SciELO - Scientific Electronic Library Online

 
vol.63The oxidation of rhenium(III) by dioxygen in the presence of tri- and tetradentate N,O-donor ligandsSynthesis of nitrogen-doped carbon nanotubes with layered double hydroxides containing iron, cobalt or nickel as catalyst precursors índice de autoresíndice de materiabúsqueda de artículos
Home Pagelista alfabética de revistas  

Servicios Personalizados

Articulo

Indicadores

Links relacionados

  • En proceso de indezaciónCitado por Google
  • En proceso de indezaciónSimilares en Google

Compartir


South African Journal of Chemistry

versión On-line ISSN 1996-840X
versión impresa ISSN 0379-4350

S.Afr.j.chem. (Online) vol.63  Durban  2010

 

RESEARCH ARTICLE

 

Simultaneous multi-element electrothermal atomic absorption determination using a low resolution CCD spectrometer and continuum light source: The concept and methodology

 

 

Dimitri A. Katskov*; G. Eunice Khanye

Department of Chemistry and Physics, Tshwane University of Technology, Pretoria, 0001 South Africa

 

 


ABSTRACT

A low resolution CCD spectrometer with continuum light source and fast-heated graphite tube atomizer was employed for simultaneous multi-element determination by electrothermal atomic absorption spectrometry (SMET AAS). The sample vaporization pulse was monitored by fast scanning of vapour spectra within the 190-410 nm wavelength range; absorption was measured at the CCD pixels corresponding to atomic resonance lines; function absorbance vs. concentration of atomic vapour was automatically linearized, and the modified signals integrated. The setup consisted of a D2 or Xe arc lamp, a spectral instrument with a half-width of transmittance profile 120 pm, a linear CCD array attached to a PC and a tube atomizer furnished with a carbon fibre collector. In the experiments simultaneous determination of 18 elements was performed in the mixed solutions at the mg L-1 to L-1 level, within 4-4.5 orders of magnitude linear concentration range. About 1-2 min was needed for the measurement and calculation. Limits of detection (LOD) for individual elements were 1.5-2 orders of magnitude higher than those in the single element ET AAS, but similar or below those in flame AAS. Further reduction of LODs and correction of possible spectral and chemical interferences are associated with optimization of the light source and atomization programme and modification of the calculation algorithm.

Keywords: Electrothermal atomic absorption spectrometry, simultaneous multi-element determination, CCD spectrometer, fast-heated graphite tube atomizer


 

 

Full text available only in PDF format.

 

Acknowledgements

The authors would like to acknowledge the National Research Foundation, South Africa, and Tshwane University of Technology for financial support of this research.

 

References

1 J.M. Harnly, J. Anal. At. Spectrom., 1999,14, 137-146.         [ Links ]

2 B. Welz, H. Becker-Ross, S. Florek and U. Heitmann, High-resolution Continuum Source AAS: The Better Way to do Atomic Absorption Spectrometry, Wiley-VCH, Weinheim, Germany, 2005.         [ Links ]

3 http://www.analytik-jena.de (accessed 1 March 2010).         [ Links ]

4 U. Heitmann, H. Becker-Ross and D. Katskov, Spectrochim. Acta, Part B, 2006, 61, 351-360.         [ Links ]

5 B.V Lvov, Atomic Absorption Spectrochemical Analysis, Adam Hilger, Elsevier, London, UK, 1970.         [ Links ]

6 K.I. Tarasov, The Spectroscope, Adam Hilger, Bristol, UK, 1974.         [ Links ]

7 A.C.G. Mitchell and M.W. Zymansky, Resonance Radiation and Excited Atoms, Cambridge University Press, Cambridge, UK, 1961.         [ Links ]

8 B. Welz and M. Sperling, Atomic Absorption Spectrometry, 3rd edn., Wiley-VCH, Weinheim, Germany, 1999.         [ Links ]

9 http://www.OceanOptics.com (accessed 1 March 2010).         [ Links ]

10 http://www.cortec.ru (accessed 1 March 2010).         [ Links ]

11 RSA Provisional Patent Application 2008/06901.         [ Links ]

12 D.A. Katskov, J. Anal. At. Spectrom., 2005, 20, 220-226.         [ Links ]

13 D.A. Katskov, Yu. M. Sadagov and M. Banda, J. Anal. At. Spectrom., 2005, 20, 227-232.         [ Links ]

14 SIMAA 6000 Simultaneous Multi-element Atomic Absorption Spectrometer, Perkin-Elmer Bodenseewerk, Überlingen, Germany, 1995.         [ Links ]

15 http://en.wikipedia.org (accessed 1 March 2010).         [ Links ]

16 D.A. Katskov, A.M. Shtepan, I.L. Grinshtein and A.A. Pupyshev, Spectrochim. Acta, Part B, 1992, 47, 1023-1041.         [ Links ]

17 http://physics.nist.gov (accessed 1 March 2010).         [ Links ]

18 J.G. Jackson, R.W. Fonseca and J.A. Holcombe, Spectrochim. Acta, Part B, 1995, 50, 1837-1846.         [ Links ]

19 G. Tyler, VARIAN Instruments at Work. Application Note. AA or ICU: Which Do You Choose? ICP-3, Varian Inc., Mulgrave, Victoria, Australia, 3 September 1991.         [ Links ]

20 P. Heitland, Spectro ICP Report No. ICP-32, Spectro Analytical Instruments, Kleve, Germany, 2001.         [ Links ]

 

 

Received 11 February 2010
Accepted 3 May 2010

 

 

*To whom correspondence should be addressed. E-mail: katskovda@tut.ac.za

Creative Commons License Todo el contenido de esta revista, excepto dónde está identificado, está bajo una Licencia Creative Commons