vol.38 issue3An investigation into specialist practice nurses' knowledge of cardiopulmonary resuscitation guidelines in a tertiary hospital in Gauteng Province, South AfricaThe incidence and outcomes of patients with acute kidney injury in a multidisciplinary intensive care unit in Durban, South Africa author indexsubject indexarticles search
Home Pagealphabetic serial listing  

Services on Demand

Article

Indicators

Related links

  • On index processCited by Google
  • On index processSimilars in Google

Share


Southern African Journal of Critical Care (Online)

On-line version ISSN 2078-676X
Print version ISSN 1562-8264

Abstract

WILSON, K; FOURTOUNAS, M  and  ANAMOURLIS, C. A comparison of the warming capabilities of two Baragwanath rewarming appliances with the Hotline fluid warming device. South. Afr. j. crit. care (Online) [online]. 2022, vol.38, n.3, pp.97-104. ISSN 2078-676X.  http://dx.doi.org/10.7196/SAJCC.2022.v38i3.549.

BACKGROUND. Accidental intraoperative hypothermia is a common and avoidable adverse event of the perioperative period and is associated with detrimental effects on multiple organ systems and postoperative patient outcomes. In a resource-limited environment, prevention of intraoperative hypothermia is often challenging. Resourceful clinicians overcome these challenges through creative devices and frugal innovations. OBJECTIVE. To investigate the thermal performance of two Baragwanath Rewarming Appliances (BaRA) against that of the Hotline device to describe an optimal setup for these devices. METHODS. This was a quasi-experimental laboratory study that measured the thermal performance of two BaRA devices and the Hotline device under a number of scenarios. Independent variables including fluid type, flow rate, warming temperature and warming transit distance were sequentially altered and temperatures measured along the fluid stream. Change in temperature (ΔT) was calculated as the difference between entry and exit temperature for each combination of variables for each warming device. RESULTS. A total of 219 experiments were performed. At a temperature of 43.0°C and a transit distance of 200 cm, the BaRA A configuration either matched or exceeded the ΔT of the Hotline over all fluid type and flowrate combinations. The BaRA B configuration does not provide comparable thermal performance to the Hotline. Measured flowrates were noticeably slower than manufacturer-quoted values for all intravenous (IV) cannulae used. CONCLUSION. A warm-water bath at 43.0°C with 200 cm of submerged IV tubing provides thermal performance comparable to the Hotline device, with all fluid type and flowrate combinations.

Keywords : perioperative hypothermia; frugal innovation; warming devices.