S.Afr.j.chem. (Online) vol.72  Durban  2019

http://dx.doi.org/10.17159/0379-4350/2019/v72a8 

SHORT COMMUNICATION

 

Air Sampling: A Mini-denuder Sampling Device for Direct Desorption in an Inlet of a Gas Chromatograph for the Analyses of Airborne Gaseous Phase Pollutants and Airborne Particulate Phase Pollutants

 

 

Estefan van Vuuren; Yvette Naudé*, §

Department of Chemistry, University of Pretoria, Private Bag X20, Hatfield, 0028, South Africa

 

 


ABSTRACT

We introduce a compact mini-denuder air sampling device consisting of a regular gas chromatograph (GC) inlet liner for the simultaneous concentration of separate airborne gaseous and particulate phases in a single-step. Denuder materials are thermally desorbed in a regular inlet of a GC using routine equipment available in any chromatography laboratory. The simplicity of the technique makes it ideal for resource-poor laboratories.

Keywords: Air sampling, gas chromatography, denuder, polydimethylsiloxane (PDMS), thermal desorption.


 

 

Introduction

Semi-volatile organic compounds are present in air, either adsorbed onto dust particles, or as free gas phase molecules.1 Diffusion denuder samplers are employed to sample particulate phase and gas phase air pollutants.1,2 Due to a difference in vapour and particle diffusivities diffusion denuders first separate the vapour phase to be retained on a sorptive surface from the particulate phase which passes through the denuder and is then collected on a downstream filter.1,2 Denuder materials are solvent-extracted or thermally desorbed in commercial thermal desorber systems for subsequent analyses.

Conventional diffusion denuders include annular, multi-capillary or honeycomb designs that are coated with adsorbents such as Tenax or Florisil.1 However, the challenges associated with adsorbents together with thermal desorption are incomplete desorption,3 or that the temperatures required to fully desorb the sample to the GC column can be as high as 300 °C to 350 °C resulting in decomposition of compounds. Denuder fabrication procedures are usually complicated.

Ortner and Rohwer originally developed a polydimethyl-siloxane (PDMS) multichannel trap for the analysis of semi-volatile air pollutants.4 Sample concentration by means of PDMS performs well in combination with thermal desorption.3 Contrary to adsorptive trapping, PDMS is an absorbant and acts as a hydrophobic solvent for analytes.3 The PDMS multichannel trap was since utilized as a diffusion denuder for sampling of the insecticide DDT in contaminated indoor air and atmospheric polycyclic aromatic hydrocarbons (PAHs) in mines.1,2 In these studies denuder materials were constructed in commercial thermal desorber glass tubes for desorption in thermal desorber systems with cryogenics. However, expensive thermal desorber systems and cryogenics may not be readily available in resource-poor laboratories. Motivated by economical restraints faced by laboratories in a developing country we designed a simple mini-denuder device consisting of a regular GC inlet liner. The denuder materials are made of a collection of PDMS (silicone rubber) tubelets arranged in parallel inside an inlet liner for collection of gaseous phase pollutants and a quartz fibre filter for the downstream collection of particulate phase pollutants for the simultaneous concentration of both phases in a single-step. The denuder materials are desorbed directly in an inlet of a GC without cryogenics. The GC inlet liner mini-denuder device is particularly well suited to resource-poor laboratories because routine chromatography equipment is used.

 

Experimental

i. Chemicals and Materials

For linear retention index determination an n-alkanes (C10-C28) mixture was used (Merck, South Africa). Air sampling of domestic fires made with three types of briquettes consisting of recycled paper or recycled paper with a soft wax coating or a hard wax coating (Clearflame briquettes, Pretoria, South Africa) was done using the mini-denuder.

ii. Mini-denuder and Air Sampling

The mini-denuder was made up of two multichannel silicone rubber traps separated by a quartz fibre filter which was held in position by a Teflon sleeve (Fig. 1A). Twelve silicone elastomer medical grade tubes (0.6 mm ODx0.3 mm ID, SILTEC, Technical Products, Georgia, U.S.A.) were arranged in parallel inside a regular glass inlet liner, straight through type, 6.3 mm OD, suitable for an Agilent GC (Supelco, South Africa). The PDMS multichannel trap inside the inlet liner was 57 mm long. The PDMS traps consisted of 0.36 ± 0.002 g silicone rubber and were prepared based on a technique described by Ortner and Rohwer.4 The trap had an analyte enrichment volume of 600 PDMS. Quartz fibre filters of 6 mm in diameter were punched from Ederol Quartz Microfibre sheets (0.43 mm thickness, pore size 0.3 μιτι, grade T293) from Munktell & Filtrak GmbH Bärenstein, Germany.

