Abstract
INTRODUCTION: The CellaVision™ DM96 (DM96) is a digital microscopy system which performs well in developed countries. However, to date it has not been evaluated in Africa, where the pathology spectrum encountered is very different. In particular, its utility in a setting with high HIV prevalence has not been assessed, which is of interest because of the morphological aberrations often seen in HIV-positive patients. OBJECTIVES: This study aimed to evaluate the accuracy of the DM96 in a South African laboratory, with emphasis on its performance in samples collected from HIV-positive patients. METHODS: A total of 149 samples submitted for a routine differential white cell count in 2012 and 2013 at the Chris Hani Baragwanath Academic Hospital in Johannesburg, South Africa were included, of which 79 (53.0%) were collected from HIV-positive patients. Results of DM96 analysis pre- and post-classification were compared with a manual differential white cell count and the impact of HIV infection and other variables of interest were assessed. RESULTS: Pre- and post-classification accuracies were similar to those reported in developed countries. Reclassification was required in 16% of cells, with particularly high misclassification rates for eosinophils (31.7%), blasts (33.7%) and basophils (93.5%). Multivariate analysis revealed a significant relationship between the number of misclassified cells and both the white cell count (p = 0.035) and the presence of malignant cells in the blood (p = 0.049), but not with any other variables analysed, including HIV status. CONCLUSION: The DM96 exhibited acceptable accuracy in this South African laboratory, which was not impacted by HIV infection. However, as it does not eliminate the need for experienced morphologists, its cost may be unjustifiable in a resource-constrained setting.
Introduction
The differential white cell count (DWCC) is a frequently requested laboratory investigation and is, historically, a labour-intensive test. Although the guidelines established by the international consensus group for haematology to identify samples where peripheral smear review may be omitted12,34
The CellaVision™ DM96 (CellaVision AB, Lund, Sweden; hereafter, 'DM96') is a digital microscopy system that has the potential to minimise the time required by a morphologist to perform a manual DWCC. It comprises: an automated microscope that scans the blood smear; a digital camera that captures images of all the cellular and particulate material on the slide; and a computer that classifies each image by means of complex algorithms. It has been demonstrated to have good performance characteristics in developed countries,5,6,7,8,91011
Research method and design
Ethical considerations
Ethical clearance was obtained from the Human Research Ethics Committee of the University of the Witwatersrand, Johannesburg, South Africa (clearance number: M090688).
Sample selection and analysis
The study was performed at the National Health Laboratory Service (NHLS) haematology laboratory at CHBAH in Soweto, Johannesburg, South Africa. A total of 149 peripheral blood samples were selected from EDTA-anticoagulated specimens submitted for a DWCC over the course of 2012 and 2013. Samples were selected to cover a wide range of white cell counts and were included only if the patient had an HIV test result available in the laboratory information system (LIS) (DisaLab Version 04.16.04.373, Laboratory System Technologies, Boksburg, Gauteng, South Africa). Slides were made and stained with May-Grünwald/Giemsa by an automated slide maker and stainer (SP-100, Sysmex, Kobe, Japan) as per standard operating procedure for the performance of a DWCC. Each smear was examined by an experienced morphologist and analysed with the DM96 within a 24-hour period. The manual DWCC was performed on 100 cells. The DM96 was set to analyse 110 images per sample. Each image was pre-classified as an unidentified cell, a neutrophil, a lymphocyte, a monocyte, a granulocyte precursor (i.e., a promyelocyte, myelocyte or metamyelocyte), a blast, an eosinophil, a basophil, a nucleated red cell, a giant platelet, a platelet aggregate, a smear cell or an artefact. Cells which the instrument identified as being band cells were classified as neutrophils. The images were then viewed by the same morphologist who performed the manual DWCC. The morphologist either verified the DM96 pre-classification or reclassified the cells. Hereafter, the initial DWCC performed by the DM96 will be referred to as the 'pre-classification DWCC' and the final DWCC following review by the morphologist will be referred to as the 'post-classification DWCC'. 'Misclassification' will refer to cells requiring reclassification by the morphologist.
For each sample, the data available in the LIS were reviewed and pertinent information recorded. This included the clinical information provided by the attending clinician, demographic details, evidence of infection (including C-reactive protein levels and culture results), recent HIV viral load and CD4 counts (where appropriate), exposure to anti-retroviral therapy, as well as bone marrow aspirate/trephine biopsy and other histology findings. The data were recorded in Excel™ spreadsheets (Microsoft Office Excel™ 2007, Redmond, Washington, United States). No identifying patient information was recorded. All data collection and analysis were performed by a haematopathologist working in the haematology laboratory of the CHBAH.
