in the CoaguChek S® device results in prosthetic heart valve thrombosis. Interact Cardiovasc Thorac Surg. 2010;11(1):123-124. https://doi.org/10.1510/icvts.2010.233288
» https://doi.org/10.1510/icvts.2010.233288 -
11.Arline K, Rodriguez C, Sanchez K. Reliability of point-of-care international normalized ratio measurements in various patient populations. Point Care. 2020;19(1):12-18. https://doi.org/10.1097/POC.0000000000000197
» https://doi.org/10.1097/POC.0000000000000197 -
12.Thikukutu M, Jonkman L, Singu B, Mwangana M, Shidhika F, Verbeeck RK. Standard-of-care versus pharmacist-supported interventions to improve anticoagulation control in Namibia. Discov Health Syst. 2025;4(1):90. https://doi.org/10.1007/s44250-025-00278-8
» https://doi.org/10.1007/s44250-025-00278-8 -
13.Xu Z, Li Y, Yang X, Wang J, Li J. Early detection of coagulation abnormalities in patients at nutritional risk: The novel role of thromboelastography. Am Surg. 2014;80(1):81-86. https://doi.org/10.1177/000313481408000129
» https://doi.org/10.1177/000313481408000129 - 14.University of Pittsburgh. Anticoagulation clinic maintenance phase data collection tool and warfarin dose alteration algorithm. Pittsburgh: University of Pittsburgh; 2007.
-
15.Roche Diagnostics. User manual [homepage on the Internet]. 2011 [cited 2024 Feb 14]. Available from: https://diagnostics.roche.com/content/dam/diagnostics/us/en/products/c/coagucheck-xs/CoaguChek-XS-Professional-System-User-Manual.pdf
» https://diagnostics.roche.com/content/dam/diagnostics/us/en/products/c/coagucheck-xs/CoaguChek-XS-Professional-System-User-Manual.pdf -
16.Beckman Coulter ACL 7000 coagulation analyzer [homepage on the Internet]. Utah: International Diagnostic Equipment; [cited 2025 Mar 26]. Available from: https://www.internationaldiagnostic.com/fr/annonces/7059370-beckman-coulter-il-acl-7000-coagulation-analyzer
» https://www.internationaldiagnostic.com/fr/annonces/7059370-beckman-coulter-il-acl-7000-coagulation-analyzer -
17.Statstutor.ac.uk. Spearman's correlation [homepage on the Internet]. [cited 2023 Mar 26]. Available from: https://www.statstutor.ac.uk/resources/uploaded/spearmans.pdf
» https://www.statstutor.ac.uk/resources/uploaded/spearmans.pdf -
18.McHugh ML. Interrater reliability: The kappa statistics. Biochem Med (Zagreb). 2012;22(3):276-282. https://doi.org/10.11613/BM.2012.031
» https://doi.org/10.11613/BM.2012.031 -
19.Donaldson M, Sullivan J, Norbeck A. Comparison of international normalized ratios provided by two point-of-care devices and laboratory-based venipuncture in a pharmacist-managed anticoagulation clinic. Am J Health Syst Pharm. 2010;67(19):1616-1622. https://doi.org/10.2146/ajhp100096
» https://doi.org/10.2146/ajhp100096 -
20.Gardiner C, Williams K, Mackie IJ, Machin SJ, Cohen H. Patient self-testing is a reliable and acceptable alternative to laboratory INR monitoring. Br J Haematol. 2005;128(2):242-247. https://doi.org/10.1111/j.1365-2141.2004.05300.x
» https://doi.org/10.1111/j.1365-2141.2004.05300.x -
21.Kim JW, Kim MH, Kim KH, et al. Comparison between the portable prothrombin time self-monitor CoaguChek XS and a standard laboratory method, Sysmex CA-1500 for monitoring anticoagulant therapy of outpatients. Korean Circ J. 2007;37(5):216. https://doi.org/10.4070/kcj.2007.37.5.216
» https://doi.org/10.4070/kcj.2007.37.5.216 -
22.Sharma P, Scotland G, Cruickshank M, et al. The clinical effectiveness and cost-effectiveness of point-of-care tests (CoaguChek system, INRatio2 PT/INR monitor and ProTime microcoagulation system) for the self-monitoring of the coagulation status of people receiving long-term vitamin K antagonist therapy, compared with standard UK practice: Systematic review and economic evaluation. Health Technol Assess. 2015;19(48):1-172. https://doi.org/10.3310/hta19480
» https://doi.org/10.3310/hta19480 -
23.Weyrauch E, Walliser G, Hartwell J. Point-of-care international normalized ratio testing is rapid and reliable: A prospective observational cohort study. Int J Acad Med. 2016;2(1):27. https://doi.org/10.4103/2455-5568.183322
» https://doi.org/10.4103/2455-5568.183322 -
24.Bhat M, Abuzied Y, Fagih Z, et al. Efficacy of point-of-care for INR testing compared to standard laboratory methods at a tertiary care hospital in Saudi Arabia. Glob J Qual Saf Healthc. 2020;3(3):98-104. https://doi.org/10.36401/JQSH-19-36
» https://doi.org/10.36401/JQSH-19-36
Accepted: 24 Nov. 2025
Published: 23 July 2026