Air samples were collected with a portable, battery operated sampling pump (manufactured in-house).1 The pump was attached to the mini-denuder with Tygon tubing. Sampling was done for 20 min at a low flow rate of ~85 mL min-1 to minimize/ prevent potential analyte breakthrough.

iii. GC-MS

Analyses were performed using a Hewlett Packard (HP) GC 1530A coupled to a HP 5973 mass selective detector (MSD) (Chemetrix, Johannesburg, South Africa). The same instrumental conditions were used for analyses of the multichannel traps (gas phase airborne pollutants) and for the quartz fibre filters (particulate phase airborne pollutants). GC separation was performed on a HP-5 30 mx 320 μηι IDx 0.25 μηι film thickness (Agilent, Chemetrix, South Africa) column, the carrier gas (helium) velocity was set at 44 cm/s (1.4 mL/min) with a column head pressure of 1 psi in the constant flow mode. The GC oven temperature programme was 30 °C (10 min) at 10 °C min-1 to 300 °C (1 min). The GC run time was 37 min. The GC-MS transfer line was at 300 °C, the source (EI+) temperature 230 °C, the MS quadrupole temperature 150 °C, the ionization energy 70 eV and the electron multiplier (EM) was set at 1750 V Chromatograms were recorded in the full scan mode at 40-450 atomic mass units. Tentative identification of compounds detected from air sampling was performed by comparison of mass spectra to that of the Wiley mass spectral library. Analytes with a mass spectral match of 80 % or greater are reported. Experimental linear retention indices were calculated and compared to those in the literature to confirm these tentative identifications.

iv. Thermal Desorption of the Mini-denuder Device Directly in the Inlet of a GC-MS

The inlet of the GC-MS was set at 250 °C in the split mode (split 100:1). The gas flow was shut off prior to opening the inlet of the GC, the inlet liner was removed from the inlet and replaced with an inlet liner containing a multichannel trap or a quartz fibre filter (Fig. 1B). The inlet was closed, the gas flow was restored, and the run started.

The portable mini-denuder sampler contained two PDMS multichannel silicone rubber traps each inside an inlet liner of a GC which are connected in series and are separated by a quartz fibre filter (Fig. 1A). The PDMS acts as an absorbent for gaseous semivolatile analytes and the filter collects the particle phase analytes downstream from the PDMS multichannel trap. Mini-denuder materials (PDMS multichannel trap or filter) were desorbed in a regular GC inlet of a GC-MSD (Fig. 1B). Inlet liners containing adsorbents, e.g. Tenax, are commercially available. However, diffusion denuders require an open-type arrangement to allow particles to pass through. Also, as discussed earlier, there are limitations associated with thermal desorption and adsorbents.

Organic compounds from domestic fires were successfully desorbed from a PDMS multichannel trap (vapour phase) and a quartz microfibre filter (particulate phase) in the inlet of a GC without the use of cryogenics or inlet modifications (Fig. 2). The second back-up trap after the filter was analyzed to check for any breakthrough of analytes and none was detected (Fig. 3).

Figure 4 is a comparison of the gaseous phase airborne pollutants on the PDMS multichannel trap with the particulate phase airborne pollutants on the filter after a single-step, simultaneous sampling of both phases from domestic fires using briquettes made of recycled paper, or recycled paper with soft wax or hard wax coatings. Differentiation between the two phases, and as well as between the three fuel types, was achieved. As is to be expected of the particular fuel-type used in the making of the domestic fires, alkanes were present in the air samples. Here, the mini-denuder device was utilized for qualitative analyses. However, the device can easily be applied quantitatively by spiking the denuder materials with known concentrations of chemical standards of interest for the quantification of analytes in air samples.1,2

A simple mini-denuder using a regular GC inlet liner for the single-step, simultaneous sampling of airborne pollutants in both the gaseous and particle phases as discrete fractions was demonstrated. The denuder materials (PDMS tubelets and a quartzfibre filter)were easilydesorbed directlyinaregularinlet of a GC-MSDwithout the need for cryogenics orinlet modifications. Thermal desorption was achieved with routine equipment available in any chromatography laboratory. The simplicityofthe technique makes itidealforresource-poorlaboratories.

 

Acknowledgements

EgmontRohwer's originaldenuderdesignedfora commercial thermal desorber system was the inspiration for developing the inletlinermini-denuderairsamplingdevice.DavidMasemulais thanked for assembling the mini-denuders and Niel Malan for providing Clearflame briquettes.

ORCID iDs

E. van Vuuren: orcid.org/0000-0003-3534-5298 Y. Naudé: © orcid.org/0000-0002-5447-4389

 

References

1 Y. Naudé and E.R. Rohwer, Novel method for determining DDT in vapour and particulate phases within contaminated indoor air in a malaria area of South Africa. Anal Chim Acta., 2012, 730, 112-119.         [ Links ]

2 G. Geldenhuys, E.R. Rohwer, Y. Naudé and P.B.C Forbes, Monitoring of atmospheric gaseous and particulate polycyclic aromatic hydrocarbons in South African platinum mines utilising portable denuder sampling with analysis by thermal desorption-comprehensive gas chromatography-mass spectrometry. J. Chromatogr A., 2015, 1380, 17-28.         [ Links ]

3 E. Baltussen, C.A. Cramers and P.J.F. Sandra, Sorptive sample preparation - A review. Anal. Bioanal. Chem,. 2002, 373(1-2), 3-22.         [ Links ]

4 E.K. Ortner and E.R. Rohwer, Trace analysis of semi-volatile organic air pollutants using thick film silicone rubber traps with capillary gas chromatography. J. High Res. Chromatogr., 1996,19, 339-343.         [ Links ]

 

 

Received 22 November 2018