Statistical analysis
Demographic data are presented as medians (interquartile range [IQR]), mean (± standard deviation [SD]) and proportions, as appropriate. The accuracy of the DM96 DWCC was evaluated by linear regression (pre- and post-classification) and Bland-Altman (post-classification) analyses comparing the absolute count for neutrophils, monocytes, lymphocytes, eosinophils, basophils and blasts with those obtained by manual counting. The misclassification rate was determined as the proportion of all counted cells requiring reclassification. A multivariate linear regression analysis was performed to assess the impact of variables of interest, including HIV infection, on misclassification rates. Any data point with a standard residual of greater than 2.5 was excluded from analysis and p-values less than 0.05 were considered statistically significant. Statistical analysis was performed using STATISTICA software, version 12.5 (Stat Soft [Pty] Ltd; Tulsa, Oklahoma, United States).
Results
Patient demographic and clinical data are summarised in Table 1. The median white cell count (WCC) was 6.76 × 10999n = 79; 53%) of patients were HIV-positive, whilst ~40% (n = 61) had a history of malignancy. The prevalence of malignancy was similar between HIV-positive and HIV-negative patients, although the spectrum of malignant disease varied substantially between these two groups. The dominant malignancy amongst HIV-positive patients was high-grade lymphomas, whereas leukaemias were more common amongst HIV-negative patients. Despite this difference, the proportion of patients with abnormal cells present in the peripheral blood was similar between the two groups. Not surprisingly, evidence of infection was substantially more common amongst HIV-positive patients.
Analysis of accuracy
Overall, pre-classification DWCC accuracy was very poor, with good correlation occurring only for neutrophils (Table 2). Correlation co-efficient (CC) values improved substantially post-classification, but remained below 0.9 for all cell types except neutrophils and blasts, and was poor for eosinophils, basophils and monocytes. However, the Bland-Altman analysis showed that agreement was within acceptable limits of bias for neutrophils, lymphocytes and blasts, with borderline acceptable agreement for monocytes. A substantial negative bias was evident for both eosinophils and basophils, but because of the low levels of these cell types, this translated into small differences in absolute values, the clinical significance of which was negligible.
Analysis of misclassification rates
Although only 3.5% of cells were classified as 'unidentified' in the pre-classification DWCC, overall 16% required reclassification (Table 3). As the most common cell type, neutrophils were the most frequently misclassified, but overall, only 7.6% of all neutrophils required reclassification. In contrast, close to 30% of monocytes, eosinophils and blasts, as well as > 90% of basophils, were misclassified. Multivariate analysis revealed a significant relationship between the number of misclassified cells and both the WCC and the presence of malignant cells in the blood (Table 4). No other variables analysed, including HIV status, had a significant association with the number of misclassified cells.
Discussion
In this study, we assessed the performance of the DM96 as compared to a manual DWCC in 149 samples collected from patients with a wide range of infections and haematological pathologies in a large South African state hospital serving a population with high HIV prevalence. Results from similar studies performed in developed countries have varied. Park et al. showed excellent correlation (CC > 0.9) for all cell types except promyelocytes and basophils.136,7,8,913
Although HIV infection has well documented effects on white cell morphology,1114
The time required to perform a DM96 DWCC has been demonstrated to be less than that required for a manual DWCC,6,8,95136
Limitations
There are some limitations to our study which should be considered when interpreting the results. A large proportion of our HIV-negative patients had a malignancy, so they cannot be regarded as being representative of the normal population. No normal control group was included, so the accuracy of the DM96 was therefore not assessed in the normal population. Our sample size is relatively small and the number of samples included with extreme WCCs and abnormal cells present most likely skewed our results to some extent.
Conclusion
This study showed that the performance of the DM96 in an African laboratory serving a population with a high HIV-prevalence was similar to that described in developed countries. However, significant intervention from experienced morphologists remains necessary for the validation of results, particularly when malignant cells are present. Given the substantial cost of this sophisticated instrument, it would be difficult to justify its routine use in a resource-constrained setting.
Competing interests
The authors declare that they have no financial or personal relationship(s) that may have inappropriately influenced them in writing this article.
Sources of support
This project was performed on samples submitted for routine testing, and therefore required no funding.
Authors' contributions
J.L.V. (University of the Witwatersrand and National Health Laboratory Services) was responsible for peripheral smear and DM96 analysis, data analysis and writing the paper. S.L. (National Health Laboratory Services) assisted with peripheral smear and DM96 analysis. N.A. (University of the Witwatersrand and National Health Laboratory Services) co-wrote the paper and provided editorial oversight.
Acknowledgements
The authors would like to acknowledge and thank all of the technical staff members of the National Health Laboratory Service haematology laboratory at the Chris Hani Baragwanath Academic Hospital.
References
-
1. Barnes PW, McFadden SL, Machin SJ, et al. The international consensus group for hematology review: suggested criteria for action following automated CBC and WBC differential analysis. Lab Hematol. 2005;11(2):83-90. http://dx.doi.org/10.1532/LH96.05019
» http://dx.doi.org/10.1532/LH96.05019 - 2. Beck S, & Doig K. Laboratory managers' views on attrition and retention of laboratory personnel. Clin Lab Sci. 2005 Fall;18(4):238-247.
-
3. Petti CA, Polage CR, Quinn TC, et al. Laboratory medicine in Africa: a barrier to effective health care. Clin Infect Dis. 2006 Feb 1;42(3):377-382. http://dx.doi.org/10.1086/499363
» http://dx.doi.org/10.1086/499363 - 4. Novis DA, Walsh M, Wilkinson D, et al. Laboratory productivity and the rate of manual peripheral blood smear review: a College of American Pathologists Q-Probes study of 95,141 complete blood count determinations performed in 263 institutions. Arch Pathol Lab Med. 2006 May;130(5):596-601.
-
5. Cornet E, Perol JP, Troussard X. Performance evaluation and relevance of the CellaVision DM96 system in routine analysis and in patients with malignant hematological diseases. Int J Lab Hematol. 2008 Dec;30(6):536-542. http://dx.doi.org/10.1111/j.1751-553X.2007.00996.x
» http://dx.doi.org/10.1111/j.1751-553X.2007.00996.x -
6. Briggs C, Longair I, Slavik M, et al. Can automated blood film analysis replace the manual differential? An evaluation of the CellaVision DM96 automated image analysis system. Int J Lab Hematol. 2009 Feb;31(1):48-60. http://dx.doi.org/10.1111/j.1751-553X.2007.01002.x
» http://dx.doi.org/10.1111/j.1751-553X.2007.01002.x -
7. Lee LH, Mansoor A, Wood B, et al. Performance of CellaVision DM96 in leukocyte classification. J Pathol Inform. 2013;4:14. http://dx.doi.org/10.4103/2153-3539.114205
» http://dx.doi.org/10.4103/2153-3539.114205 -
8. Ceelie H, Dinkelaar RB, van Gelder W. Examination of peripheral blood films using automated microscopy; evaluation of Diffmaster Octavia and Cellavision DM96. J Clin Pathol. 2007 Jan;60(1):72-79. http://dx.doi.org/10.1136/jcp.2005.035402
» http://dx.doi.org/10.1136/jcp.2005.035402 -
9. Kratz A, Bengtsson HI, Casey JE, et al. Performance evaluation of the CellaVision DM96 system: WBC differentials by automated digital image analysis supported by an artificial neural network. Am J Clin Pathol. 2005 Nov;124(5):770-781. http://dx.doi.org/10.1309/XMB9K0J41LHLATAY
» http://dx.doi.org/10.1309/XMB9K0J41LHLATAY -
10. UNAIDS. Fact sheet: 2014 Global statistics [document on the Internet]. c2015 [cited 2015 Sep 04]. Available from: http://www.unaids.org/sites/default/files/media_asset/20150714_FS_MDG6_Report_en.pdf.
» http://www.unaids.org/sites/default/files/media_asset/20150714_FS_MDG6_Report_en.pdf -
11. Bain B. The haematological features of HIV infection. Br J Haematol. 1997 Oct;99(1):1-8. http://dx.doi.org/10.1046/j.1365-2141.1997.2943111.x
» http://dx.doi.org/10.1046/j.1365-2141.1997.2943111.x -
12. Westgard QC. Quality requirements: Desirable biological variation database specifications [page on the Internet]. c2014 [cited 2015 Oct 24]. Available from: https://www.westgard.com/biodatabase1.htm.
» https://www.westgard.com/biodatabase1.htm -
13. Park SH, Park CJ, Choi MO, et al. Automated digital cell morphology identification system (CellaVision DM96) is very useful for leukocyte differentials in specimens with qualitative or quantitative abnormalities. Int J Lab Hematol. 2013 Oct;35(5):517-527. http://dx.doi.org/10.1111/ijlh.12044
» http://dx.doi.org/10.1111/ijlh.12044 -
14. Rollins-Raval MA, Raval JS, Contis L. Experience with CellaVision DM96 for peripheral blood differentials in a large multi-center academic hospital system. J Pathol Inform. 2012;3:29. http://dx.doi.org/10.4103/2153-3539.100154
» http://dx.doi.org/10.4103/2153-3539.100154
Accepted: 10 Dec. 2015
Published: 09 Mar. 2016